User's Manual

Installation Safety Information 24 Specifications
245
Installation Safety Information
WARNING If multiple instruments are connected to a patient, the sum of the leakage currents may exceed the
limits given in IEC/EN60601-1, IEC60601-1-1, UL60601-1. Consult your service personnel.
Connectors
The actual placement of boards and configuration of connections for your monitor depends on how
your hardware has been configured. See the symbols table on page 244 to see which symbols are used
to mark the connections.
WARNING Connect only medical devices to the ECG output connector socket.
Connecting the ECG sync out to external equipment should only be done by a qualified user. Do
not touch the patient when you have contact to the ECG output connector socket.
Silence Alarms Alarms Alarms Off
ECG connector
Error LED Identifies year
and month of
manufacture
LAN connection
indicator for
connection to a wired
network
IntelliVue
Instrument
Telemetry wireless
network
Always use separate
collection for waste
electrical and
electronic equipment (WEEE)
Monitor supports 12-
lead ECG
built-in short
range radio
interface
built-in wireless
network
Symbols
2002-
06
Earthing The monitor must be earthed during operation. The earthing is for functional purposes
and does not provide protection against electric shock. The protection against electric
shock in this device is provided by double and/or reinforced insulation. If a three-wire
receptacle is not available, consult the hospital electrician. Never use a three-wire to two-
wire adapter.
Combining equipment Combinations of medical equipment with non-medical equipment must comply with
IEC 60601-1-1. Never use a multiple portable socket-outlet or extension cord when
combining equipment unless the socket outlet is supplied specifically for use with that
equipment.
Network Cables All network cables must be unshielded.