Instruction manual

4-4
4
CHECKING AND
ADJUSTMENT
2) Adjustment
Press the [MENU] and [7] key to enter
the calibration screen.
<DTM-350>
Take a F1 measurement to a target on
the horizon and press the [ENT] key.
Vertical angle is shown in “V0 dir=
Horiz” setting in this function.
V1: Face-left vertical angle (tilt-off value)
H1: Face-left horizontal angle (tilt-off value)
X1: Face-left X axis tilt value
Y1: Face-left Y axis tilt value
Upon finishing the measurement, the
message on the bottom line is changed
from “DO NOT TOUCH!” to “Turn to
FACE2.
Take a F2 measurement to the same
target and press the [ENT] key.
V2: Face-right vertical angle (tilt-off value)
H2: Face-right horizontal angle (tilt-off value)
X2: Face-right X axis tilt value
Y2: Face-right Y axis tilt value
When the observation on F2 is completed
four parameters are displayed.
[1:Redo]/[ESC] returns to the first
observation screen.
[4:OK]/[ENT] sets those parameters on
the instrument.
1:Job 5:Comms
ogo :T
ett :C
ata :N
X: 70″ Y: -10″
ACV: 23″
ACH: 20″
X1 38″1: 1″
V1: 0°26′40
H1: 79°00′50″
Ai Horiz EN
X1: 18″Y1 21″
V1: 0°26′4″
H1: 79°00′52″
Turn to CE
X2: 10″Y2: -55″
V2:179°33′25″
H2:281°04′30″
Ai Hor NT