Operating instructions

15
Data Transfer and Statistical Evaluation
in the Automatic Mode
Data Transfer
The first value transferred must be greater than 30 scale
intervals of the weighing instrument’s display (e.g., > 3.0
g, if the weighing instrument displays decigrams).
The weighing instrument must display a stability symbol
(fluctuation > 1 scale interval). The first value read into
the statistics memory is used as the reference value for the
statistics. Afterwards, data transfer will be blocked until
it is detected that the weighing instrument has been
unloaded. This means that the load on the pan is below
<30% of the reference value.
The next value will be transferred and read into the
statistics memory if it is > 70% of the reference value.
Required Printer Configuration
You must set the following printer configuration parame-
ters in the “Statistics Applications” menu:
Statistics program = 1 (-a3-)
Data transfer automatic = 1 (-P6-)
Required Weighing Instrument Configuration
The weighing instrument must be configured as follows
(see instructions for the weighing instrument connected):
Auto print
Data output without stability