User's Manual

5 Patient Alarms and INOPs Technical Alarm Messages (INOPs)
74
!!/!!! REPLACE BATTERY T
Severe yellow/red INOP tone.
During this INOP, alarms cannot be
paused or switched off.
Telemetry The battery in the telemetry device is almost empty and must
be replaced.
RESP EQUIP MALF
Numeric is replaced by -?-
INOP tone.
RESP Contact your service personnel. The RESP hardware is faulty.
RESP ERRATIC
Numeric is replaced by -?-
RESP The monitor has detected too many artifacts in the measured
Resp signal. Check that the RA and LL electrodes are correctly
attached and have not dried out.
RESP LEADS OFF
Numeric is replaced by -?-
INOP tone.
RESP Not all the required leads for Resp monitoring are attached.
Make sure that the RA and LL leads are attached.
RL LEAD OFF
Numeric is replaced by -?- for 10
seconds; INOP tone.
ECG The RL electrode has become detached from the patient or the
lead set has been changed. Reattach the electrode or select
New
Lead Setup
in the Setup ECG menu to confirm the
new lead set.
Settings Malfunc.
INOP tone.
Monitor The monitor cannot use the predefined settings for
monitoring. Contact your service personnel.
<SO
2
Label> CAL FAILED
Numeric is replaced by -?-
INOP tone.
SO
2
The calibration failed. Check the catheter-to-Optical-Module
connection. Manually restart the calibration. Try another
catheter and Optical Module. If the catheter is already inserted,
perform an in-vivo calibration.
<SO
2
Label> CAL MODE
Numeric is replaced by -?-
INOP tone.
SO
2
Pre-insertion calibration is complete, but the catheter tip is still
inside the optical reference. The catheter is now ready for
insertion.
<SO
2
Label>CAL REQUIRED
Numeric is replaced by -?-
INOP tone.
SO
2
There is no valid calibration data in the Optical Module.
Perform either a pre-insertion or an in-vivo calibration.
<SO
2
Label> CANNOT MEAS
Numeric is replaced by -?-
INOP tone.
SO
2
The signal is out of the normal range, and no oxygen saturation
can be derived. Perform an in-vivo calibration. If the INOP
persists, try another Optical Module and catheter.
<SO
2
Label>CONFIG ERROR
Numeric is replaced by -?-
INOP tone.
SO
2
The Optical Module has been configured to SaO
2
Mode. Use
Change to Venous in the setup menu to reconfigure to venous
saturation mode.
<SO
2
Label> CONN OPTMOD
Numeric is replaced by -?-
INOP tone.
SO
2
The Optical Module was disconnected during data storage.
Reconnect the Optical Module for at least 20 seconds.
<SO
2
Label> EQUIP MALF
Numeric displays -?-.
INOP tone
SO
2
The SO
2
/SvO
2
Module or Optical Module is faulty. Unplug
and replug the Optical Module and SO
2
/SvO
2
module.
Exchange the modules. If the INOP persists, contact your
service personnel.
SO
2
INCOMPATIBLE
INOP tone
SO
2
The SO
2
Module or Optical Module is not supported. Contact
your service personnel.
<SO
2
Label> IN-VIVO CAL SO
2
The in-vivo calibration is not yet complete. Lab values must be
stored to the Optical Module to complete the calibration.
Either continue with the next steps of the current calibration or
recall the previous calibration.
INOP Message, Indication Source What to do