User's Guide

POLI User’s Guide
25
1. Turn on your POLI by pressing and holding the [ /↓] key, and allow the
instrument to boot up fully until the main measurement screen with sensor names
and readings is shown.
Important! Make sure all of the instrument’s sensors have warmed up before performing
the bump test. The instrument will take the time to warm up the sensors prior to enabling
access to bump test menus. You can tell a sensor has warmed up if you see a reading next
to it name on the display. If it has not yet warmed up, you see three dashes (--) next to
it.
2. Install a calibration adapter on the gas inlet at the top of the diffusion version
POLI.
3. Connect the pump version instrument to the T calibration tube and connect it to
the calibration gas.
4. Turn on the standard gas to initiate flow.
5. Press [+/OK] key to start the bump test, or press [ /↓] to quit. While the bump
test is being performed, the readings for each sensor are shown. Once the bump
test completes, pass/fail test results and readings are shown for each sensor.
Important! If one or more sensors fail a bump test, be sure to calibrate those sensors.
6. The bump test is now complete. Press [+/OK] to exit and return to the calibration
menu.
If all of alarms and sensors have passed and no sensor is due for a calibration, the
instrument is now ready for use.
6.2.2 Air (Zero) Calibration
This operation sets the zero point of the sensor calibration curve for clean air. It should be
performed before other calibrations.
Note: If you use a zero air cylinder, you must use the Calibration Adapter (and a T
calibration tube if the instrument has a pump). Using a calibration adapter is not
necessary for calibration in fresh air.
Note: Make sure pressure in the
calibration gas cylinder is
higher than 100 psi when using
a T calibration tube (pumped
models only).