Test Equipment Preparation
FEB 2005 1X SC 4812T Lite BTS Optimization/ATP F-1
PRELIMINARY
Purpose
This appendix provides information on pre–testing set–up for the
following Test Equipment Items (not required for the Cybertest Test
Set):
S Agilent E7495A Test Equipment Set–up
S Agilent E4406A Transmitter Test Set
S Agilent E4432B Signal Generator
S Advantest R3267 Spectrum Analyzer
S Advantest R3562 Signal Generator
S Agilent 8935 Analyzer (formerly HP 8935)
S HP 8921 with PCS Interface Analyzer
S Advantest R3465 Analyzer
S Motorola CyberTest
S HP 437 Power Meter
S Gigatronics 8541C Power Meter
S GPIB Adapter
Pre–testing set–up information covered includes verification and setting
GPIB Addresses, Inter–unit Cabling, Connectivity Testing, Pre–test
Control Settings, and Equipment Calibration for items that are not
calibrated with the Calibrate Test Equipment Function of the LMF.
The following procedures cover verification and changing GPIB
Addresses for the various items of CDMA Test Equipment supported by
the LMF.
Agilent R7495A Test Equipment
Set–up
This Test Equipment requires a warm-up period of at least 30 minutes
before BTS Testing or Calibration begins.
Using the Agilent E7495A with the LMF
The Agilent E7495A does not require the use of the 19MHz Frequency
Reference; if connected, it will be ignored. The Even Sec SYNC
Connection is required.
The Agilent E7495A Signal Generator is only calibrated down to –80db.
In order to achieve accurate FER Testing, ensure that the RX Set–up
includes at least 40db of Attenuation. This will ensure the Signal
Generator will output sufficient power to operate in the calibrated range.
Set the IP Address as described in Table F-1.
Table F-1: Set IP Address on Agilent E7495A Test Set
n Step Action
1 Use the System Button > Controls >IPAdmin to set an IP Address on the E7495A as
128.0.0.49, and Netmask to 255.255.255.128.
F