Verifying and Setting GPIB Addresses – continued
FEB 2005 1X SC 4812T Lite BTS Optimization/ATP F-7
PRELIMINARY
Agilent E4432B Signal
Generator GPIB Address
Refer to Figure F-4 and perform the procedure in Table F-5 to verify
and, if necessary, change the Agilent E4432B GPIB Address.
Numeric Keypad
Softkey Buttons
Softkey Label Display AreaActive Entry Area
Backspace Key
Utility Key
Figure F-4: Setting Agilent E4432B GPIB Address
Table F-5: Verify that and Change Agilent E4432B GPIB Address
n Step Action
1 In the MENUS section of the Instrument Front Panel, press the Utility Key.
The Softkey Labels displayed on the right side of the instrument screen changes.
2 Press the GPIB/RS232 Softkey Button to the right of the instrument screen.
The Softkey Labels changes.
The current instrument GPIB Address will be displayed below the GPIB Address Softkey Label.
3 If the current GPIB Address is not set to 1, perform the following to change it:
3a Press the GPIB Address Softkey Button.
The GPIB Address Label and current GPIB Address changes to boldface.
In the on–screen Active Entry Area, Address: is displayed followed by the current GPIB Address.
3b
On the Front Panel Numeric Keypad, enter the Signal Generator GPIB Address of 1.
The GPIB Address Label changes to Enter.
Characters typed in the Keypad replace the current GPIB Address in the Active Entry Display.
NOTE
To correct an entry, press the Backspace Key at the lower right of the Keypad to delete one character at
a time.
3c Press the Enter Softkey Button to set the new GPIB Address.
The new GPIB Address will be displayed under the GPIB Address Softkey Label.