User manual no.: LMI-53-05/11/12/A BALANCES APP/2C SERIES MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG 26 – 600 Radom, Bracka 28, POLAND phone +48 48 38 48 800, fax. +48 48 385 00 10 e-mail: export@radwag.com www.radwag.
NOVEMBER 2012 -2-
TABLE OF CONTENTS 1. 2. 3. 4. 5. 6. 7. 8. 9. TECHNICAL DATA ............................................................................... 5 BASIC INFORMATION ......................................................................... 6 2.1. 2.2. 2.3. 2.4. 2.5. 2.6. Intended use............................................................................................... 6 Inappropriate use........................................................................................ 6 Warranty .............
12.3.1. Balances with internal adjustment system.................................. 41 12.3.2. Balances with external adjustment system................................. 42 12.4. Manual adjustment ................................................................................... 43 12.4.1. Internal adjustment..................................................................... 43 12.4.2. External adjustment ................................................................... 44 12.5.
1. TECHNICAL DATA Balance with external adjustment system Balance with internal adjustment system APP 10/2C/1 APP 25/2C/1 APP 30/2C/1 APP 35/2C/1 APP 6/35/2C/1 APP 50/2C/1 APP * 10/2C APP * 25/2C APP 30/2C APP 35/2C APP * 6/35/2C - Max.
2. BASIC INFORMATION 2.1. Intended use Precision balance APPC series is designed to determine the weighing value of loads in laboratory environment. It is intended for application as a nonautomatic weighing instrument only, i.e. the material to be weighed is manually and carefully placed in the center of the weighing pan. Weighing result should be read only after stable reading has been obtained – stability pictogram is visible on balance’s display. 2.2.
2.4. Monitoring metrological parameters of the instrument Metrological characteristics of the balance requires periodical inspection carried out by its operator. Inspection frequency is conditioned by ambient conditions in which the balance is used, types of performed processes and accepted quality management system in an organization. 2.5.
4. UNPACKING, ASSEMBLING AND STARTUP 4.1.
• • Remove the weighing pan from balance’s housing and remove the transport protections, as shown on above figure. After unpacking, unscrew bolt (1) locking balance’s adjustment mechanism. • Assembly the weighing pan. Pay attention to correct positioning of the weighing pan on balance’s mandrels – the weighing pan must not touch balance’s housing, and it should rest steadily.
• Level the balance by turning its adjustable feet. Then using a wrench (3) tighten up the nut (2) on the adjustable foot (1) to balance’s basis until refusal point. Setting Before plugging to mains, level the balance using adjustable feet. Turn the feet in a way that the air bubble of the level is located centrally. • Only then plug the balance to mains (power socket is located at the back of balance’s housing). 4.3. • • • • 4.4. Standard delivery components Balance. Weighing pan. Power adapter.
4.6. Connecting peripheral equipment The balance must be unplugged from the mains before connecting or disconnecting any peripheral equipment (printer, PC computer). Use only peripheral equipment recommended by the manufacturer with your balance. These have been ideally coordinated to your balance. On connecting a peripheral device, plug the balance to mains. 5.
Balance APPC series can be used to weigh loads under the weighing platform – i.e. under hook weighing option. Balance software enables printing a report from adjustment process containing a set of data on adjustment conditions and criteria (see GLP procedures).
6. BALANCE KEYBOARD ON/OFF key enables switching on and off balance’s display. If switched off balance components other than the display are powered, and balance is in stand-by mode. F key. Function key, which enables quick entering the settings of an active working mode. MODE key for selecting balance’s working mode. UNITS key, changes measuring units. PRINT/ENTER key - Sends current display status to a peripheral device (PRINT) or accepts selected value of a parameter or function (ENTER).
7. START-UP When plugged to mains, the balance displays name and number of software, and afterwards proceeds to weighing mode. Balance temperature stabilization period. Before start of measuring processes, it is necessary to wait until the balance is thermally stabilized. It is a period of so called balance self-heating. In case of balances APP series self-heating period takes approximately 15 minutes.
