Technical Specs

24
Guidance and manufacturer’s declaration electromagnetic immunity
The leva PDHS is intended for use in the electromagnetic environment specified below. The
customer or the user of the leva PDHS should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601
test level
Compliance level
Electromagnetic environment
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV
contact
±8 kV air
±6 kV contact
±8 kV air
1
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply lines
±1 kV for
input/output lines
Not applicable
2
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1kV
line(s) to line(s)
±2 kV
line(s) to earth
Not applicable
3
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
line
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Not applicable
3
Mains power quality should be that of
a
typical commercial or hospital
environment
. If the user of the leva
PDHS
requires continued operation
during
power mains interruptions, it is
recommended
that the leva PDHS be
powered
from an uninterruptible
power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
1. If the system stops communicating during use, it is recoverable by turning off and on again.
2. I/O cable is less than 3meters.
3. The system is powered by internal battery only.
Electromagnetic Interference
Wireless communications equipment such as wireless home network devices,
mobile phones, and cordless telephones and their base stations can affect
leva PDHS. Keep leva PDHS away from other wireless equipment based on
the table below: