5.17.3 Gate Time
Press Gate Timesoftkey then select ‘Gate Time’ to configure the desired gate time.
The gate time indicates the sampling interval time of tested signal, the longer the gate
time is, the more the sampling data, so the users may get more stable measurement
result and higher measurement resolution. And the shorter the gate time is, the better the
tested signal is tracked, but short gate time will result in low measurement resolution.
Generally, gate time should be longer than the period of tested signal.
5.17.4 Trigger Level
Press Trig levelsoftkey then select ‘Trig levl’ to configure the desired trigger level
value. Set the trigger level to be 0 if using AC coupling and adjust the trigger level if using
DC coupling. The influence for trigger level adjustment is small if the amplitude of tested
signal is higher. But if the amplitude of measured signal is lower or frequency is higher,
you need adjust the trigger level carefully to get a better result.
5.17.5 Sensitivity
Press Sensitivsoftkey then select ‘Sensitive’ to configure the desired sensitivity value.
The bigger the value is, the higher the sensitivity is. The influence for sensitivity
adjustment is small if the amplitude of tested signal is higher. But if the amplitude of
measured signal is lower and noise contained in signal, you need adjust the sensitivity to
get a better result.
Generally speaking, improve sensitivity if test value of frequency is less than standard
frequency of measured signal, or lower sensitivity if test value of frequency is greater
than standard frequency of measured signal.
5.17.6 Coupling Mode
Press Coupledsoftkey to choose between AC or DC. If frequency of measured signal
is higher with DC offset, select AC mode and set the trigger level to be 0. If frequency of
measured signal is lower than 1Hz or amplitude lower than 100mVpp, select DC mode
and adjust trigger level properly to get a better result.
5.17.7 Low-pass Filter
Press Filter On/Offsoftkey to turn on or off the low-pass filter. When frequency of
measured signal is lower with high-frequency noise, frequency testing value will be
greater than standard frequency value of test signal, low-pass filter should be turned on
to filter the high-frequency noise, then get an accurate result. But when frequency of
measured signal is higher with lower amplitude, the low-pass filter will attenuate the