8. USER MENU Balance menu is divided into 7 basic function groups. Each group has individual name. 8.1. Moving through user menu An operator moves through balance menu using keys on overlay’s keyboard: Setup key. Entering balance’s main menu. Selecting group of parameters one by one downwards or changing parameter value by one value downwards. Selecting group of parameters one by one upwards or changing parameter value by one value upwards. Selecting group of parameters for activating.
Name of function groups and their content: P1 CAL [Adjustment/Calibration] P1.1 P1.2 P1.3 P1.4 P1.5 P1.6 | | | | both/nonE/tenno/tinnE | 1 h ÷ 12h | YES/no iCAL ECAL tCAL ACAL CALt CALr P2 GLP P2.1 P2.2 P2.3 P2.4 P2.5 P2.6 P2.7 P2.8 USr PrJ Ptin PdAt PUSr PPrJ PId PFr P3 rEAd P3.1 AuE P3.2 ConF P3.3 Auto P3.4 Ldi9 P3.5 rAPd P4 Print P4.1 bAud P4.2 CntA P4.3 Cntb P4.4 rEPL P4.5 PStb P4.
P5 Unit P5.1 StUn P5.2 mg P5.3 └g P5.4 Ct P5.5 lb P5.6 oZ P5.7 oZt P5.8 dwt P5.9 t P5.10 nno P5.11 Gr [Measuring units] | g/mg/kg/ct/lb/oz/ozt/dwt/t/mom/G [basic measuring units – e.g. gram] | YES/no [mg - milligram] | YES/no [kg– kilogram] | YES/no [ct – carat] | YES/no [lb – pound] | YES/no [oz – ounce] | YES/no [ozt – ounce troy] | YES/no [dwt – pennyweight] | YES/no [t – tael] | YES/no [mom - momme] | YES/no [gr – grain] [Working modes] P6 Func P6.1 FFun P6.2 PcS P6.3 HiLo P6.4 PrcA P6.5 Prcb P6.6 P6.
8.2. Return to weighing mode CAUTION Changes introduced in balance memory will be saved on return to weighing with procedure of saving changes. Press ESC/TARE key for several times, until the display shows command SAvE?. When display indicates the above command for saving changes, press one of below options: PRINT/ENTER – save changes or ESC/TARE – abandon introduced changes, and leave to balance menu. After pressing an appropriate key, balance goes back to weighing mode. 9.
• mass indication of a load placed on balance’s weighing pan can be tarred for multiple times by pressing ESC/TARE key ((pay attention not to exceed maximal capacity of a balance by applying multiple tare function), During times between carrying out the following measurement series do not unplug the balance from mains. It is recommended to switch off balance’s display by pressing ON/OFF key.
press key for a few times. The display indicates available measuring units in turns After selecting the basic measuring unit press PRINT key. The balance returns to displaying parameter name. Available settings: • • For verified balances, a user can select from the following measuring units: [g], [kg], [ct] For non-verified balances, a user can select from the following measuring units: [g], [kg], [ct], [lb], [oz], [ozt], [dwt], [t], [mom], [gn].
Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing). CAUTION: On switching on, the balance will start operating with set basic measuring unit. 9.2. Temporary measuring unit Function enables selecting a measuring unit which will be indicated next to mass readout during weighing process. Set measuring unit will be in use from the moment of its activation until its change or switching off and on the balance.
- 22 -
9.3.1. Selection of working modes available for a user Function enables selecting (by pressing MODE key) which working modes are accessible during weighing function – if all – then select (ALL) and if only one from all available (described below). After setting the parameter press PRINT key. Balance goes back to displaying submenu name P6.1 FFun.
If a mode other than ALL is selected, procedure of returning to weighing mode differs from other ones. If for instance PcS mode is selected, follow procedure in accordance with point 14.1 point, to return to weighing: Press Esc/TARE key, balance returns to weighing. 9.3.2. Means of selecting one of available working modes Balance user can enable or disable modes which are not used in the weighing process. This provides quicker access to modes used on a balance.
YES - mode enabled; no - mode disabled On selecting, press PRINT key. The balance returns to displaying submenu name. Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing). 9.4. Dual range balance (APP 6/35/2C) Balance APP 6/35/2C series is a dual range weighing instrument. The accuracy of the I weighing range is d1=1g, and of the II weighing range it is d2=5g.
