DreamStation Go Heated Humidifier Manual

13
6. Cleaning & Disinfection for Hospitals/Institutions
If you are using the water tank base, water tank lid, and lid components on multiple patients, complete the following
steps to clean and disinfect between each new user.
Caution
Only the hospital and institution cleaning and disinfection procedures listed in this manual are
recommended by Respironics. Use of other cleaning and disinfecting processes, not specied by Respironics,
may aect the performance of the product.
Cleaning the water tank base
1. Remove the water tank from the device and then remove the tank lid.
2. Wash the tank base in a solution of warm water and a mild liquid dishwashing detergent (1 teaspoon of liquid
dishwashing detergent per gallon of warm water). Rinse with clean water. Wipe the top and bottom of the tank
base. Allow the tank base to air dry.
Caution
Use a mild liquid dishwashing detergent only.
3. Inspect the tank base for damage and ensure that the CPAP seal is in place. If it shows signs of wear or damage,
contact your supplier for a replacement.
Cleaning the water tank lid and components
Cautions
Use a mild liquid dishwashing detergent only (1 teaspoon of liquid dishwashing detergent per gallon of warm
water).
Hand wash only.
1. Remove the manifold (1) by pulling up on both
manifold tabs.
2. Remove the bae (2) by pressing in on the side
tabs and pulling up on the bae.