20. How to Implement Sampling Setup
Press the Acquire button and the menu is displayed in the screen, shown as Fig. 38.
Fig. 38 ACQU MODE Menu
The description of the Acqu Mode Menu is shown as follows:
Function Menu Setting Description
Sample
Sample General sampling mode.
Peak detect
Use to capture maximal and minimal samples. Finding highest
and lowest points over adjacent intervals. It is used for the
detection of the jamming burr and the possibility of reducing the
confusion.
Average 4, 16, 64, 128
It is used to reduce the random and don’t-care noises, with the
optional number of averages.
The description of the Record Length Menu is shown as follows:
Function Menu Setting Description
Record Length
1000
Choose the record length
10 K
100 K
1 M
10 M
Change the ACQU Mode settings and observe the consequent variation of the wave form displayed.