User's Manual

Table Of Contents
3 Patient Alarms and INOPs
69
** ST-<n> LOW ECG/ST The ST depression in lead <n> is lower than
the limit.
numeric flashes and low alarm limit
is highlighted, yellow alarm lamp,
alarm tone
* SVT
** SVT
ECG/Arrhythmia A run of supraventricular beats greater than the
SVT run limit has been detected and the HR
has exceeded the SVT HR limit.
numeric flashes, yellow alarm lamp,
alarm tone
** Tblood HIGH C.O. The blood temperature value has exceeded the
high alarm limit.
numeric flashes, high alarm limit is
highlighted, yellow alarm lamp,
alarm tone
** Tblood LOW C.O. The blood temperature value has fallen below
the low alarm limit.
numeric flashes, low alarm limit is
highlighted, yellow alarm lamp,
alarm tone
** tcpO HIGH
** tcpCO HIGH
tcGas The tcpO
2
or tcpCO
2
value has exceeded the
high alarm limit.
numeric flashes, high alarm limit is
highlighted, yellow alarm lamp,
alarm tone
** tcpO LOW
** tcpCO LOW
tcGas The tcpO
2
or tcpCO
2
value has fallen below
the low alarm limit.
numeric flashes, low alarm limit is
highlighted, yellow alarm lamp,
alarm tone
* TELE ALARM
** TELE ALARM
*** TELE ALARM
Telemetry This is a generic alarm from the telemetry
system. The specific alarm cause is indicated in
the alarm message in the Telemetry Data
Window.
yellow or red alarm lamp and alarm
tone
** <Temp Label> HIGH TEMP The temperature has exceeded the high alarm
limit.
numeric flashes and high limit is
highlighted, yellow alarm lamp,
alarm tone
** <Temp Label> LOW TEMP The temperature has fallen below the low alarm
limit.
numeric flashes and low limit is
highlighted, yellow alarm lamp,
alarm tone
* VENT BIGEMINY
** VENT BIGEMINY
ECG/Arrhythmia A dominant rhythm of N, V, N, V (N =
supraventricular beat, V = ventricular beat) was
detected.
numeric flashes, yellow alarm lamp,
short yellow audible alarm
*** VENT FIB/TACH ECG A fibrillatory waveform for 4 consecutive
seconds was detected.
numeric flashes, red alarm lamp,
alarm tone
* VENT RHYTHM
** VENT RHYTHM
ECG/Arrhythmia A dominant rhythm of adjacent Vs > vent
rhythm limit and ventricular HR < VTach HR
limit was detected.
numeric flashes, yellow alarm lamp,
short yellow audible alarm
* VENT TRIGEMINY
** VENT TRIGEMINY
ECG/Arrhythmia A dominant rhythm of N, N, V, N, N, V (N =
supraventricular beat, V = ventricular beat) was
detected.
numeric flashes, yellow alarm lamp,
short yellow audible alarm
*** VTACH ECG, Arrhythmia Ventricular tachycardia has been detected
(Consecutive PVCs exceed V-Tach Run limit
and HR exceeds V-Tach HR limit).
numeric flashes, red alarm lamp,
alarm tone
** VueLink ALARM
*** VueLink ALARM
at Information Center
VueLink A yellow (**) or red (***) patient alarm is
present on the VueLink module. Check the
monitor display for more detailed alarm
information.
(on monitor) yellow or red alarm
lamp; the alarm text is defined by
the VueLink device driver
** ΔQTc HIGH ECG/QT ΔQTc value has exceeded the ΔQTc high limit
for more than 5 minutes
numeric flashes, yellow alarm lamp,
alarm tone
Alarm Message From Condition Indication