Datasheet

103
11. Technical Specifications
All these specifications apply to the Waveform Generator unless otherwise explanation.
To reach these specifications, the instrument must have been operating continuously for
more than 30 minutes within the specified operating temperature.
All the specifications are guaranteed unless those marked with “typical”.
Frequency Characteristic
Frequency resolution: 1 μHz
Max. Sampling Rate: PeakTech 4120 - 4125: 125 MSa/s
PeakTech 4165: 250 MSa/s
Sine
P 4120
1 μHz—5 MHz
P 4121 / 4124
1 μHz—10 MHz
P 4125
1 μHz—25 MHz
P 4165
1 μHz—60 MHz
Square
P 4120 - 4125
1 μHz—5 MHz
P 4165
1 μHz—30 MHz
Ramp
1 μHz—1MHz
Pulse
P 4125 - 4124
1 μHz—5 MHz
P 4165
1 μHz—15 MHz
White Noise
25 MHz bandwidth (-3 dB) (typical)
Arbitrary
P 4120 / 4121
1 μHz— 5 MHz
P 4124 - 4165
1 μHz—10 MHz
Amplitude Characteristic
Output Amplitude
P 4120 /
P 4121
High Z
1mVpp 25Vpp
50 Ω
1mVpp 12.5Vpp
P 4124
High Z
1 μHz ~ 10 MHz: 1 mVpp - 20Vpp
50 Ω
1 μHz ~ 10 MHz: 0,5 mVpp - 10Vpp
P 4125
High Z
1 μHz ~ 10 MHz: 1 mVpp - 20Vpp
10 MHz ~ 25 MHz: 1 mVpp - 10Vpp
50 Ω
1 μHz ~ 10 MHz: 0,5 mVpp - 10Vpp
10 MHz ~ 25 MHz: 0,5 mVpp - 5Vpp
P 4165
High Z
1 μHz ~ 25 MHz: 1 mVpp - 20Vpp
25 MHz ~ 60 MHz: 1 mVpp - 10Vpp
50 Ω
1 μHz ~ 25 MHz: 0,5 mVpp - 10Vpp
25 MHz ~ 60 MHz: 0,5 mVpp - 5Vpp
Waveforms
Standard Waveforms
Sine, Square, Ramp, Pulse, Noise
Arbitrary Waveforms
Exponential rise, Exponential fall, Sin(x)/x,
Staircase, etc. 45 built-in waveforms, User-Definable
Waveform
Numbers of channel
1 (P 4120 & 4121) or 2 (P4124 4165)