Instruction Manual
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
906 Titrando
■■■■■■■■
27
6
Screw the 6.1805.200 PTFE M8 tubing of the solvent bottle onto the
dosing tip.
Inserting electrode, adsorber tube and septum stoppers
6.2730.020
6.2730.010
6.0338.100
6.1403.040
6.2104.020
1
2
3
4
5
6
Proceed as follows:
1
Introduce the 6.2730.020 septum stopper (with septum inserted)
into the front opening of the vessel lid.
2
Insert the O-rings of the electrode and of the adsorber tube into the
middle openings of the vessel lid.
3
Screw the two 6.2730.010 screw nipples into the openings with the
O-rings. Do not screw too tightly.
4
Introduce the 6.0338.100 double Pt electrode into the left-hand
opening and then tighten the screw nipple until it seals.
5
Screw the 6.2104.020 electrode cable tightly onto the electrode.