MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 2 Monday, June 22, 2009 8:17 AM : IMPRINT :::::::::::::::::::::::::::::::::: All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of Siemens. Subject to change without notice. Copyright 2009 by Siemens 2 www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 3 Monday, June 22, 2009 8:17 AM CONTENTS SITRANS FX300 1 Safety instructions 6 1.1 Intended use ..................................................................................................................... 6 1.2 Certifications .................................................................................................................... 7 1.3 Safety instructions from the manufacturer .............................................................
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 4 Monday, June 22, 2009 8:17 AM CONTENTS SITRANS FX300 4 Electrical connections 31 4.1 Safety instructions.......................................................................................................... 31 4.2 Connecting the signal converter .................................................................................... 32 4.3 Electrical connection of current and pulse output ........................................................ 33 4.3.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 5 Monday, June 22, 2009 8:17 AM CONTENTS SITRANS FX300 8 Technical data 66 8.1 Functional principle........................................................................................................ 66 8.2 Technical data................................................................................................................. 67 8.3 Dimensions and weights ............................................................................................
Safety instructions MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 6 Monday, June 22, 2009 8:17 AM 1 SAFETY INSTRUCTIONS SITRANS FX300 1.1 Intended use The vortex flowmeters are made to measure the flow of gases, steam and liquids.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 7 Monday, June 22, 2009 8:17 AM SAFETY INSTRUCTIONS 1 SITRANS FX300 1.2 Certifications CE marking The device fulfils the statutory requirements of the following EC directives: • Pressure Equipment Directive 97/23/EC • Low Voltage Directive 73/23/EEC • EMC Directive 89/336/EC as well as • EN 61010 • EMC specification acc.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 8 Monday, June 22, 2009 8:17 AM 1 SAFETY INSTRUCTIONS SITRANS FX300 1.3 Safety instructions from the manufacturer 1.3.1 Copyright and data protection The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date. The contents and works in this document are subject to German copyright. Contributions from third parties are identified as such.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 9 Monday, June 22, 2009 8:17 AM SAFETY INSTRUCTIONS 1 SITRANS FX300 1.3.3 Product liability and warranty The operator shall bear responsibility for the suitability of the device for the specific purpose. The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation and operation of the devices (systems) will cause the warranty to be void.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 10 Monday, June 22, 2009 8:17 AM 1 SAFETY INSTRUCTIONS SITRANS FX300 1.3.5 Warnings and symbols used Safety warnings are indicated by the following symbols. DANGER! This information refers to the immediate danger when working with electricity. DANGER! This warning refers to the immediate danger of burns caused by heat or hot surfaces. DANGER! This warning refers to the immediate danger when using this device in a hazardous atmosphere.
Device description MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 11 Monday, June 22, 2009 8:17 AM DEVICE DESCRIPTION 2 SITRANS FX300 2.1 Scope of delivery INFORMATION! Inspect the cartons carefully for damage or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer. INFORMATION! Check the packing list to check if you received completely all that you ordered.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 12 Monday, June 22, 2009 8:17 AM 2 DEVICE DESCRIPTION SITRANS FX300 2.2.1 Devices with connection flange The measuring system consists of a measuring sensor and a signal converter. These elements form a permanent mechanical unit. Figure 2-2: Flanged devices with display 1 Version with temperature sensor 2 Version with temperature sensor and optional pressure sensor 3 Version with temperature sensor, optional pressure sensor and shut-off valve 2.2.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 13 Monday, June 22, 2009 8:17 AM DEVICE DESCRIPTION 2 SITRANS FX300 2.2.3 Device description Figure 2-4: Device description 1 2 3 4 5 6 Signal converter Cable feedthrough grey, standard version Pressure sensor, optional Shut-off valve, optional Measuring sensor Centering ring 2.2.4 Free air delivery measurement - FAD (optional) A (air) compressor draws air from the ambient atmosphere, compresses it and delivers it at the required pressure.