Specifications

TEC modules Two stages
Maximal T
>50 °C below ambient
Regulated T
48 °C below ambient (85% cooling)
Regulation precision
±0.1 °C
Hot side cooling Forced air cooling (fan)
Optional heat exchanger for liquid coolant
Maximum temperature difference between CCD and ambient air may exceed
50 °C when the cooling runs at 100% power. However, temperature cannot be
regulated in such case, camera has no room for lowering the CCD temperature
when the ambient temperature rises. The 45 °C temperature drop can be
achieved with cooling running at approx. 85% power, which provides enough
room for regulation.
Power Supply
The 12 V DC power supply enables camera operation from arbitrary power
source including batteries, wall adapters etc. Universal 100-240 V AC/50-
60 Hz, 60 W “brick” adapter is supplied with the camera. Although the camera
power consumption does not exceed 30 W, the 60 W power supply ensures
noise-free operation.
Camera head supply 12 V DC
Camera head power consumption 30 W
Adapter input voltage 100-240 V AC/50-60 Hz
Adapter output voltage 12 V DC/5 A
Adapter maximum power 60 W
1. Maximum power consumption includes 100% cooling.
2. The camera contains its own power supplies inside, so it can be
powered by unregulated 12 V DC power source – the input voltage
can be anywhere between 10 and 14 V. However, some parameters
(like cooling efficiency) can degrade if the supply drops below 12 V.
3. G2 CCD camera measures its input voltage and provides it to the
16