User's Manual

Table Of Contents
36 Specifications
405
Measurement Link (MSL) Connectors ODU out (Proprietary)
Voltage 56 V ±10 %
Power 45 W
Power Sync. 5 V CMOS Level; 78.125 kHz (typ.)
LAN signals IEEE 802.3 10-Base-T compliant
Serial signals RS-422 compliant
Video Interface (standard) Connector DVI (digital and analog, single link)
Digital video signals single link TMDS
Analog video signals 0.7 Vpp@75 Ohm
HSYNC/VSYNC signals TTL
DDC signals none
DDC power 5 V +/-5% @0-55 mA
Video Interface (Independent
display interface option)
Connector DVI (digital and analog, single link)
Pixel clock frequency 31,5 MHz - 119 MHz
Digital video signals single link TMDS
Analog video signals 0.7 Vpp@75 Ohm
HSYNC/VSYNC signals TTL
DDC signals none
DDC power 5 V +/-5% @0-55 mA
ECG Sync Output/Analog ECG Output (1/4" stereo phone jack with tip, ring, sleeve)
General Connector 1/4" phone with tip, ring, sleeve
Isolation functional isolation
Analog ECG Output
(ring, tip)
Full scale on display signal gain x measured ECG voltage
Gain error <15 %
Baseline offset <100 mV
Bandwidth 1 to 100 Hz
Output voltage swing ±4 V (min)
Signal delay <22 ms
Signal delay with older versions of
the M3001A MMS
[identifiable with the serial number
prefix DE227 or DE441 and
option string #A01]
<30 ms
Pacemaker Pulse filtered and included in ECG output signal
MX600/700/800 Interface Specifications