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 14 Monday, June 22, 2009 8:17 AM 2 DEVICE DESCRIPTION SITRANS FX300 2.2.5 Gross heat meter In almost all applications with saturated steam, the steam is used for heating. It is much more interesting to know how great the heat flow volume is that is available to the process, than to know how great the flow is in kg/h. As the enthalpy of steam changes with the temperature, it cannot be assumed to be constant.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 15 Monday, June 22, 2009 8:17 AM DEVICE DESCRIPTION 2 SITRANS FX300 2.2.6 Dual seal To cover the requirements of ANSI/ISA-12.27.01-2003 “Requirements for Process Sealing Between Electrical Systems and Flammable or Combustible Process Fluid” Dual seal, a membrane vent is integrated in the neck of the instrument.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 16 Monday, June 22, 2009 8:17 AM 2 DEVICE DESCRIPTION SITRANS FX300 2.3 Nameplate INFORMATION! Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
Installation MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 17 Monday, June 22, 2009 8:17 AM INSTALLATION 3 SITRANS FX300 3.1 Notes on installation INFORMATION! Inspect the cartons carefully for damage or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer. INFORMATION! Check the packing list to check if you received completely all that you ordered. INFORMATION! Look at the device nameplate to ensure that the device is delivered according to your order.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 18 Monday, June 22, 2009 8:17 AM 3 INSTALLATION SITRANS FX300 3.4 Installation condtitions INFORMATION! For accurate volumetric flow measurement the measuring device needs a completely filled pipe and a fully developed flow profile. Please observe the instructions regarding inlet and outlet pipe runs as well as the installation position.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 19 Monday, June 22, 2009 8:17 AM INSTALLATION 3 SITRANS FX300 3.4.1 Measurement of liquids Prohibited installation Figure 3-2: Upper pipe bend CAUTION! Prohibited: Installing the device in an upper pipe bend 1, because there is a risk of gas bubbles 2 forming. Gas bubbles can lead to pressure surges and inaccurate measurement. Figure 3-3: Downpipe and outlet CAUTION! Prohibited: Installing the device in a downstream pipe 3 or upstream pipe of a outlet 4.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 20 Monday, June 22, 2009 8:17 AM 3 INSTALLATION SITRANS FX300 Recommended installations for measurement of liquids CAUTION! It is absolutely necessary to comply with the required inlet and outlet runs. 1 2 3 4 If the device is installed in a downpipe, a standpipe must be installed immediately after it. Installing the device in an inclined standpipe. Installing the device in a vertical standpipe. Installing the device in the lower pipe bend.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 21 Monday, June 22, 2009 8:17 AM INSTALLATION 3 SITRANS FX300 3.4.2 Measurement of vapours and gases Prohibited installation 1 Lower pipe bends 2 Condensate DANGER! Prohibited: Installing the device in a lower pipe bend 1, because there is a risk of condensate forming 2. Condensate can lead to cavitation and inaccurate measurement. Under certain circumstances the device can be destroyed and the measured product can leak.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 22 Monday, June 22, 2009 8:17 AM 3 INSTALLATION SITRANS FX300 3.4.3 Pipelines with control valve INFORMATION! To ensure smooth and correct measurement, the manufacturer recommends not installing the measuring device downstream from a control valve. This would run the risk of vortex formation, which would distort the measuring result.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 23 Monday, June 22, 2009 8:17 AM INSTALLATION 3 SITRANS FX300 3.4.5 Turning the connection housing DANGER! All work on the device electrics may only be carried out by appropriately trained personnel. The regional occupational health and safety directives must always be observed. CAUTION! Do not damage the electrical cable by overtwisting it. Do not remove the electrical connector.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 24 Monday, June 22, 2009 8:17 AM 3 INSTALLATION SITRANS FX300 3.4.6 Turning the display DANGER! All work on the device electrics may only be carried out by appropriately trained personnel. The regional occupational health and safety directives must always be observed. INFORMATION! If the measuring device is installed in a vertical pipe, you will have to turn the display by 90°; if installed below a pipe, turn 180°.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 25 Monday, June 22, 2009 8:17 AM INSTALLATION 3 SITRANS FX300 3.4.7 Heat insulation CAUTION! The area above the converter support must not be heat-insulated. The heat insulation 3 may only extend to the maximum height 1 shown below up to the connecting screws of the measuring sensor. Figure 3-7: Heat insulation on connection piece and signal cable 1 Max. height of insulation up to intermediate piece between measuring sensor and signal converter 2 Max.