MD211C2 MD211C3 INSTALLATION & MAINTENANCE GUIDE
Index Warning, Caution ..........................................................................................................................English-1 Intended Use ................................................................................................................................English-2 Registration Information ...............................................................................................................English-3 Recommended use ............................................
TO PREVENT FIRE OR SHOCK HAZARDS, DO NOT EXPOSE THIS UNIT TO RAIN OR MOISTURE. ALSO, DO NOT USE THIS UNIT’S POLARIZED PLUG WITH AN EXTENSION CORD RECEPTACLE OR OTHER OUTLETS UNLESS THE PRONGS CAN BE FULLY INSERTED. REFRAIN FROM OPENING THE CABINET AS THERE ARE HIGH VOLTAGE COMPONENTS INSIDE. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL. CAUTION TO REDUCE THE RISK OF ELECTRIC SHOCK, MAKE SURE POWER CORD IS UNPLUGGED FROM WALL SOCKET.
Intended Use The MD211C2/MD211C3 Color displays are intended to be used for displaying and viewing of digital images for diagnosis by trained physicians. To guarantee the display performance as specified, it must only be used in conjunction with NEC approved display controllers. MD211C2/MD211C3 cannot be used for a life-support system. This device must not be used in digital mammography. This device is designed for exclusive interconnection with IEC 60601-1-1 certified equipment.
English Registration Information Declaration Declaration of the Manufacturer NEC Display Solutions Europe GmbH declares that the product listed is in conformity with the essential requirements and provisions of the Council Directive 93/42/EEC, including the amendments of the Council Directive 2007/47/EC, and conforms to the applicable clauses of the following standards: Means of Conformity Device Classification: Class I, non-measuring function Applicable Rules: Annex IX, Rules 1.4 (Section 1) and 1.
Recommended use Safety Precautions and Maintenance FOR OPTIMUM PERFORMANCE, PLEASE NOTE THE FOLLOWING WHEN SETTING UP AND USING THE LCD MONITOR: WARNING: Do not modify this equipment without authorization of the manufacturer. WARNING: To disconnect this product from the mains supply disconnect the mains plug from the socket outlet. WARNING: To avoid risk of electric shock, this equipment must only be connected to a supply mains with protective earth. WARNING: No modification of this equipment is allowed.
b) This model is equipped with an integrated front sensor and ambient light sensor complying with the IEC 62563-1 standard requirements for measurement devices with a calibration traceable to a primary standards laboratory. Therefore, these sensors can be re-calibrated using external luminance and illuminance meters as a reference, which were calibrated following country specific legal obligations for metrological control.
Contents Your new NEC monitor box* should contain the following: • MD211C2/MD211C3 monitor with tilt/swivel/pivot/height adjust stand • Power Cord • DVI-D cable*1 • DisplayPort cable • USB Cable • Quick Reference Guide • CD-ROM x 2 • Screw (x 4) (to mount the monitor to a flexible arm (page 11)) Screws Power Cord*2 CD-ROM x 2 Quick R eferenc e Guide Quick Reference Guide DVI-D cable*1 USB Cable DisplayPort cable * Remember to save your original box and packing material to transport
To connect the LCD monitor to your system, follow these instructions: NOTE: Make sure to read “Recommended Use” (page 4) before installation. In order to display the maximum resolution, a display controller that can output a resolution of 1200 x 1600 (in portrait mode) or 1600 x 1200 (in landscape mode) for MD211C2 and 1536 x 2048 (in portrait mode) or 2048 x 1536 (in landscape mode) for MD211C3 is needed. 1. Turn off your computer. 2.
4. Connect all cables to the appropriate connectors (Figure C.1). When using the USB cable, connect the B type connector to the USB upstream port on the right back side of the monitor and the A type connector to the downstream port on the computer (Figure C.1a). If using the cord from a USB device, plug into one of the downstream ports of the monitor. NOTE: Incorrect cable connections may result in irregular operation, damage display quality/components of the LCD module and/or shorten the module’s life.
8. Connect one end of the power cord to the AC inlet on the back of the monitor and the other end to the power outlet. NOTE: Please refer to the Caution section of this manual for proper selection of an AC power cord. Figure D.1 9. Turn on the monitor with power switch on the top (Portrait) or left side (Landscape) of the monitor (Figure E.1). Turn on the computer. NOTE: If you have any problems, please refer to the Troubleshooting section of this Manual. Power switch Figure E.
Screen Rotation Before rotating, the screen must be raised to the highest level and tilt to avoid knocking the screen on the desk or pinching your fingers. Disconnect all cables. To raise the screen, place hands on each side of the monitor and lift up to the highest position (Figure RL.1). To rotate the screen, place hands on each side of the monitor and turn clockwise from Landscape to Portrait or counterclockwise from Portrait to Landscape (Figure R.1).