The balance returns to weighing with the accuracy of the I weighing range d1=1g and II weighing range pictogram is blanked. 10. BALANCE MAIN PARAMETERS It is possible to adjust balance’s operation to ambient conditions at a workstation (by setting filter range and value release parameters) or user requirements (autozero settings, last digit visibility). These parameters are set in group P3 rEAd. 10.1.
10.2. Value release Due to various ambient conditions at workstations, and in order to set balance to present conditions, it is necessary to determine the value release parameter as: FAST_rEL, Fast or rEL. Depending on selected option, weighing time will be shorter or longer. Fast_rEL Fast rEL - fast + reliable - fast - reliable. Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing).
10.3. Autozero function In order to ensure balance’s precise mass indication, autozero (Auto) software parameter has been introduced. The application of this function is automatic control and correction of zero indication. When function is enabled, it compares balance indications at declared time interval e.g. 1 s, on conditions that weighing pan is unloaded and display indication is close to zero. If results vary less than declared AUTOZERO range e.g.
10.4. Blanking of the last digit on the display Function enables blanking of the last displayed digit. ALuuAYS nEuEr uu_StAb - always - never - when measurement result is stabilized. Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing).
10.5. Weighing mode RAPID Weighing mode RAPD is located in menu parameter P3.rEAd and it offers two options: RAPID = NO – balance operates with settings as set in menu AUE and CONF, i.e. standard weighing.
11. OTHER PARAMETERS Balance APPC series features a set of parameters which enable influencing its functionality. The parameters are contained in group: P7 othEr and they are: backlight and “beep” signal and printout of all balance parameters through RS 232 port. In order to edit the above parameters enter menu group P7 Other. 11.1.
(see point - 8. 2. – return to weighing). CAUTION: Balance software has implemented option of automatic backlight switch off bl = Aut. In case of this setting, the backlight is switched automatically off if weighing results does not change within 10 seconds. Backlight is automatically activated at the moment of weighing result change on balance display. 11.2.
11.3. OFF On “Beep” sound – reaction on pressing function key - signal of pressing function key disabled - signal of pressing function key enabled. Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing).
11.4. Printout of all balance parameters through RS 232 port After pressing F key balance parameters are sent to a peripheral device connected to balance’s RS 232 port. Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing).
12. BALANCE ADJUSTMENT As the gravitational acceleration force has different value at various latitudes and altitudes, a balance should be adjusted to present working conditions. This process should take place at the first installation of the instrument on site and at its every relocation – as in case of ambient temperature changes.
- balance software detects necessity of carry out adjustment and signalizes it by Celsius degree pictogram or by black colour pictogram on balance display, - since pictogram occurrence, time interval of 5 minutes is counted within which weighing procedure can be completed, - as the 5 minutes time elapses, balance display indicates message CAL_30 and starts counting down from 30..29..
• adjustment triggered by elapsing time. It is possible to declare time intervals which are a criteria for balance adjustment. Accessible settings are: adjustment after 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12 hours since last carried out adjustment procedure, CAUTION Declaring time interval for balance adjustment is available only in non-verified balances. In case of verified balances the time interval between following adjustments is set for 3 hours. 12.2. Menu adjustments settings P1 CAL P1.1 P1.2 P1.
• P1.2 ECAL – External adjustment (with an external weight) adjustment with external weight, which value is saved in factory menu of balance, function unavailable in verified balances. balances with internal adjustment • balances with external adjustment P1.3 tCAL – Adjustment test mass comparison of internal adjustment weight with its value saved in balance memory. balances with internal adjustment • balances with external adjustment P1.
nonE tEmP timE both – none of factors will cause start of adjustment – adjustment triggered by change of temperature – adjustment triggered by time interval set in – adjustment triggered by time and temperature.
• P1.5 CALt - Time of automatic adjustment (balances with internal adjustment system) Determination of time interval, after which automatic adjustment process is initiated. • P1.6 CALr – printout of adjustment report Printout activated on completion of automatic adjustment procedure. balances with internal adjustment no YES balances with external adjustment – report printout disabled – report printout enabled Go back to weighing mode with procedure of saving changes (see point 8. 2.