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 26 Monday, June 22, 2009 8:17 AM 3 INSTALLATION SITRANS FX300 3.5 Inlet and outlet runs 3.5.1 Minimum inlet runs 1 General inlet run without disturbing flow ≥ 20 DN 2 Behind a control valve ≥ 50 DN 3 After a pipe diameter reduction ≥ 20 DN 1 After a single bend 90° ≥ 20 DN 2 After a double bend 2x90° ≥ 30 DN 3 After a double three-dimensional bend 2x90° ≥ 40 DN 26 www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 27 Monday, June 22, 2009 8:17 AM INSTALLATION 3 SITRANS FX300 3.5.2 Minimum outlet runs Figure 3-8: Minimum outlet runs 1 Upstream of pipe expanders, pipe bends, control valves, etc. ≥ 5 DN 2 Upstream of metering points ≥ 5…6 DN INFORMATION! The interior of the pipe at the metering points must be free of burrs and other flow impediments. The measuring device has an internal temperature sensor.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 28 Monday, June 22, 2009 8:17 AM 3 INSTALLATION SITRANS FX300 3.6 Installation 3.6.1 General installation notes CAUTION! Installation, assembly, start-up and maintenance may only be performed by appropriately trained personnel. The regional occupational health and safety directives must always be observed.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 29 Monday, June 22, 2009 8:17 AM INSTALLATION 3 SITRANS FX300 3.6.2 Installing devices in flange design Figure 3-12: Installing devices in flange design 1 Gasket 2 Bolts with fixing nuts • • • • • • Use bolts and fastening nuts 2 to attach the measuring device to one side of the flange. While doing so, insert the gaskets 1 between measuring sensor and flange and align them.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 30 Monday, June 22, 2009 8:17 AM 3 INSTALLATION SITRANS FX300 3.6.3 Installing devices in sandwich design Figure 3-13: Installation using centering ring 30 1 2 3 4 5 6 Measuring sensor Centering ring Bolts with fixing nuts Hole Hole Gasket • • • • • • • • • • Push the first bolt 3 through the hole 5 of both flanges. Screw on the nuts and washers to both ends of the bolt 3 but do not tighten them. Install the second bolt through the holes 4.
Electrical connections MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 31 Monday, June 22, 2009 8:17 AM SITRANS FX300 ELECTRICAL CONNECTIONS 4 4.1 Safety instructions DANGER! All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate! DANGER! Observe the national regulations for electrical installations! DANGER! For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 32 Monday, June 22, 2009 8:17 AM 4 ELECTRICAL CONNECTIONS SITRANS FX300 4.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 33 Monday, June 22, 2009 8:17 AM SITRANS FX300 ELECTRICAL CONNECTIONS 4 4.3 Electrical connection of current and pulse output • Current output: In some cases, a shielded or twisted cable may be necessary. The cable shield may only be earthed (grounded) at one place (on the power supply unit). • Pulse output: When using the pulse output, two separate signal circuits are necessary if the pulse output is utilized together with analogue signals.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 34 Monday, June 22, 2009 8:17 AM 4 ELECTRICAL CONNECTIONS SITRANS FX300 4.3.1 Power supply INFORMATION! The supply voltage has to be between 14 VDC and 36 VDC. This is based on the total resistance of the measuring loop. To determine this, add up the resistances of each component in the measuring loop (not including the measuring device). The required supply voltage can be calculated using the formula below: Uext. = RL * 22mA + 14V where Uext.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 35 Monday, June 22, 2009 8:17 AM ELECTRICAL CONNECTIONS 4 SITRANS FX300 The pulse output is a passive "open collector" output which is electrically separated from the current interface and the measuring sensor. It can be configured as a high current output or NAMUR output using a jumper on the amplifier board. Jumper in NAMUR setting Ri = 900 Ω Umax = 36 VDC The maximum frequency of the pulse output is fmax = 0.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 36 Monday, June 22, 2009 8:17 AM 4 ELECTRICAL CONNECTIONS SITRANS FX300 4.4 Grounding connections The grounding can be done either by connecting the PE terminal in the housing or the PE terminal on the connection piece between measuring sensor and signal converter. Both of these electrical connections are equally effective from a technical point of view.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 37 Monday, June 22, 2009 8:17 AM SITRANS FX300 ELECTRICAL CONNECTIONS 4 4.5 Protection category The measuring device meets all requirements of protection category IP66/67. DANGER! After all servicing and maintenance work on the measuring device, the specified protection category must be ensured again. Figure 4-6: Cable feedthrough Therefore it is essential to observe the following points: • Use only original gaskets.
