- Introduction and Basic Operation
- Introducing the IntelliVue Cableless Measurements
- Operating and Navigating
- Switching the Devices On and Off
- Screen Layout
- Standard Layout
- 1 Connection status indicator
- 2 Indicator that alarming capability has been transferred to the host (to the monitor or, for the telemetry device, to the Information Center). No patient alarms will be announced on the Cableless Measurement Device.
- 3 Battery indicator
- 4 Measurement values
- 5 Measurement-related symbols (see the measurement chapters for details)
- 6 Patient identification (from a monitor or telemetry device)
- 1 Connection status indicator
- 2 Battery indicator
- 3 Measurement values
- 4 Measurement-related symbols (see the measurement chapters for details) The Alarms Off symbols indicate that no physiological alarms are available from the Cableless Measurement Devices when not assigned to a host.
- 5 Patient identification (from a monitor or telemetry device)
- INOP Layout
- Low-Activity Screen
- Standard Layout
- When a Cableless Measurement Device Cannot be Activated
- Using the Hard Keys
- Using the SmartKeys
- Using the Main Setup Menu
- Operating Modes
- Using the Patient Menu
- Using Profiles
- Setting the Date and Time
- Charging IntelliVue Cableless Measurement Devices
- Monitoring with the IntelliVue Cableless Measurements
- Monitoring SpO2
- NOTE
- SpO2 Sensors
- Connecting SpO2 Sensors
- 1 Connect the sensor to the single patient Mobile CL SpO2 Cradle (if not already connected).
- 2 Insert the SpO2 Pod into the Mobile CL SpO2 Cradle. The correct orientation is indicated by a matching blue dot inside the cradle.
- CAUTION
- Removing the Pod from the Cradle
- Applying the Sensor
- 1 Choose a finger of the patient that matches the sensor dimension in a way that the sensor optical components are properly aligned and the sensor is neither too loose nor applies too much pressure to the finger. For small pediatric patients consider...
- 2 Remove colored nail polish from the application site.
- 3 Apply the sensor to the patient. The application site should match the sensor size so that the sensor can neither fall off, nor apply excessive pressure. See the sections below for details on applying the different sensors.
- 4 Check that the light emitter and the photodetector are directly opposite each other. All light from the emitter must pass through the patient's tissue.
- WARNING
- Measuring SpO2
- Selecting Measurement Modes
- Understanding SpO2 Numerics and Symbols
- Assessing a Suspicious SpO2 Reading
- Changing the Averaging Time
- Understanding SpO2 Alarms
- Perfusion Numeric (only available on the Patient Monitor)
- Monitoring NBP
- Introducing the Oscillometric NBP Measurement
- Preparing to Measure NBP
- 1 Attach the cradle to the NBP cuff and insert the NBP Pod into the cradle.
- 2 Plug the air tubing into the NBP Pod. Hold the flat part of the connector vertically when plugging it into the inlet, then twist it clockwise to lock into place. Avoid compression or restriction of pressure tubes. Air must pass unrestricted through...
- 3 Make sure that you are using a Philips-approved correct sized cuff and that the bladder inside the cover is not folded or twisted.
- 4 Apply the cuff to the upper arm at the same level as the heart. If it is not at heart level, you must use the measurement correction formula to correct the measurement.
- CAUTION
- Additional Information
- Using a Carrying Pouch
- 1 Apply the cuff as described in step 4 above.
- 2 Secure the pouch on the patient, with the upper ties around the patient's head and arm and the lower ties around the patient's torso.
- 3 Insert the NBP Pod into the cradle.
- 4 Attach the extension hose to the NBP Pod in the cradle.
- 5 Insert the NBP cradle into the pouch
- 6 Connect the extension hose to the cuff.
- Starting and Stopping Measurements
- Enabling Automatic Mode and Setting Repetition Time
- Enabling Sequence Mode and Setting Up The Sequence
- 1 In the NBP menu, select Mode and select Sequence from the pop-up menu.
- 2 Select Setup Sequence to open the Setup Sequence menu.
- 3 Select each phase in turn (A, B, C and D) and select the number of measurements and the time interval between the measurements.
- 4 To have measurements continue after the sequence, set the number of measurements for your last phase to Cont and this phase will run indefinitely.
- CAUTION
- Understanding the NBP Numerics and Symbols
- Correcting the Measurement if Limb is not at Heart Level
- Switching Pulse from NBP On/Off
- Assisting Venous Puncture
- Calibrating NBP
- Technical Alarms (INOPs)
- Care and Cleaning
- Maintenance and Troubleshooting
- WARNING
- Inspecting the Equipment and Accessories
- 1 Examine unit exteriors for cleanliness and general physical condition. Make sure that the housings are not cracked or broken, that everything is present, that there are no spilled liquids and that there are no signs of abuse.
- 2 Inspect all accessories (cables, sensors and so forth). If any show signs of damage, do not use them.
- WARNING
- Maintenance Task and Test Schedule
- Troubleshooting
- Disposing of the IntelliVue Cableless Measurement Devices
- Integrated Battery Handling
- Accessories
- Specifications
- WARNING
- Intended Use
- Manufacturer's Information
- Symbols
- Installation Safety Information
- Safety Specifications
- EMC And Radio Regulatory Compliance
- IntelliVue CL SpO2 Pod Specifications
- IntelliVue CL NBP Pod Specifications
- Alarm Specifications
- Telemetry Device Battery Runtime Specifications
- IntelliVue CL Charging Station Specifications
- IntelliVue CL Transmitter Base Station Specifications
- Safety and Performance Tests
- Electromagnetic Compatibility (EMC) Specifications
- Accessories Compliant with EMC Standards
- Electrosurgery Interference/Defibrillation
10 Specifications
65
IntelliVue CL NBP Pod Specifications
Complies with IEC 60601-2-30:1999/EN 60601-2-30:2000.
Viewing Area 25.6 mm x 19.2 mm
Dot Size 0.2 mm x 0.2 mm
Resolution 128 x 96 pixel
Backlight white LED
Sounds
Hardkey operation tone
INOP tone
Battery
Integrated rechargeable Li-Ion battery with battery
gauge and cycle counter
Runtime (fully charged battery) Continuous measurement:
Typically 24 hours
Minimum 12 hours
Intermittent measurement:
Typically > 32 hours with repetition interval of 2.5
minutes
Charging Time max 2.5 hours
Short Range Radio Specifications
Type built in interface with integrated antenna
Technology IEEE 802.15.4
Frequency band 2.4 GHz ISM (2.400 - 2.483 GHz)
Modulation DSSS (O-QPSK)
Receiver bandwidth 5 MHz
Effective radiated power (ERP) max. 0 dBm (1 mW)
Electronic Article Surveillance (EAS)
EAS tag inside the housing
Real Time Clock Accuracy
less than 5 seconds per day (typically) as long as device
is in power state "Device on" or "Device off".
Automatically synchronized with assigned patient
monitor/telemetry device.
Performance Specifications