Verifying and Setting GPIB Addresses – continued
F-10 1X SC 4812T Lite BTS Optimization/ATP FEB 2005
PRELIMINARY
Agilent 8935 Series E6380
(formerly HP 8935) Test Set
GPIB Address
Refer to Figure F-7 and perform the procedure in Table F-7 to verify
and, if necessary, change the Agilent 8935 GPIB Address.
Preset
Cursor Control
Shift
Inst Config
Local
FW00885
Figure F-7: Agilent 8935 Test Set
This procedure assumes that the Test Equipment is set–up
and ready for testing.
NOTE
Table F-7: Verify and/or Change Agilent 8935 (formerly HP 8935) GPIB Address Procedure
n Step Action
NOTE
The HP I/O Configuration MUST be set to Talk & Listen, or no device on the GPIB will be
accessible.
Consult Test Equipment OEM Documentation for additional information as required.
1
To verify that the GPIB Addresses are set correctly, press Shift and LOCAL on the Agilent 8935.
The current HP–IB Address is displayed at the top of the screen.
NOTE
HP–IB is the same as GPIB.
2 If the current GPIB Address is not set to 18, perform the following actions to change it:
2a Press Shift and Inst Config.
2b Turn the Cursor Control Knob to move the cursor to the HP–IB Adrs Field.
table continued on next page