Start-up MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 38 Monday, June 22, 2009 8:17 AM 5 START-UP SITRANS FX300 5.1 Start After the device is switched on, the display shows the following in sequence 1. Testing... 2. Device type Sofware Version - Revision The device performs a self-test and switches to measurement mode. Here, all of the parameters preset for the customer are analysed and checked for plausibility, and the current measured value is displayed. 5.
Operation MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 39 Monday, June 22, 2009 8:17 AM OPERATION 6 SITRANS FX300 6.1 Display and operating elements With the cover open the device is operated by using the mechanical keys on the front; with the cover closed a bar magnet is used. CAUTION! The switching point of the magnetic sensors is directly under the glass panel above the appropriate symbol. Only touch the symbol vertically and from the front using the bar magnet.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 40 Monday, June 22, 2009 8:17 AM 6 OPERATION SITRANS FX300 The mechanical keys and keys for the bar magnet have the same functionality. In this documentation the keys are represented as symbols to describe the operating functions: Mechanical Bar magnet Symbol → ↑ ^ Table 6-1: Description of keys 6.2 Operating principles 6.2.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 41 Monday, June 22, 2009 8:17 AM OPERATION 6 SITRANS FX300 6.2.3 Navigation within the menu structure Navigation within the menu is by means of the and ^ buttons. Pressing button takes you one menu level lower, ^ takes you one menu level higher. If you are already located at the lowest level (function level), you can use the button to go in the change mode, which can be used to set data and values.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 42 Monday, June 22, 2009 8:17 AM 6 OPERATION SITRANS FX300 Example: changing the default parameter from m3/h to l/min Procedure Display Procedure Display 107.2 m3/h 5 ^ 0000600.00 L/min 1 3x 1.1.1 1 Language 6 ^ Display Unit 2 3x↑ 1.1.4 4 Max. Flow 7 ^ 1.1.4 Max. Flow 3 → m3/h Unit 8 ^ Save Yes 4 3x↑ L/min Unit 9 ^ 1787.0000 L/min 6.2.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 43 Monday, June 22, 2009 8:17 AM OPERATION 6 SITRANS FX300 Totalizer / pulse output The base units for the totalizer and the pulse output are m3 for volume, m3 norm. for standard volume and kg for mass. The unit and the conversion factor can be changed in menu item 3.2.8 "Tot.Conf.". User defined units (User Def.) can be selected and entered. The conversion factor of the unit must always be entered based on the base unit.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 44 Monday, June 22, 2009 8:17 AM 6 OPERATION SITRANS FX300 6.3 Overview of the most important functions and units INFORMATION! A complete list of all functions and short descriptions is provided in the appendix. All default parameters and settings are adapted for the specific customer. Level Designation Explanation 1.1.1 Language Select the menu language 1.1.4 Max. Flow Maximum flow rate The value set is represented by a 20 mA analogue current output.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 45 Monday, June 22, 2009 8:17 AM OPERATION 6 SITRANS FX300 Unit totalizer Volume Standard volume Mass Liquids, steams, gases Gas Liquids, steams, gases m3 L ft3/h ft3 gal ImpGal cm3 dm3 bbl User Def. m3 norm L norm ft3 std. User Def. kg t lb g User Def. Table 6-6: Totalizer units Temperature - Pressure - Power - Energy - Density units Temp. Absolute pressure (over) Power Energy Density °C °F K User Def.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 46 Monday, June 22, 2009 8:17 AM 6 OPERATION SITRANS FX300 6.4 Error messages Error message Cause Measure No Signal No signal from vortex amplifier Check connector In the event of measuring sensor problems contact service. Low freq Sample frequency too low Contact service. High Freq Sample frequency too high Contact service.