English Flexible Arm Installation This LCD monitor is designed for use with a flexible arm. To prepare the monitor for alternate mounting purposes: • Follow the instructions provided by the manufacturer of the monitor mount. • To meet the safety requirements, the monitor must be mounted to an arm that supports the weight of the monitor. See page 23-24 for details. Remove the monitor stand before mounting. Remove the Monitor Stand for Mounting To prepare the monitor for alternate mounting purposes: 1.
Controls OSD (On-Screen Display) control buttons on the front of the monitor function as follows: To access the OSD menu, press the EXIT button. To change the signal input, press the SELECT button. All buttons are located at the back of the monitor. The Key Guide appears pressing any button when not in the OSD control menu. NOTE: The OSD must be closed in order to change the signal input. 4 5 6 7 E D C B 1 2 A 3.
• Set the OSD language before using OSD functions. • Use the control keys (LEFT/RIGHT or UP/DOWN or EXIT) to access the “LANGUAGE SELECTION” menu. • Press the LEFT/RIGHT or UP/DOWN buttons to select the desired OSD language. • To exit from this OSD menu, press the EXIT button. NOTE: Setting the OSD language is only necessary upon initial setup. The OSD language will stay the same until changed by the user (See page 14). Luminance LUMINANCE Adjusts the overall image and background screen luminance.
EDID SELECTION During communication via EDID* with the signal source (computer), the preferred signal timing of the monitor is provided. AUTO: The preferred timing depends on the physical orientation of the monitor. Landscape orientation: 1600x1200 pixels, portrait orientation: 1200x1600 pixels for MD211C2. Landscape orientation: 2048x1536 pixels, portrait orientation: 1536x2048 pixels for MD211C3. The orientation is automatically detected by a sensor.
This control completely locks out access to all OSD control functions. When attempting to activate OSD controls while in the Lock Out mode, a screen will appear indicating the OSD controls are locked out. There are two types of OSD LOCK OUT: OSD LOCK OUT with no control: To activate the OSD Lock Out function, press SELECT, then “RIGHT” button and hold down simultaneously. To deactivate the OSD Lock Out, press SELECT, then “RIGHT” button and hold down simultaneously while in the OSD menu.
Using the PICTURE MODE function Choose the Picture mode which is the most suitable for the type of displayed content. These are the available modes: MEDICAL, Gamma 2.2 and PROGRAMMABLE. • Each PICTURE MODE includes PRESET, LUMINANCE, WHITE, Color Gamut (RED,GREEN,BLUE) , GAMMA, BLACK, UNIFORMITY, AMB.LIGHT COMP., RESPONSE IMPROVE and (6-AXIS ADJUST) settings. You can change these settings in Tag1 and Tag2 on the Advanced OSD. The type of PRESET PRESET PURPOSE MEDICAL Factory setting.
Allows the monitor to display an internal test pattern similar to AAPM TG18-QC, without using an external signal source. A quick overall visual evaluation as well as an inspection of the gray levels according to DICOM is available with this built-in method. - If the monitor is connected with a video signal source (computer): Make sure that no OSD menu is displayed, then press the RESET button for more than 3 seconds or press the RESET button twice.
Advanced OSD • When both OSD and Key Guide are off, push the “RESET” and “EXIT” buttons two times simultaneously. • You will see the Advanced OSD. This menu is larger than the normal OSD and has a tagged structure from Tag 1 to 9. • Push the “EXIT” button. To make an adjustment, ensure that the tag is highlighted, then press “SELECT”. To move to another tag, press “EXIT”, then press “LEFT” or “RIGHT” to highlight another tag.
Tag2 Adjusts color gamut. When the selected settings are outside of the LCD gamut, the indicator on the OSD will turn magenta. UNIFORMITY This function electronically compensates the slight variations in the white uniformity level that may occur across the display area of the screen. These variations are a characteristic of LCD panel technology. This function evens out the luminance uniformity of the display. Selectable UNIFORMITY levels are OFF, LOW and HIGH.
Tag4 H.POSITION Controls the Horizontal Image Position within the display area of the LCD. Press “LEFT” or “RIGHT” to adjust. V.POSITION Controls the Vertical Image Position within the display area of the LCD. Press “LEFT” or “RIGHT” to adjust. H.RESOLUTION Adjusts the horizontal size by increasing or decreasing the setting. Press “RIGHT” button to expand the width of the image on the screen. Press “LEFT” button to narrow the width of the image on the screen. V.
Selects the video detection method, when more than one video signal source is connected. Press “LEFT” or “RIGHT” to select. FIRST: When current video input signal is not present, then the monitor searches for a video signal from a different video source. If a video signal is supplied by the other video source, the monitor automatically switches to the new video source. The monitor will not look for any other video signals, as long as the current video signal is present.
Tag7 Tag8 GRAYSCALE MODE This function changes the display image to monochrome. There are 4 types of settings. OFF: The Display image is COLOR. MODE1: The signal Input is 8 bit monochrome. The display shows a grayscale image. This mode uses only the Green channel of an R/G/B input signal. MODE2: Input is R/G/B Color, but the monitor changes the color space to YUV. The displayed image will become monochrome. Only Y data passes the R/G/B LUT. 10BIT: The signal Input is 10 bit monochrome.