12.3. Adjustment test 12.3.1. Balances with internal adjustment system Adjustment test is a comparison of internal adjustment weight with its value stored in balance’s memory. This process is automatic and its result is shown on the display (if a balance is connected to a computer or a printer through RS 232 socket, data on adjustment test is printout). Press ESC/TARE key to return to previous display. Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing).
12.3.2. Balances with external adjustment system It is mass comparison of external adjustment weight with its value stored in balance’s memory during adjustment process. This process is automatic and its result is shown on the display (if a balance is connected to a computer or a printer through RS 232 socket, data on adjustment test is printout). Press ESC/TARE key to return to previous screen. Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing).
12.4. Manual adjustment 12.4.1. Internal adjustment Press Cal key or follow below guidelines The balance performs internal adjustment automatically. During adjustment process do not load the weighing pan with any load. On completion of adjustment process, balance saves its result in memory and returns to weighing mode.
12.4.2. External adjustment External adjustment in balances APPC series should be carried out with an external mass standard / weight class F2 Start of external adjustment process, Balances with internal adjustment balances with external adjustment Balance will display a command ordering unloading of weighing pan unloAd (weighing pan should be empty). When weighing pan is unloaded, press PRINT key. The balance determines mass of empty pan.
no YES . – report is not printed – report is printed Remember, that if the parameter is set for YES, then a report is generated and sent automatically. Content of report from adjustment process depends on settings of GLP parameters. Any option in the GLP submenu which attribute is YES is included in a report from adjustment process. P2 GLP – Good Laboratory Practice P2.1 P2.2 P2.3 P2.4 P2.5 P2.6 P2.7 P2.
13. DETERMINING CONTENT OF A PRINTOUT FOR GLP PROCEDURE P2 GLP is group of the parameters which enables declaring variables that are present on an adjustment printout and printout activated after a measurement. P2 GLP P2.1 P2.2 P2.3 P2.4 P2.5 P2.6 P2.7 P2.8 • USr PrJ Ptin PdAt PUSr PPrJ PId PFr | | | | | | | | YES YES YES YES YES YES P2.1 USr The parameter enables determining name of a operator who works with the balance. User name contains maximally 8 alphanumeric characters.
Accessible characters and their equivalents displayed by a balance: An instance of a user name inserted to a balance using capital letters takes below form: OP1_WILK (operator 1 WILK) An instance of a user name inserted to a balance using small letters takes below form: op1_wilk (operator 1 wilk) • P2.2 PrJ The parameter enables determining name of a project (e.g. related to a specific type of weighing process).
An instance of project name inserted to a balance using small letters takes below form: prradwag • P2.3 Ptin Option determining presence of measurement time on a printout (data on measurement time is acquired from a connected printer). • P2.4 PdAt Option determining presence of measurement date on a printout (measurement date is acquired from a connected printer). • P2.5 PUSr Option determining presence of user name on a printout. • P2.
14. BALANCE WORKING MODES - Weighing mode - bASic Parts counting - PcS Checkweighing - HiLo Percent setup with reference mass – PrcA or PrcB Density determination of solids - d_Co; Density determination of liquids - d_Li.
Pressing MODE key causes displaying the first of available modes. Each next pressing of or key causes displaying name of next available mode. The way of setting a working mode is described in further section of this manual. 14.1. Parts counting of the same unit mass Balance in its standard version is equipped with an option for counting parts and small objects of the same unit mass. Parts counting mode does not cooperate with other working modes of balance APPC series.
• in order to select one of the options, e.g.
• If option FrEE (sample quantity is declared by a user) is selected, insert sample quantity used for determining mass of single part, Press Use key to select a digit to be altered, key to select digit value, • • Accept inserted value by pressing PRINT key, Display indicates a command LoAd – load the weighing pan with as many details as inserted while determining sample quantity, • Press PRINT key – balance indicates sample’s quantity (pcs pictogram is visible on the display), • add remaining par
Return to weighing CAUTION: While determining sample quantity, and before accepting it, it is necessary is displayed. Only if to wait until stable measurement pictogram visible on balance’s display, it is allowed to accept the declared quantity of parts by pressing PRINT key. Otherwise balance will not accept the measurement.
