User's Manual

Table Of Contents
36 Specifications
417
Temp
Complies with EN 12470-4:2000. Specified without transducer.
CO2
The CO
2
measurement in M3014A and M3015A complies with EN ISO 21647:2004 + Cor.1:2005
(except alarm system; alarm system complies with IEC 60601-2-49:2001).
Tachycardia Difference to high limit 0 to
50 bpm
5 bpm steps max. 14 seconds
Clamping at 150 to 300 bpm 5 bpm steps
Bradycardia Difference to low limit 0 to
50 bpm
5 bpm steps max. 14 seconds
Clamping at 25 to 100 bpm 5 bpm steps
Invasive Pressure Alarm
Specifications
Range Adjustment Delay
Temp Performance Specifications
Temp Range -1 to 45°C (30 to 113°F)
Resolution 0.1°C (0.2°F)
Accuracy ±0.1°C (±0.2°F)
Average Time Constant
Less than 10 seconds
Temp Alarm Specifications Range Adjustment
Temp High/Low Alarms
-1 to 45°C (30 to 113°F) -1 to 35°C (30 to 95°F), 0.5°C (1.0°F) steps
35 to 45°C (95 to 113°F), 0.1°C (0.2°F) steps
M3015A Microstream CO
2
Performance Specifications
CO2 Range 0 to 150 mmHg (0 to 20 kPa), or 20% CO
2
, whichever is lower
Accuracy Up to 5 minutes during warmup: ±4 mmHg or 12%, whichever is greater
After 5 minutes warmup:
0 to 40 mmHg (0 to 5.3 kPa):±2.2 mmHg (±0.3 kPa)
Above 40 mmHg (5.3 kPa):±(5.5% + (0.08%/mmHg above 40 mmHg)) of
reading
These specifications are valid for 21% O
2
and N
2
balance, up to 35°C ambient
temperature, up to 60 rpm in adult mode and 100 rpm in neonatal mode.
Outside of these conditions the accuracy reaches at a minimum ±4 mmHg or
±12% of the reading, whichever is greater.
Resolution Numeric: 1.0 mmHg (0.1 kPa)
Wave: 0.1 mmHg (0.01 kPa)
Stability Included in Accuracy specifications
awRR Range 0 to 150 rpm
Accuracy 0 to 40 rpm: ±1 rpm
41 to 70 rpm: ±2 rpm
71 to 100 rpm: ±3 rpm
>100 rpm: ±5% of reading
Warm-up Time 5 minutes for full accuracy specification