5 Patient Alarms and INOPs Technical Alarm Messages (INOPs)
64
CHARGER MALFUNC
INOP tone, battery LED may flash
Batteries There is a problem with the battery charger in the monitor.
Connect the monitor to mains power and contact your service
personnel.
Check Alarm Lamps
INOP tone.
Monitor Perform a visual check of the alarm lamp to establish whether
there is a problem. Contact your service personnel to check the
internal connections to the alarm lamps.
CHECK BATT TEMP
INOP tone
Battery The temperature of one or both batteries is too high. Check
that ventilation openings (if applicable) are not blocked and
monitor is not exposed to heat.
Check DrugSettings
INOP tone
Monitor There was a problem loading the drug settings. Check that the
settings are complete and correct.
Check ECG Settings
INOP tone
Telemetry Synchronization of ECG settings between the monitor and
Information Center has failed. Check that the ECG settings in
use are appropriate.
!!CHECK ECG SOURCE
INOP tone
Monitor The telemetry device and the monitor both have valid ECG
signals. Unpair the telemetry device and the monitor if they are
no longer used for the same patient.
Check Flex Texts
INOP tone
Monitor Check the names of the monitor menus, for example the labels
for screens, profiles, event or trend group names, before you
resume monitoring. If they are unexpected, there may be a
problem with the monitor software. Contact your service
personnel.
CheckInternVoltage
at Information Center
Monitor Potential problem with alarm lamps, display or interfaces
detected. Contact your service personnel. This INOP will
appear on the monitor as
Check Monitor Func.
Check Keyboard
INOP tone
Monitor Perform a visual and functional check of the keyboard. Contact
your service personnel.
Check Main Board 2
INOP tone.
Monitor There is a problem with the second main board in the monitor.
Contact your service personnel.
Check Monitor Func
INOP tone.
Monitor Potential problem with alarm lamps, display or interfaces
detected. Contact your service personnel. This INOP may
appear on the Information Center as
CheckInternVoltage.
Check Monitor Temp
INOP tone
Monitor The temperature inside the monitor is too high. Check that the
monitor ventilation is not obstructed. If the situation
continues, contact your service personnel.
Check Mouse Device
INOP tone.
Monitor Perform a visual and functional check of the mouse input
device. Contact your service personnel.
Check MSL Voltage
INOP tone
Monitor/
Multi-
Measuremt
Module
There is a problem with the voltage of the Measurement Link
(MSL). Contact your service personnel.
Check Network Conf
INOP tone
Monitor The monitor is receiving network topology information from
more than one source, e.g. the Database Server and an
Application Server. Contact your service personnel.
Check Nurse Relay
INOP tone
Monitor There is a problem with the connection to the nurse relay.
Contact your service personnel.
!!Check Pairing
INOP tone
Monitor There is a problem with device pairing. Check that the monitor
and telemetry device are correctly paired.
INOP Message, Indication Source What to do