14.2.
SET LOW LIMIT (LOW THRESHOLD) - digit selection; - set digit value; - accept inserted digit value SET HIGH LIMIT (HIGH THRESHOLD) - digit selection; - set digit value; - 55 - - accept inserted digit value
CAUTION: If by mistake the value of low threshold is set higher than high threshold, balance will indicate error message and return to weighing.
14.3. Percent setup with reference mass Balance software enables controlling deviation (in %) of weighed loads with determined reference mass. Mass of a reference load can be determined by its weighing (PercA) or entered to balance memory by a user (PercB). 14.3.1.
From now on the display will not indicate mass of weighed load but deviation of mass placed on the pan in relation to the reference (in %). 14.3.2.
• Determine the value of reference mass, insert it to balance memory and accept it by pressing PRINT key – display will indicate: 0,00 % From now on the display will not indicate mass of weighed load but deviation of mass placed on the pan in relation to the reference (in %).
14.4. Density determination of solids and liquids The software enables determining density of solids (with determined dimensions) and liquids. Before carrying out density determining process, it is necessary to place a balance on a dedicated rack which provides access to assembling a hook in the bottom section of the balance (the rack is additional equipment of a balance, and it is manufactured on individual client’s order).
14.4.1. Density determination of solids Procedure: • • • • Press Mode key, then use navigating arrows to select submenu F6 d_Co, Select one of the 3 available liquid types H20 (distilled water), C2H5OH 0 (spirit 100% ± 0.1% in temp.
• After selecting liquid type the balance displays command Load A and zero indication (if indication on balance’s display is other than zero, press ESC/ZERO/TARE key), • Then load the sample on the top weighing pan (as indicated on below figure), and weigh the sample in the air: - 62 -
• • When the balance displays stability marker press Print/Enter key, When the balance displays command Load L – load the sample on the bottom weighing pan of the density kit (in the liquid).
• 3 The balance displays density measurement result expressed in [g/cm ]. CAUTION: Pay special attention to presence of air bubbles during measuring process. If air bubbles are on sample’s surface during measuring process, they may be a source of unreliable measurements.
In order to weigh a sample again, press Units key: 14.4.2. Density determination of liquids Basic element for determining density of liquids is a glass sinker with precisely determined volume. Before starting liquid density determination, insert the value of sinker’s volume to balance memory. In order to measure the density of liquid, first determine mass of the sinker in the air. Then measure mass of the same sinker in tested liquid.
• After inserting sinker’s volume and storing the data in balance’s memory (using Print/Enter key), balance’s display indicates a command Load A and zero (if the indication is other than zero , press ESC/ZERO/TARE key), • Hang the sinker on the density kit and in the air to determine its mass in the air, - 66 -
• • • When the balance indicates stable measurement pictogram press Print/Enter key, As the display indicates a command Load L – place the beaker with tested liquid on the rack’s basis under the balance and hang the sinker on the hook. The sinker has to be fully immersed in the tested liquid, Then determine the density measurement result by pressing Print/Enter 3 key; the balance indicates the density of tested liquid in [g/cm ].
To repeat inserting sinker’s volume procedure press F key: To return to weighing mode press Mode key, and then Zero/Tare key: In order to once again weigh the sinker in the air press Units key: - 68 -
14.5. Totalizing Totalizing mode is basically adding the following components in order to prepare a mixture. After adding following component of the mixture, the balance totalizes mass of the mixture and exposes its total mass on the display. During totalizing process mass of each added component is sent through RS232 interface to a connected computer of a printer (e.g. Kafka printer). CAUTION Totalizing mode functions only in grams.
- 70 -
Return to weighing mode - 71 -
14.6. Working mode FILL Working mode FiLL enables two settings: FiLL = NO – mode disabled, i.e. not available in the working modes. FiLL = YES – mode enabled, i.e. available in the working modes. If FiLL is enabled, then the balance operates with settings, that are stored in balance’s settings as default and ensuring quicker operation. The settings automatically decrease Autozero range to 1 reading interval.
