User Manual
The description of the Save Function Menu is shown as the following table:
Function Menu
Setting
Description
Type
Wave
Setting
Image
Record
Choose the saving type (the Record Type see
“27.1. How to Record/Playback Waveforms” on
page 132).
When the type is Wave, the menu shows as following:
Source
CH1
CH2
Math
Choose the waveform to be saved.
Object & Show
Object
1~15
Choose the address which the waveform is saved
to or recall from.
Show
ON
OFF
Recall or close the waveform stored in the current
object address. When the show is ON, if the
current object address has been used, the stored
waveform will be shown, the address number and
relevant information will be displayed at the top
left of the screen; if the address is empty, “None is
saved” will be displayed.
Save
Save the waveform of the source to the selected
address. You can also press Copy button to do it.
Storage format is BIN.
Storage
Internal
External
Save to internal storage or USB storage. When
External is selected, save the waveform according
to the current record length (see “Record Length
Menu” on page 124). The file name is editable.
The waveform file could be open by PeakTech
®
waveform analysis software (on the supplied CD).
When the type is Setting, the menu shows as following:
Setting
Setting1
…..
Setting8
The setting address
Save
Save the current oscilloscope setting to the
internal storage
Load
Recall the setting from the selected address
When the type is Image, the menu shows as following:
Save
Save the current display screen. You can also
press Copy button to do it. The file can be only
stored in a USB storage, so a USB storage must
be connected first. The file is stored in BMP
format and the file name is editable.
-130-