Use Instructions

Menu (Overview)
51 / 77
7 Menu (Overview)
Mode and user-dependent device settings can be changed in the menu.
User Menu Operating Mode Options Default
Starting pressure
(Operat-
ing mode dependent)
Sets the initial nominal
pressure value for each
mode
Standard 1 - 15 mmHg 15 mmHg
High Flow/Bariatric 1 - 15 mmHg 15 mmHg
Advanced Flow 1 - 15 mmHg 15 mmHg
Pediatric 1 - 12 mmHg 8 mmHg
Vessel Harvest 1 - 15 mmHg 10 mmHg
TAMIS 1 - 15 mmHg 12 mmHg
Gas Flow Rates
(Operating
mode dependent)
Sets the flow rates for each
mode
Flow rate 1 will be loaded
per default in operating
screen, holding the flow
button will allow scrolling
through the rates
Standard
Rate 1: 1 - 5 l/min
Rate 2: Rate 1+1 - 39 l/min
Rate 3: Rate 2+1 - 40 l/min
Rate 1: 3 l/min
Rate 2: 20 l/min
Rate 3: 40 l/min
High Flow/Bariatric
Rate 1: Rate 1 - 5 l/min
Rate 2: Rate 1+1 - 44 l/min
Rate 3: Rate 2+1 - 45 l/min
Rate 1: 5 l/min
Rate 2: 25 l/min
Rate 3: 45 l/min
Advanced Flow
Rate 1: 1 - 5 l/min
Rate 2: Rate 1+1 - 49 l/min
Rate 3: Rate 2+1 - 50 l/min
Rate 1: 5 l/min
Rate 2: 25 l/min
Rate 3: 50 l/min
Pediatric
Rate 1 0.1 - 2 l/min
Rate 2: Rate 1+1 - 19 l/min
Rate 3: Rate 2+1 - 20 l/min
Rate 1: 0.1 l/min
Rate 2: 2 l/min
Rate 3: 20 l/min
Vessel Harvest
Rate 1: 1 - 5 l/min
Rate 2: Rate 1+1 - 9 l/min
Rate 3: Rate 2+1 - 10 l/min
Rate 1: 1 l/min
Rate 2: 4 l/min
Rate 3: 10 l/min
TAMIS
Rate 1: 0.5 - 5 l/min
Rate 2: Rate 1+1 - 39 l/min
Rate 3: Rate 2+1 - 40 l/min
Rate 1: 3 l/min
Rate 2: 10 l/min
Rate 3: 40 l/min
Venting valve
(Operating
mode dependent)
Standard
- Always ON
- In Veress mode OFF
In Veress mode OFF
High Flow/Bariatric
- Always ON
- In Veress mode OFF
In Veress mode OFF
Advanced Flow
- Always ON
- In Veress mode OFF
In Veress mode OFF
Pediatric
- Always ON
- In Veress mode OFF
Always ON
Vessel Harvest
- Always ON
- Always OFF
Always OFF
TAMIS
- Always ON
- Always OFF
Always ON
Smoke Evacuation Level
(Operating mode depend-
ent)
Sets the default starting
suction rate for each mode
Standard Evacuation level 1-4 Level 1
High Flow/Bariatric Evacuation level 1-4 Level 1
Advanced Flow Evacuation level 1-4 Level 1
Pediatric Evacuation level 1-2 Level 1
Vessel Harvest Not available
TAMIS Evacuation level 1-4 Level 1
en
DRAFT