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 47 Monday, June 22, 2009 8:17 AM OPERATION 6 SITRANS FX300 6.5 Menu structure 6.5.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 48 Monday, June 22, 2009 8:17 AM 6 OPERATION SITRANS FX300 6.5.2 Entering values in change mode → Moves the insertion point one position to the right; after the last position, the insertion point returns to the beginning. ↑ Cycles through available values and characters; moves the decimal point to the right. ^ Accepting the entry. 6.5.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 49 Monday, June 22, 2009 8:17 AM OPERATION 6 SITRANS FX300 6.5.4 Menu item Quick Setup Level Designation 1.1.1 Language → 1.1.2 Location → 1.1.3 Maes.Inst → 1.1.4 Selection / entry field Select menu language German ↑... German language French ↑… French language English ↑...^ English language 0000000000 Location↑…↑...↑...^ Enter the name of the location (max. 10 characters) Select type of flow measurement Volume ↑… Volume measurement Norm.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 50 Monday, June 22, 2009 8:17 AM 6 OPERATION SITRANS FX300 6.5.5 Menu item Tests Level Designation Selection / entry Explanation 2.1.1 Test I → 4 mA ^ Test current output 8 mA ^ 12 mA ^ 16 mA ^ 20 mA ^ 2.1.2 Test P → 0.5003 Hz ^ Test pulse output Table 6-10: Menu item Test Every current value has to be confirmed with ^. Only after this is the current output set to the respective value. The following is then displayed: • "Contin.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 51 Monday, June 22, 2009 8:17 AM OPERATION 6 SITRANS FX300 6.5.6 Menu item Setup (firmware version - basic) Level Designation Selection / entry 3.1 3.1.1 3.1.2 3.1.3 Set display Error Msg. → Cycle.Disp. → Code 1 → Display error Yes ↑... Shows error messages in measuring mode in plain text alternating with the measured values No ↑...^ Flashing cursor in the top left corner of the display indicates errors. Display measured values Yes ↑...
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 52 Monday, June 22, 2009 8:17 AM 6 OPERATION Level SITRANS FX300 Designation Selection / entry Explanation (available if Maes.Inst = Mass, see menu item 1.1.3) kg Unit ↑...^ Select totalizer unit for mass flow measurement 0000000000 kg ↑…↑...↑...^ Enter preset totalizer value Reset yes / Reset no ↑...^ Reset totalizer / Do not reset totalizer/ Disp. off / Disp. on ↑...^ Display totalizer / Do not display totalizer 3.3 HART settings 3.3.1 Poll.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 53 Monday, June 22, 2009 8:17 AM OPERATION 6 SITRANS FX300 Level Designation 3.5.6 Dens. Opr. → Selection / entry Explanation Density at operating pressure and operating temperature kg/m3 Unit ↑...^ Set density unit 00000.0000 kg/m3 ^ 1 Operating density Dens. Norm 2 → 00000.0000 kg/m3 ↑…↑...↑...^ Enter density for reference conditions (pressure and temperature) 3.6.1 Remote 0.000 length/m (max. = 30m) 3.6.2 Cable Coef 14.074 Cable Coef 3.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 54 Monday, June 22, 2009 8:17 AM 6 OPERATION SITRANS FX300 6.5.7 Menu item Setup (firmware version - steam) Level Designation Selection / entry 3.1 3.1.1 3.1.2 3.1.3 Set display Error Msg. → Cycle.Disp. → Code 1 → Display error Yes ↑... Shows error messages in measuring mode in plain text alternating with the measured values No ↑...^ Flashing cursor in the top left corner of the display indicates errors. Display measured values Yes ↑...
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 55 Monday, June 22, 2009 8:17 AM OPERATION 6 SITRANS FX300 Level Designation Selection / entry Explanation (available if Maes.Inst = Volume, see menu item 1.1.3) m3 Unit ↑...^ Select totalizer unit for volume flow measurement 0000000000 m3 ↑…↑...↑...^ Enter preset totalizer value Reset Yes / Reset No ↑...^ Do not reset counter/ reset counter Disp. off / Disp. on ↑...^ Do not display totalizer Display totalizer / m3 norm Unit ↑...
