Specifications

Section-5 Maintenance
October 17, 2013
56
Copyright©2013,HarrisBroadcast
WARNING:Disconnectprimarypower priortoservicing. 8882693004
SeeSection2.13onpage221andfollowingsectionsforLPUandTCUwebbrowserconnectionandlogin
information.Harrisr
ecommendsuseoftheChromeorFirefoxwebbrowser.
Calibrateusi
ngthelocalTCUGUIorwebbrowserbynavigatingtotheLPUUAXHome>Config>System
Configuration>Cal’SystemCalibration’screenandpressthepulldownmenuarrowtoshowthecalibrationoptions
(seeFigure53onle
ftbelow).
Forwardc
alibrationmustbeperformedfirstandisfollowedbyreflectedcalibration.Afterselectingthecalibration
type,presstheStartCalsoftkeytobeginthecalibrationprocess(seeFigure53onrigh
tbelow).
Figure 5-3 on
left - System Calibration options on right - System Calibration started
Powerc
alibrationsinclude:
Forward‐Powerattheoutputofthetransmitteramplifyingstages,beforethebandpassfilter(BPF).
Reflected‐Poweratthe outputofthetransmitteramplifyingstagesandbeforethebandpassfilter(BPF).
UDCOut‐PoweratoutputofLPUmodulatorstage(SMAconnectoronupperrearofLPU)
LPUForward‐Powerattheoutput(typeN)oftheLPUamplifiersection
PAB#Forward‐Powerattheoutputofindividualpoweramplifierblocks
PAB#Reflected‐Powerattheoutputofindividualpoweramplifierblocks
Note
Re-calibration is not typically required in the field once the transmitter is in service. Re-calibration would
be indicated only for verification, frequency change, or after major service work involving replacement of
one or more power metering components has been performed.
EquipmentUsed:
AnHPEPMseriesPowerMeterand8482HPowerSensor,orAgilentE4418BandType“N”toSMAadapter(or
equivalent).MetershouldbecapableofdisplayingmeasuredvaluesinWattsorindBm.
Precisiondi rectional couplers(precisioninthesensethatthecouplingratioshavebeenmeasuredforthe
operatingfrequenciestoahighdegreeofprecision).
Caution
THE REFLECTED POWER FAULTS ARE DISABLED DURING THE POWER
CALIBRATION PROCEDURE. EXIT FROM THE CALIBRATION PROCEDURE
WHEN CALIBRATION IS COMPLETE TO AUTOMATICALLY REACTIVATE
REFLECTED POWER PROTECTIONS.
Note
Read and understand the applicable power calibration procedure before beginning. The procedures below
assume use of the LCD display and the buttons on the front of the LPU. The procedure can also be per
-
formed using the web browser screens. Screen captures, such as seen in Figure 5-3 on page 5-6, should be
assumed to be via web browser.