DualScan TM U L T R A Talking Ear and Forehead T H ERM O ME TER plus Pulse Meter
User’s Manual for DualScan ™ Ultra Digital Thermometer - Thank you for purchasing our DualScan™ Ultra thermometer. For optimal use of this product, be sure to read this user manual before any operation. - Remove safety tab from battery. Open battery case and gently pull out the plastic safety tab from underneath the battery. Thermometer will not activate until this has been performed.
WARNING Use of this thermometer is not intended as a substitution for consultation with your physician. Measurement results are for reference only. They cannot be used as a clinical assessment or medical diagnostic basis. Contact your physician if you have or suspect any health concerns. Don't touch or blow infrared sensor. *A soiled Infrared sensor may cause inaccuracy. Clean a soiled infrared sensor with a soft wet cloth in a gentle manner.
RECOMMENDATIONS - Don't use the product for other purposes then those propose herein. - This is not a waterproof device. Clean it only with dampened cloth gently. - It is unsafe and hazardous to leave the product exposed to any chemical solvent, direct sunshine or high temperature. - Don't drop, handle roughly, or cause any vibration or impact to the product. - Don't use a mobile phone nearby when the thermometer is in use.
LCD DISPLAY Indicates AM or PM when time of day is displayed below it Numerical Readings for Temperature / Time / Pulse Memory Number heading (Memory Numbers displayed below) Calendar Year / Number of the previous reading in Memory Calendar Month Calendar Day Indicates Ear Temperature Button was used Indicates Audio Talking Mode in use Indicates Forehead Temperature Button was used Indicates Settings Button in use Human Temperature measurement indicator Temperature Degree Scale (ºF or ºC) Indicates Heart R
PRIMARY FUNCTIONS Calendar Date falls within Gregorian calendar year 2001–2099. Time of Day is shown in 24 hour shifts. The initial date for a powered unit is set as January 1st of the manufacturing year. Room Temperature is displayed with Time of Day (alternating every 5 seconds) when not in use otherwise. Measurement ranges from 41.0°F to 139.8°F (5.0°C to 59.9°C). Body Temperature displayed when Ear or Forehead Buttons are used with measurements ranging from 89.6°F to 109.3°F (32.0°C to 42.9°C ).
Large Heart your heart beat as you leave your thumb on the Heart Rate Button. When Large Heart symbol your Pulse Rate will be displayed in heart beats per minute. • The Heart Rate is usually read within 15 seconds if the Large Heart symbol between your thumb and the sensor is not optimal and will result in an ERR (error) message on the display indicating the Heart Rate was not measured.
HOW TO TAKE TEMPERATURE READINGS To measure Ear Temperature Ear Button cover from the unit, and insert the slimmer Ear Forehead Probe tip just slightly Probe Cover into the ear opening, then press and hold the Ear Button to measure the body temperature. Grip unit comfortably and depress ONLY the Ear Button.
Display of Measurement Results Temperature Range: Results shown from 89.6°F to 109.3°F (32.0°C to 42.9°C). “HI” is displayed for all results above 109.3°F (42.9°C), and “LO” displayed for result below 89.6°F (32.0°C) Fever Diagnosed: a RED LED lights with results from 100.4°F (38.0°C) to 109.3°F (42.9°C) No Fever: a GREEN LED lights with results from 96.0°F (35.6°C) to 100.3°F (37.9°C) Thirty seconds after a measurement the unit automatically returns to Clock Mode. button is pressed.
To retrieve Readings from the Memory Press and release the SET Button to retrieve the most recent reading recorded in the Memory. It will show the Body Temperature, the Type of Reading (ear or forehead), the Fever/No Fever colored light, the Date it was recorded, and the Memory Number (highest number in memory). To retieve further readings from the memory just keep pressing the SET Button.
To set to Sleep Mode The unit enters Sleep Mode one minute after the last button is pressed. Sleep Mode (enabled) Sleep Mode (disabled) To change between Fahrenheit or Celsius Press the Ear Button and the TALK Button together. The unit will change from which ever scale it is in, to the other…Celsius or Fahrenheit. This setting change will be spoken if the Talking Feature has not been disabled.
RECOMMENDATIONS For users not familiar with digital thermometers, it is recommended that you measure the same ear 3 times during initial usage. The maximum reading shall Wait about 30 seconds before measuring the same patient again to factor in changes of skin condition. For users in sound health, both ears will be virtually the same in terms of body temperature. Clean ear before taking measurement. REPLACE BATTERIES Remove the battery compartment door as shown below.
MAINTENANCE, STORAGE AND CALIBRATION Maintenance: 1. Remove all stains from the device with a soft, dry cloth. 2. Clean the tip of the probe as follows: Wipe the surface gently with a swab of cotton or a soft cloth moistened with alcohol until no dirt or wax remains. Allow the alchol to dry before using thermometer again. 3. Do not wash the thermometer with water, abrasive detergent or detergent containing benzene or other harsh solvents. Storage: 1.
UNIT SPECIFICATIONS • Product Name: DualScan Ultra Thermometer • Model: 70120 • Power supply: DC3V (2 x AAA battery) • Temperature Sensing Part: IR sensor • Measuring range: 32.0°C - 42.9°C (89.6°F -109.3°F) • Room Temperature: 5.0°C - 59.9°C 141.0°F -139.8°F) • Measuring accuracy: +1- 0.2°C (or +1- 0.4°F) from 35.5°C - 42.0°C (or 95.9°F -107.6°F), and +/- 0.3°C (+/- 0.5°F), from 32.0°C - 35.4°C (or 89.6°F - 95.8°F) or from 42.1°C - 42.9°C (or 107.7°F - 109.
EXPLANATION OF SYMBOLS Type B Applied part Disposal in accordance with Directive 2002/96/EC (WEEE) The name and the address of the manufacturer The name and the address of the authorised representative in the European Community 0197 Complies with the European Medical Device Directive (93142/EEC) and amended by directive 2007/47/EC requirements. Refer to Instruction Manual.
GUIDANCE AND MANUFACTURER'S DECLARATION Guidance and manufacturer's declaration - electromagnetic emissions below. The customer or the user must ensure that it is used in such an environment. Emissions Test Compliance Electromagnetic environment - guidance RF emissions CISPR 11 Group2 The device must emit electromagnetic e energy in order to perform its intended function.
Guidance and manufacturer's declaration - electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user should assure that it is used in such an environment.
Guidance and manufacturer's declaration - Electromagnetic immunity The device is intended for use in. the electromagnetic environment specified below. The customer or the user should assure that it is used in such an environment. Portable and mobile RF Communications equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance.
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [Vi] V/m.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic and people. For transmitters rated at a maximum output power not listed above , the recommended separation distance d in meters (m) can be estimated using theequation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) accordable to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies.
TROUBLESHOOTING GUIDE Troubles Check lists Battery drained No response/ Battery in wrong polarity? Automatic reset when pulling out insulator Poor battery contact Battery symbol on LCD “Lo” on screen (result below 32.0℃ or 89.6℉) Low battery Countermeasures Change batteries Take out battery, reinsert battery correctly Change new battery Follow user’s manual to Thermometer correctly angled correctly angle for the for the eardrum / forehead? eardrum / forehead.
LIMITED WARRANTY Mobi Technologies, Inc. warrants this product to be free from defects in material and workmanship for a period of ninety (90) days from the date of purchase. Should this product prove to be defective at any time during the warranty period, Mobi Technologies, Inc. will, at its option, either replace or repair it without charge. After the warranty period, a service charge will be applied for replacement of parts or labor for repair.