4 Managing Patients
107
Note that a Screen with an embedded Other Bed screen element cannot be displayed on the D80
Intelligent Display.
Changing the Screen may automatically change the bed shown in the other bed window. If you switch
to a different Screen, you should check that the correct other bed is displayed.
Other Bed Pop-Up Keys
Select the Other Patients SmartKey or the Other Bed window or embedded screen element to access
the associated pop-up keys:
Next Wave lets you view waveforms not currently shown in the other bed window.
More Vitals lets you view more numerics not currently shown in the other bed window.
Next Bed lets you view waveforms and numerics from the next available bed in the Care Group.
My Care Group lets you call up the Care Group window to select another bed.
Alarming Beds lets you view a list of all beds in your care group with an unacknowledged alarm.
The beds are listed in order of alarm severity.
Silence Bed lets you silence active alarms at the other bed. (Whether you see this key depends on
the revision and configuration of the Information Center your monitors are connected to.)
WARNING
The Silence Bed pop-up key in the Other Bed window silences alarms at a remote bed. Be aware that
accidental use of this key could silence alarms for the wrong patient.
To silence own bed alarms use the Silence permanent key on screen.
Visual Alarm Status Information in the Other Bed Window
If individual alarms are switched off at the other bed, this is indicated by the crossed alarm symbol
beside the measurement numeric.
If alarms are switched off at the other bed, the message
Alarms Off is shown in the other bed
window.
In the embedded Other Bed window, the crossed speaker symbol in the upper right hand indicates
that the volume of the audible alarm status change notification for the other beds in the care group
is set to zero at the overview monitor.