User's Manual

Table Of Contents
3 Patient Alarms and INOPs
74
MMS, MMS Extensions and FMS INOPs
CHECK BATT TEMP
INOP tone
The temperature of one or both batteries is too high. Check that ventilation openings (if applicable)
are not blocked and monitor is not exposed to heat.
Chk MSL Connection Check the MSL connection between X2/MP2 and the battery extension for damage or loose
connections. Check also if a second X1 or X2 has been connected accidentally (e.g. in companion
mode).
ExtBat EMPTY The estimated battery-powered operating remaining time is less than 10 minutes. Replace the battery in
the battery extension immediately.
ExtBat INCOMPAT. The battery in the battery extension cannot be used.
ExtBat LOW The estimated battery-powered operating time remaining is less than 20 minutes.
ExtBat MALFUNCTION The monitor cannot determine the status of the battery in the battery extension. If this INOP persists,
replace the faulty battery in the battery extension. If the condition persists and the monitor is not
connected to mains power or a host monitor, this INOP is re-issued two minutes after you
acknowledge it.
ExtBat MISSING There is no battery in the Battery Extension
!!INSERT BATTERY
Severe yellow INOP tone
During this INOP, alarms cannot
be paused or switched off.
X2/MP2 only: There is no battery in the battery compartment. You cannot operate the monitor on
AC mains while the battery compartment is open (not sealed with a battery). Load a battery
immediately.
MSL Power High The power consumption of the devices connected to the Battery Extension is too high. If this
situation continues, the Battery Extension will be switched off. Contact your service personnel.
MSL Power Off The power consumption of the devices connected to the Battery Extension was too high for too long.
The Battery Extension has been switched off. Contact your service personnel.
INOP Message, Indication What to do
INOP Message, Indication What to do
FMS UNPLUGGED
INOP tone
Make sure that the Flexible Module Rack is connected to the monitor. All FMS measurements are off
while the FMS is unplugged.
FMS UNSUPPORTED
INOP tone
The Flexible Module Rack is not supported by your monitor. Contact your service personnel.
MEASSRV UNSUPPORTD
INOP tone
The Multi-Measurement module is not supported by the monitor. Contact your service personnel.
MMS Ext. UNPLUGGED
INOP tone
The MMS extension has been disconnected from the Multi-Measurement Module.
MMS Ext. Unpowered
INOP tone
The MMS extension cannot operate while the Multi-Measurement Module is running on battery
power.
MMS Ext.EQUIP MALF
INOP tone
Loss of communication between the Multi-Measurement Module and the MMS extension. Contact
your service personnel.
MMS UNPLUGGED
INOP tone
Make sure that the Multi-Measurement Module is connected to the monitor. All MMS measurements
are off while the MMS is unplugged.