4-8 LIFEPAK 20e Defibrillator/Monitor Operating Instructions
No Shock Advised
Note: If you do not press the
button within 60 seconds, the AED
disarms the shock button, and the
DISARMING message appears.
When the button is pressed you will
see the message
ENERGY DELIVERED
indicating energy transfer was
completed.
When energy transfer is complete the
shock counter increases by 1. This will
continue to increase incrementally with
every energy transfer.
After a shock is delivered you will see
and hear
START CPR. A countdown
timer (min:sec format) continues for the
duration specified in the
CPR TIME 1
setup option.
Note: If the CPR metronome is set
to be on during CPR times, you will
hear audible “tocks” and ventilation
prompts or tones.
When the CPR countdown time ends,
you will see and hear
PUSH ANALYZE.
This message stays on the screen and
the voice prompt will repeat every 20
seconds until you press the
ANALYZE
button.
If the AED detects a nonshockable
rhythm, you will see and hear
NO
SHOCK ADVISED
. The AED will not
charge, and a shock can not be
delivered.