36 Specifications
430
Electrosurgery Interference/Defibrillation/Electrostatic Discharge
The equipment returns to the previous operating mode within 10 seconds without loss of any stored
data. Measurement accuracy may be temporarily decreased while performing electrosurgery or
defibrillation. This does not affect patient or equipment safety. Do not expose the equipment to x-ray
or strong magnetic fields (MRI).
Fast Transients/Bursts
The equipment will return to the previous operating mode within 10 seconds without loss of any
stored data (M1032 within 30 seconds). If any user interaction is required, the monitor indicates this
with a technical alarm (INOP).
Restart time
After power interruption, an ECG wave will be shown on the display after 30 seconds maximum.
Frequency of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
Equation
d = 1.2P
for respiration and BIS:
d = 3.5P
d = 1.2P
for respiration and BIS:
d = 3.5P
d = 2.3P
for respiration and BIS:
d = 7.0P
Rated max. output
power of transmitter
Separation distance Separation distance Separation distance
0.01 W 0.1 (0.4) m 0.1 (0.4) m 0.2 (0.7) m
0.1 W 0.4 (1.1) m 0.4 (1.1) m 0.7 (2.2) m
1 W 1.3 (3.5) m 1.3 (3.5) m 2.3 (7.0) m
10 W 3.8 (11.1) m 3.8 (11.1) m 7.3 (22.1) m
100 W 12.0 (35.0) m 12.0 (35.0) m 23.0 (70.0) m