User's Manual

Table Of Contents
3 Patient Alarms and INOPs
78
Pressure INOPs
<SpO Label> NON-
PULSAT.
Numeric is replaced by -?-
INOP tone
Check the perfusion at measurement site. If necessary, stimulate circulation or change measurement
site. If the INOP is due to NBP measurement on the same limb, wait until the NBP measurement is
finished.
<SpO Label> POOR
SIGNAL
Numeric is replaced by -?-
(questionable numeric)
The signal condition of the SpO
2
measurement is poor and measurement accuracy may be
compromised.
<SpO Label> PULSE?
Numeric is replaced by -?-
INOP tone
The detectable pulsations of the SpO
2
signal are outside the specified pulse rate range.
<SpO Label> SEARCHING
Numeric unavailable
SpO
2
is analyzing the patient signal to derive Pulse, SpO
2
and Perf values. Please wait until the search
analysis is complete.
<SpO Label> SENSOR
MALF
Numeric is replaced by -?-
INOP tone
The SpO
2
sensor or adapter cable is faulty. Try another adapter cable and sensor. If the INOP persists,
contact your service personnel.
<SpO Label> SENSOR
OFF
Numeric is replaced by -?-
INOP tone
The SpO
2
sensor is not properly applied to the patient. Apply the sensor following the instructions
supplied by the manufacturer.
<SpO Label>
UNKN.SENSOR
Numeric is replaced by -?-
The connected sensor or adapter cable is not supported by the SpO
2
measurement. Use only specified
sensors and cables.
<SpO Label>
UNPLUGGED
Numeric is replaced by -?-
INOP tone
An SpO
2
measurement label has been deactivated, either by unplugging a module, or by deactivating
the label in the Measurement Selection window.
The measurement automatically disappears from the display.
To switch the measurement on again, either replug the module or reactivate the measurement label in
the Measurement Selection window.
<SpO Label> UPGRADE
Numeric is replaced by -?-
Numeric is unavailable
The SpO
2
measurement is currently in upgrade mode. Monitoring is not possible in this mode.
INOP Message, Indication What to do
INOP Message, Indication What to do
ABP INOPs See <Press Label> INOPs
Ao INOPs See <Press Label> INOPs
ART INOPs See <Press Label> INOPs
BAP INOPs See <Press Label> INOPs
CPP CHK SOURCES
Numeric is replaced by -?-
INOP tone
Not all measurements or values required to perform the calculation are available. Check the
measurement sources.
CPP CHK UNITS
Numeric is replaced by -?-
The monitor has detected a conflict in the units used for this calculation. Check the unit settings.
CVP INOPs See <Press Label> INOPs