Note: Only if a limit value is defined, a limit check is performed for the harmonic (see
HARMonic:RESult<m>:HARMonics<n>:LIMCheck? auf Seite 375).
Tip: To query the result of the limit check for the entire signal, use HARMonic:
RESult<m>:LIMViolation? auf Seite 378.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
<n> 1..64
Order of harmonic (possibly restricted, see HARMonic:
RESult<m>:NOHarmonics?)
Rückgabewerte:
<State> ON | OFF
ON
A limit violation occurred - the limit was exceeded.
OFF
No limit violation occurred.
Beispiel:
HARM:LIM:TYPE EN50160
HARM:RES1:HARM2:LIMC?
//Result: 1
//The EN50160 defines a limit of 2% for the
//second order harmonic.
HARM:RES1:HARM2:LIM?
//Result: 2
HARM:RES1:HARM2:LIMV?
//Result: 1
//The calculated value exceeds the defined
// limit for the second harmonic.
HARM:RES1:LIMV?
//Result: 1
//Since the limit check for the second harmonic
// failed, the limit check for the entire
// signal failed.
Verwendung:
Nur Abfrage
HARMonic:RESult<m>:HARMonics<n>:PHASe?
Queries the calculated phase value for the specified harmonic.
Suffix:
<m>
.
1|2 (RTH1002) , 1..4 (RTH1004)
<n> 1..64
Order of harmonic (possibly restricted, see HARMonic:
RESult<m>:NOHarmonics?)
Rückgabewerte:
<Phase> Std-einheit: degrees
Verwendung: Nur Abfrage