Test Equipment Preparation – continued
FEB 2005 1X SC 4812T Lite BTS Optimization/ATP F-3
PRELIMINARY
Power Sensor Calibration
Table F-3 describes the E7495A Power Sensor Calibration.
Table F-3: E7495A Power Sensor Calibration Procedure
n Step Action
1 Display the Power Meter Screen.
2 Zero the Power Meter by performing the following actions.
Make sure you are connected as shown in Figure F-1.
2a Press the Zero Softkey.
2b Press the Continue Softkey.
3 Calibrate the Power Meter by performing the following actions.
3a Press Ref CF.
3b Enter the Reference Cal Factor, reading it off the label on the Power Sensor Head.
3c Press Calibrate.
3d Connect the Power Sensor.
Refer to Figure F-2.
3e Press Continue.
3f Press Cal Factor.
3g Enter the Cal Factor from the label on the Power Sensor Head.
Select a Cal Factor that is within the operating frequency of the Base Station.