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 56 Monday, June 22, 2009 8:17 AM 6 OPERATION Level Designation 3.3.3 HART TV → SITRANS FX300 Selection / entry Explanation HART tertiary variable Temp. ↑... Pressure ↑... Density 3.3.4 HART 4V → ↑...^ HART quaternary variable Temp. ↑... Pressure ↑... Density ↑...^ 3.4 3.4.1 3.4.2 3.4.5 Set fluid and medium Fluid 1 → Medium → Dry. Fact. → (available if Medium = Sat.Steam, see menu item 3.4.2) Fluid type Steam ^ Steam Process medium Sat.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 57 Monday, June 22, 2009 8:17 AM OPERATION 6 SITRANS FX300 Level Designation 3.5.6 Dens. Opr. → 3.5.10 P-Excit. V → (available if P-Sensor = Internal, see menu item 3.5.2) 3.5.11 P-Sen.P1V1 → (available if P-Sensor = Internal, see menu item 3.5.2) Selection / entry Explanation Density at operating pressure and operating temperature kg/m3 Unit ↑...^ Set density unit 00000.0000 kg/m3 ^ 1 Operating density 0005.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 58 Monday, June 22, 2009 8:17 AM 6 OPERATION SITRANS FX300 6.5.8 Menu item Setup (firmware version - gas) Level Designation Selection / entry 3.1 3.1.1 3.1.2 3.1.3 Set display Error message → Cycle.Disp. → Code 1 → Display error Yes ↑... Shows error messages in measuring mode in plain text alternating with the measured values No ↑...^ Flashing cursor in the top left corner of the display indicates errors. Display measured values Yes ↑...
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 59 Monday, June 22, 2009 8:17 AM OPERATION 6 SITRANS FX300 Level Designation Selection / entry Explanation (available if Maes.Inst = Mass, see menu item 1.1.3) kg Unit ↑...^ Select totalizer unit for mass flow measurement 0000000000 kg ↑…↑...↑...^ Enter preset totalizer value Reset Yes / Reset No ↑...^ Reset totalizer Do not reset totalizer/ Disp. off / Disp. on ↑...^ Do not display totalizer Display totalizer / 3.3 HART settings 3.3.1 Poll.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 60 Monday, June 22, 2009 8:17 AM 6 OPERATION SITRANS FX300 Level Designation Selection / entry 3.4.2 Medium (available if Fluid = Gas / Wet Gas, see menu item 3.4.1) Air ↑... Air Ammonia ↑... Ammonia ↑... Argon ↑... Argon Set process medium etc. 3.4.3 % Gas (available if Fluid = Gas Mix., see menu item 3.4.1) 3.4.4 %Rel.Hum . (available if Fluid = Wet Gas, see menu item 3.4.1) 3.4.6 FAD Unit.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 61 Monday, June 22, 2009 8:17 AM OPERATION 6 SITRANS FX300 Level Designation Selection / entry 3.5 3.5.1 Pressure, temperature and density T-sensor → Internal temperature sensor No ↑... Yes 3.5.2 P-sensor → 3.5.5 3.5.6 Temp. Opr. → Pres. Opr. → No temperature sensor available ↑...^ Temperature sensor available Pressure sensor Intern ↑... --- 3.5.4 Explanation Internal pressure sensor ↑...
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 62 Monday, June 22, 2009 8:17 AM 6 OPERATION Level Designation 3.5.12 P-Sen.P2V2 (available if P-Sensor = Internal, see menu item 3.5.2) SITRANS FX300 Selection / entry Explanation Pressure sensor: 2nd calibration point 0005.00000 P2 kg/cm2g ↑...^ ↑…↑... 0048.00048 V2 mV ↑…↑...↑...^ 3.6 3.6.1 Remote 0.000 length/m 3.6.2 Cable coeff 14.074 Cable Coef 62 (max. = 30m) www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.