15. RS 232 FUNCTIONS P4 Print P4.1 P4.2 P4.3 P4.4 P4.5 P4.6 bAud CntA CntB rEPl PStb Lo | | | | | | 4800 YES YES YES YES 0.005 P4.1 bAud - baud rate setting (speed of transmission) P4.2 CntA - continuous data transmission in basic measuring unit P4.3 CntB - continuous data transmission in current measuring unit P4.4 rEPl - setting operation mode for RS 232 output (manual / automatic) P4.5 PStb - stable / unstable measurement for RS 232 transmission P4.
Select required baud rate value - 2400 bit/s - 4800 bit/s - 9600 bit/s - 19200 bit/s. Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing). 15.2. Continuous transmission 15.2.1. Continuous transmission in basic measuring unit no – continuous transmission disabled YES – continuous transmission enabled. Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing).
15.2.2. Continuous transmission – in current measuring unit no – continuous transmission disabled YES – continuous transmission enabled. Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing).
15.3. no YES Setting operation mode for RS 232 interface - manual operation (enabled on pressing PRINT key) - automatic operation (enabled on stabilization of weighing result). Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing).
15.4. Determination of data type sent through RS 232 output no YES - sending stable or instantaneous (unstable) measuring result - sending stable weighing result. Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing).
15.5. Minimal mass required for activation of automatic printout Balance software is equipped with a function for setting automatic operation. While in automatic operation, data will not be transmitted through RS 232 to a connected computer or printer until mass reading is above Lo value. • Press • Press key to select a digit which should be changed; key to change value of selected digit. Go back to weighing mode with procedure of saving changes (see point - 8. 2. – return to weighing).
16. COOPERATION WITH A PRINTER OR A COMPUTER CAUTION A peripheral device that is connected to RS 3232 output of a balance, has to be powered from the common low voltage power network equipped with common anti-shock protection in a way to preclude possible occurrence of different potentials in zero cables of the peripheral device and the balance. Transmission parameters programmed in a balance: Baud rate Data bits Stop bit Parity control - 2400 – 19200 bit / s -8 -1 - none.
16.1.
16.2. Printing date with specified date and time Every weighing result can be printed with specification of measurement date and time. The presence of date and time on a printout is conditioned by connection of printers type Kafka 1/Z or Kafka SQ S to a balance. When connected, go to balance menu and set parameters from group P2 GLP: - YES - YES PdAt Ptin 16.3.
16.4.1. Format of data sent on pressing of PRINT key Depending on setting of P4.5 PStb parameter only stable measurement or instantaneous mass indication will be sent. 13 14 - 16 17 18 unit CR LF 4 - 12 mass sign Printout format 2 3 space stability marker 1 space CAUTION: For verified balances printout of instantaneous measurement data is blocked.
17 18 - 20 21 22 CR LF space sign mass unit 8 – 16 unit stability marker Command stability marker 7 space 6 mass 5 sign 4 space command 1–3 – 1 ÷ 3 characters – [space] if stable – [?] if unstable – [^] if there is an error of range exceeding on + – [v] if there is an error of range exceeding on – [space] for positive values or [-] for negative values – 9 marks alignment to the right – 3 marks alignment to the left 17.
18.
19. ERROR MESSAGES Er1 Hi – mass out of range on balance start and adjustment Er2 nuLL – value from AD converter ≤ 0 (A/D ≤ 0) Er3 FuL1 – values from A/D converter ≥ maximal converter range (A/D ≥ 0) Er4 FuL2 – maximal capacity of the balance exceeded (over load) Er5 rout – mass value out of range (during adjustment, determining reference for sample quantity and percent setup, etc.
20. ADDITIONAL EQUIPMENT 20.1. Anti-Vibration table It is a very stable basis which eliminates all kinds of vibration and ground shakes. The inner part of the table holds a marble plate, which is a foundation for balance positioning. 20.2. Additional display Features: Length of conductor between additional display and balance – 1,5m, plastic casing, possibility of tilting the additional display head. 20.3. Computer software PW-WIN 2004 – gathering data. Data is gathered in tabular or chart format.
MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG WAGI ELEKTRONICZNE 26-600 Radom 28 Bracka POLAND phone +48 48 38 48 800 fax. + 48 48 385 00 10 e-mail: export@radwag.com www.radwag.