Operator’s Manual ZM-520PA/521PA/530PA/531PA 59
To cancel suspending alarms during 2 minute alarm suspension:
1. Press the Function key while the “ALARMS SUSPENDED” message is displayed. The
confirmation screen appears.
2. Press the
key to cancel alarm suspension.
Press the Screen key to not cancel alarm suspension.
Pause Monitoring
When the FUNCTION KEY is set to “SUSPEND ALARM & PAUSE” on the SYSTEM SETUP
screen, you can pause monitoring on the receiving monitor from the transmitter when the patient
cannot be monitored, such as during X-ray examination.
NOTE
To use the Function key for PAUSE, you must set “PROTOCOL on the SYSTEM SETUP
screen of the transmitter to 51 and the receiving monitor must be able to receive protocol
51. To use protocol 51, the ORG-9100A/ORG-9110A multiple patient receiver software
version 03-03 or later is required.
Monitoring on the receiving monitor resumes when one of the following conditions is met.
When OFF is selected for “AUTO RESUME AFTER PAUSE” on the SYSTEM SETUP screen,
monitoring does not automatically resume.
Heart rate is properly monitored on the transmitter for the interval selected for “AUTO
RESUME AFTER PAUSE”.
• SpO
2
is properly monitored on the transmitter for the interval selected for “AUTO RESUME
AFTER PAUSE”.