Monitor Specifications MD211C2 Monitor Notes LCD Module 54.0 cm/21.3 inches 54.0 cm/21.3 inches 1200 x 1600 (Portrait) 1600 x 1200 (Landscape) Active matrix; thin film transistor (TFT) liquid crystal display (LCD); UA-SFT (IPS) 0.270 mm dot pitch; 900 cd/m2 max luminance; 400 cd/cm2 calibrated luminance; 1400:1 contrast ratio (typical). DisplayPort Connector: Digital RGB DisplayPort Complies with Standard V1.
Specifications - MD211C3 Monitor Specifications MD211C3 Monitor Notes LCD Module 54.0 cm/21.3 inches 54.0 cm/21.3 inches 1536 x 2048 (Portrait) 2048 x 1536 (Landscape) Active matrix; thin film transistor (TFT) liquid crystal display (LCD); UA-SFT (IPS) 0.212 mm dot pitch; 800 cd/m2 max luminance; 400 cd/cm2 calibrated luminance; 1400:1 contrast ratio (typical). DisplayPort Connector: Digital RGB DisplayPort Complies with Standard V1.
DisplayPort: DisplayPort is designed to be the future-ready, scalable solution for high performance digital display connectivity. It enables the highest resolutions, the fastest refresh rates and deepest color depths over standard cables. DVI-D: The digital-only subset of DVI ratified by the Digital Display Working Group (DDWG) for digital connections between computers and displays. As a digital-only connector, analog support is not provided off a DVI-D connector.
Troubleshooting No Picture • The signal cable should be completely connected to the display controller/computer. • The display controller should be completely seated in its slot. • The monitor does not support a DisplayPort converter signal. • Monitor power switch and computer power switch should be in the ON position. • Make certain the computer is not in a power-saving mode (touch the keyboard or mouse).
Stand alone calibration can recalibrate the display without using a computer by the integrated front sensor or connecting an external sensor. There are 5 functions in STAND-ALONE CALIBRATION. • Self Calibration: Calibrates the integrated front sensor based on the USB color sensor. Alternatively, manually adjust the integrated front sensor with an external near range color analyzer reference, which is traceable to a primary standards laboratory.
Self Calibration Calibrates the integrated front sensor based on the USB color sensor. Or to manually adjust the integrated front sensor with an external near range color analyzer reference, which is traceable to a primary standards laboratory. All PICTURE MODES are updated when Self Calibration is performed. NOTE: AUTO CALIB. mode requires connecting the USB color sensor to the sensor port. EXPERT CALIB. mode requires an external near range color analyzer. 1.
The white point and luminance can be copied from one display to one or more additional displays. Using this feature reduces the variation between different displays allowing them to match each other more closely. B B A A USB color sensor MD211C2/ MD211C3 Source display MD211C2/ MD211C3 Source display Sensor Port (located at the back of the monitor) Figure C.1 Figure C.2 Display A – SOURCE display of white point and luminance to copy. Display B – MD211C2/MD211C3 which performs a copy. 1.
Gamma Adjust By using the integrated front sensor, Gamma Adjust recalibrates the LCD gamma. Gamma Adjust is recommended if MAX GSDF ERROR exceeds the medical standards. All PICTURE MODES are updated when Gamma Adjust is performed. 1. To show the CALIBRATION menu, select the STAND ALONE CALIBRATION in the advanced menu (page 27). 2. Select GAMMA ADJUST in the MODE selection (Figure 1). 3. Press SELECT to start the calibration. This may take several minutes depending on user setting. Figure 1 4.
DICOM PART14 can be performed by using the integrated front sensor. 1. To show the CALIBRATION menu, select the STAND ALONE CALIBRATION in the advanced menu (page 27). 2. Use “LEFT” or “RIGHT” to select DICOM MEASURE in the MODE selection (Figure D.1). 3. Select the GAMMA that you want to measure and press SELECT. NOTE: This function is effective only if GAMMA (Advanced menu Tag1) is DICOM or PROGRAMMABLE. 4. After the DICOM MEASURED RESULT message appears (Figure D.2).
5. Input the value measured at 1, and press SELECT (Figure AS.3). When not having measured, press SELECT without changing the value. 6. After the CALIBRATION SUCCEEDED message appears (Figure AS.4), press SELECT. You can set DATE and TIME by using control keys, and press SELECT. 7. To end the calibration mode, press EXIT. Figure AS.3 Figure AS.4 Illuminance meter Figure AS.
NEC DISPLAY SOLUTIONS is strongly committed to environmental protection and sees recycling as one of the company’s top priorities in trying to minimize the burden placed on the environment. We are engaged in developing environmentallyfriendly products, and always strive to help define and comply with the latest independent standards from agencies such as ISO (International Organisation for Standardization) and TCO (Swedish Trades Union).