Service MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 63 Monday, June 22, 2009 8:17 AM SERVICE 7 SITRANS FX300 7.1 Replacing signal converter / LCD display The signal converter has to be replaced by a converter of the same type. The following parameters have to be observed for it: • The item number must match: 2.143670.xxx • • • • The software version 2 must match.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 64 Monday, June 22, 2009 8:17 AM 7 SERVICE SITRANS FX300 7.2 Spare parts availability The manufacturer adheres to the basic principle that operational spare parts for each device or each important accessory part will be kept available for a period of 10 (ten) years after delivery of the last production run for that device. Operational spare parts are defined as parts that are subject to faults in normal operation. 7.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 65 Monday, June 22, 2009 8:17 AM SERVICE 7 SITRANS FX300 7.4.2 Form (for copying) to accompany a returned device Company: Address: Department: Name: Tel. no.: Fax no.: Manufacturer's order no. or serial no.: The device has been operated with the following medium: This medium is: water-hazardous toxic caustic flammable We checked that all cavities in the device are free from such substances.
Technical data MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 66 Monday, June 22, 2009 8:17 AM 8 TECHNICAL DATA SITRANS FX300 8.1 Functional principle Vortex flowmeters are used to measure the volumetric flow of gases, vapours and liquids at completly filled pipes. The measuring principle is based on the Karman vortex street. The measuring tube contains a bluff body, behind which vortex shedding occurs. The frequency f of the vortex shedding is proportional to the flow rate v.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 67 Monday, June 22, 2009 8:17 AM TECHNICAL DATA 8 SITRANS FX300 8.2 Technical data INFORMATION! • The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local representative. • Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website .
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 68 Monday, June 22, 2009 8:17 AM 8 TECHNICAL DATA SITRANS FX300 Operating conditions Temperature Product -40…+240°C / -40…+465°F Ambient Non-Ex: -40…+85°C / -40…+185°F Ex: -40…+60°C / -40…+140°F Storage -50…+85°C / -58…+185°F Pressure Product Max. 100 bar / 1450 psi; Information on higher pressures on request. Ambient Atmosphere Chemical properties Density Taken into consideration when sizing. Viscosity < 10 cP Reynold's number 10000...
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 69 Monday, June 22, 2009 8:17 AM TECHNICAL DATA 8 SITRANS FX300 Process connections Flange version DIN EN 1092-1 DN15...300 in PN16…100 ASME B16.5 ½...12" in 150…600 lbs JIS B 2220 DN15...300 in JIS 10…20 K For detailed information on combination flange/pressure rating, refer to chapter "Dimensions and weights". Sandwich version DIN DN15...100 in PN100 (higher pressures on request) ASME ½...4" in 600 lbs (higher pressures on request) JIS DN15...
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 70 Monday, June 22, 2009 8:17 AM 8 TECHNICAL DATA SITRANS FX300 8.3 Dimensions and weights 8.3.1 Flange versions Option: version with 2 transmitters a = 133 mm / 5.24" b = 105 mm / 4.13" c = 179 mm / 7.05" Dimension H x 2 Specified weight + 2.80 kg Flange version EN 1092-1 Size Pressure rating DN PN Dimensions [mm] d D L Weight [kg] H l With Without pressure sensor 70 15 40 17.3 95 200 265 174.3 6.1 5.5 15 100 17.