36 Specifications
422
Pressure Wave (AWP) Range -20 to 120 cmH
2
O
Accuracy ±2% or ±0.5 cmH
2
O whichever is greater
Scales 10, 20, 40, 60, 80, 100, 120 cmH
2
O
Scale Default 20 cmH
2
O (neonatal)
40 cmH
2
O (pediatric)
40 cmH
2
O (adult)
Respiration Rate
numeric (RRspir)
Range 2 to 120 breaths/min
Resolution 1 breath/min
Inspired Minute Volume
numeric (MVin)
Range 0.01 to 5 l/min (neonatal)
0.06 to 30 l/min (pediatric)
0.4 to 60 l/min (adult)
Resolution 0.01 l/min (neonatal/pediatric)
0.1 l/min (adult)
Expired Minute Volume
numeric (MVexp)
Range 0.01 to 5 l/min (neonatal)
0.06 to 30 l/min (pediatric)
0.4 to 60 l/min (adult)
Resolution 0.01 l/min (neonatal/pediatric)
0.1 l/min (adult)
Inspired Volume
numeric (TVin)
Range 5 to 100 ml (neonatal)
30 to 400 ml (pediatric)
200 to 3000 ml (adult)
Resolution 1 ml (neonatal/pediatric)
10 ml (adult)
Expired Volume
numeric (TVexp)
Range 5 to 100 ml (neonatal)
30 to 400 ml (pediatric)
200 to 3000 ml (adult)
Resolution 1 ml (neonatal/pediatric)
10 ml (adult)
Lung Compliance
numeric
Range 0 to 500 ml/cmH
2
O
Resolution 1 ml/cmH
2
O
Airway Resistance
numeric
Range 0 to 500 cmH
2
O/l/s (neonatal)
0 to 250 cmH
2
O/l/s (pediatric)
0 to 100 cmH
2
O/l/s (adult)
Resolution 1 cmH
2
O/l/s
Peak Inspiratory
Pressure numeric (PIP)
Range 0 to 120 cmH
2
O
Resolution 1 cmH
2
O
Negative Inspiratory
Pressure numeric
Range -20 to 0 cmH
2
O
Resolution 1 cmH
2
O
Mean Airway Pressure Range 0 to 120 cmH
2
O
Resolution 1 cmH
2
O
Plateau Pressure
numeric
Range 0 to 120 cmH
2
O
Resolution 1 cmH
2
O
Positive End-Expiratory
Pressure numeric
(PEEP)
Range 0 to 99 cmH
2
O
Resolution 1 cmH2O
Spirometry Performance Specifications