User's Manual

Table Of Contents
5 ECG, Arrhythmia, ST and QT Monitoring
125
Capture 12-Lead Pop-up Keys
Here is a summary of all the pop-up keys available during 12-Lead ECG procedure:
EASI ECG Lead Placement
Using a standard 5-electrode set in EASI lead placement you can monitor up to 12 standard ECG leads
simultaneously and continuously at the bedside. EASI provides a monitoring method for trending ST
segment changes that can provide an early indication of ischemia.
WARNING
EASI-derived 12-lead ECGs and their measurements are approximations to conventional 12-lead
ECGs. As the 12-lead ECG derived with EASI is not exactly identical to the 12-lead conventional
ECG obtained from an electrocardiograph, it should not be used for diagnostic interpretations.
Respiratory monitoring is also possible with the EASI placement; respiration is measured between the
I and A electrodes.
Place the electrodes as accurately as possible to obtain the best quality EASI measurements.
When EASI lead placement is selected,
EASI is shown beside the 1mV calibration bar on the ECG
wave on the display, and
EASI is marked on any recorder strips and printouts.
EASI Monitoring During INOP Conditions
If one of the derived EASI leads has an INOP condition (for example,
LEAD OFF), a flat line is
displayed. After 10 seconds, the directly acquired EASI AI, AS, or ES lead (depending on which is
available) is displayed with the corresponding lead label. This causes an arrhythmia relearn.
Pop-up Keys Selecting this pop-up key lets you
Admit Patient enter the age and gender for the patient in the Patient Demographics
window, to allow analysis of the 12-Lead ECG.
Change View switch between the single column and two-column display of the realtime
wave data.
Capture Waves capture the current ECG waves and open the preview window.
Show Waves return to the realtime 12-Lead ECG view.
Show Preview show the preview of the captured 12-Lead wave data.
Setup 12 Lead choose settings for filter, gain, display layout, etc.
Enter OrderNo. enter an order number that is stored with the capture and displayed in the
title bar of the pop-up keys.
Print Report print a 12-Lead report.
Store & Send store the captured 12-Lead ECG in the monitor and send it to the
Information Center.
Show Stored view a previously stored 12-Lead ECG, allowing a visual comparison
between the current and the stored data.