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 71 Monday, June 22, 2009 8:17 AM TECHNICAL DATA 8 SITRANS FX300 Size Pressure rating DN PN Dimensions [mm] d D L Weight [kg] H l With Without pressure sensor 100 100 104.3 265 250 310 174.5 36 35.4 150 16 159.3 285 300 325 196.5 35.8 35.2 150 40 159.3 300 300 325 196.5 41.8 41.2 150 63 157.1 345 300 325 196.5 59.8 59.2 150 100 154.1 355 300 325 196.5 67.8 67.2 200 10 206.5 340 300 350 208.5 38.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 72 Monday, June 22, 2009 8:17 AM 8 TECHNICAL DATA SITRANS FX300 a = 133 mm / 5.24" b = 105 mm / 4.13" c = 179 mm / 7.05" Flange version ASME B16.5 Size Pressure rating NPS Class Dimensions [mm / inches] d D L H Weight [kg / lbs] l With Without pressure sensor 72 ½ 150 15.8 / 0.6 90 / 3.5 200 / 7.9 265 / 10.4 174.3 / 6.9 5.1 / 11.2 4.5 / 9.9 ½ 300 15.8 / 0.6 95 / 3.7 200 / 7.9 265 / 10.4 174.3 / 6.9 5.5 / 12.1 4.9 / 10.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 73 Monday, June 22, 2009 8:17 AM TECHNICAL DATA 8 SITRANS FX300 8.3.2 Sandwich version a = 133 mm / 5.24" b = 105 mm / 4.13" c = 179 mm / 7.05" Dimension H x 2 Specified weight + 2.80 kg Sandwich version EN Size Pressure rating DN PN Dimensions [mm] d D L Weight [kg] H l With Without pressure sensor 15 100 16 45 65 265 174.25 4.1 3.5 25 100 24 65 65 265 174.25 4.9 4.3 40 100 38 82 65 270 174.5 5.5 4.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 74 Monday, June 22, 2009 8:17 AM 8 TECHNICAL DATA SITRANS FX300 a = 133 mm / 5.24" b = 105 mm / 4.13" c = 179 mm / 7.05" Sandwich version ASME Size Pressure rating NPS Class Dimensions [inches] d D L H Weight [lbs] l With Without pressure sensor 74 ½ 150 0.63 1.77 2.56 10.43 6.82 9.04 7.72 ½ 300 0.63 1.77 2.56 10.43 6.82 9.04 7.72 ½ 600 0.55 1.77 2.56 10.43 6.82 9.04 7.72 1 150 0.94 2.56 2.56 10.43 6.82 10.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 75 Monday, June 22, 2009 8:17 AM TECHNICAL DATA 8 SITRANS FX300 8.4 Flow tables Measuring ranges Size DN EN 1092-1 Qmin NPS ASME B16.5 Qmax Qmin Qmax [gph] [m3/h] Water 15 ½ 0.45 5.0 119 1321 25 1 0.81 11.40 214 3012 40 1½ 2.04 28.57 539 7547 50 2 3.53 49.47 933 13069 80 3 7.74 108.37 2045 28629 100 4 13.30 186.21 3514 49192 150 6 30.13 421.86 7960 111445 200 8 52.66 792.42 13911 209335 250 10 90.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 76 Monday, June 22, 2009 8:17 AM 8 TECHNICAL DATA SITRANS FX300 Flow rate limits Product Nominal sizes EN ASME Liquids DN15…300 ½…12" Gas, steam DN15…300 ½…12" Minimum flow rates [m/s] Maximum flow rates [ft/s] [m/s] [ft/s] 0.5x(998/ρ)0.5 1 1.64x(998/ρ)0.5 2 7x(998/ρ)0.47 1 23x(998/ρ)0.47 2 16.4x(1.29/ρ)0.5 7x(998/ρ)0.47 3 23x(998/ρ)0.47 4 6x(1.29/ρ)0.5 3 4 ρ = operating density [kg/m3] 1 2 3 4 76 Minimum flow rate 0.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 77 Monday, June 22, 2009 8:17 AM TECHNICAL DATA 8 SITRANS FX300 Measuring range saturated steam: 1...7 bar Overpressure [bar] 1 3.5 5.2 7 Density [kg/m³] 1.13498 2.4258 3.27653 4.16732 Temperature [°C] 120.6 148.2 160.4 170.6 Flow rate min. max. min. [kg/h] max. min. [kg/h] max. min. [kg/h] max. DN EN 1092-1 NPS ASME B16.5 [kg/h] 15 ½ 5.87 36.97 7.68 79.0 8.93 106.68 10.06 135.69 25 1 11.82 129.39 17.28 276.40 20.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 78 Monday, June 22, 2009 8:17 AM 8 TECHNICAL DATA SITRANS FX300 Measuring range saturated steam: 15...100 psig Overpressure [psig] 15 50 75 100 Density [lbs/ft³] 0.0719 0.1497 0.2036 0.2569 Temperature [°F] 249.98 297.86 320.36 338.184 Flow rate min. max. [lbs/h] min. max. [lbs/h] min. max. [lbs/h] min. max. [lbs/h] [lbs/h] DN EN 1092-1 NPS ASME B16.5 15 ½ 12.9 82.70 16.83 1720.12 19.62 234.0 22.04 295.23 25 1 26.
Notes MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 79 Monday, June 22, 2009 8:17 AM NOTES 9 SITRANS FX300 06/2009 - SFIDK.PS.050.F2.02 A5E02100423 www.siemens.