DeltaGT MI 3309 BT Instruction manual Ver. 1.1, Code no.
Distributor: Manufacturer: Metrel d.d. Ljubljanska cesta 77 SI-1354 Horjul E-mail: metrel@metrel.si http://www.metrel.si © 2013 Metrel Mark on your equipment certifies that this equipment meets the requirements of the EU (European Union) regulations concerning safety and electromagnetic compatibility No part of this publication may be reproduced or utilized in any form or by any means without permission in writing from METREL.
MI 3309 BT DeltaGT Table of contents Table of contents 1 General description................................................................................................. 6 1.1 1.2 1.3 1.4 2 Warnings ................................................................................................................ 7 Battery and charging............................................................................................... 8 New battery cells or cells unused for a longer period .................
MI 3309 BT DeltaGT Table of contents 5.1 Performing measurements in single test mode ..................................................... 32 5.2 Measurements and inspections ............................................................................ 33 5.2.1 Visual inspection ............................................................................................ 33 5.2.2 Earth continuity resistance ............................................................................. 33 5.2.
MI 3309 BT DeltaGT 9 Table of contents Maintenance .......................................................................................................... 81 9.1 9.2 9.3 9.4 10 Periodic calibration ............................................................................................... 81 Replacing the fuses .............................................................................................. 81 Service ..................................................................................
MI 3309 BT DeltaGT General description 1 General description The multifunctional portable appliance tester DeltaGT is intended to measurements for testing the electrical safety of portable electrical equipment.
MI 3309 BT DeltaGT General description 1.1 Warnings In order to reach a high level of operator safety while carrying out various measurements using the instrument, as well as to keep the test equipment undamaged, it is necessary to consider the following general warnings: Warning on the instrument means »Read the Instruction manual with special care to safety operation«.
MI 3309 BT DeltaGT General description 1.2 Battery and charging The instrument uses six AA size rechargeable NiCd or NiMH battery cells. Alkaline battery cells are not allowed. Battery condition is always displayed in the upper right corner of the display. If the battery power becomes too weak, the instrument indicates this as shown in Figure 1.1. This indication appears for a few seconds and then the instrument turns itself off. Figure 1.
MI 3309 BT DeltaGT General description 1.3 New battery cells or cells unused for a longer period Unpredictable chemical processes can occur during the charging of new battery cells or cells that have been left unused for a longer period (more than 3 months). NiMH and NiCd cells can be subjected to these chemical effects (sometimes called the memory effect). As a result the instrument operation time can be significantly reduced during the initial charging/discharging cycles of the batteries.
MI 3309 BT DeltaGT General description 1.
MI 3309 BT DeltaGT Instrument description 2 Instrument description 2.1 Front panel Figure 2.1: Front panel Legend: 1 2 3 4 5 6 7 8 LCD FAIL PASS TEST UP DOWN MEM TAB 9 ON / OFF ESC 10 128 x 64 dots matrix display with backlight. Red indicator Indicates PASS / FAIL of result. Green indicator Starts testing / confirms selected option. Selects parameter / changes value of selected parameter. Store / recall / clear tests in memory of instrument.
MI 3309 BT DeltaGT Instrument description 2.2 Connector panel Figure 2.2: Connector panel Legend: 1 S/EB 2 IEC 3 LN 4 PE 5 FUSE compartment 6 MAINS 7 USB connector 8 PS/2 connector Probe and Earth continuity terminal IEC test terminal LN terminal (for connection of fixed installed appliances) PE terminal (for connection of fixed installed appliances) Fuses: 2 x T16 A / 250 V; breaking capacity: 1500 A (for protection against overload and short circuit) Mains supply connector and test terminal.
MI 3309 BT DeltaGT Instrument description 2.3 Back side Figure 2.3: Back side Legend: 1 2 3 4 5 Inserts for side belt Battery compartment cover Fixing screw for battery compartment cover Back side information label Holder for inclined position of the instrument Figure 2.
MI 3309 BT DeltaGT Instrument description 2.4 Meaning of symbols and messages on the instrument display Before performing a measurement, the instrument performs a series of pre-tests to ensure safety and to prevent any damage. These safety pre-tests are checking for any external voltage and load condition on test terminals. If a pre-test fails, an appropriate warning message will be displayed. Warnings and protective measures are described in this chapter. WARNING! Improper supply voltage warning.
MI 3309 BT DeltaGT Instrument description WARNING! A high resistance between L and N was measured in the fuse pre-test. This indication means that the device under test has very low power consumption or it is: not connected; switched off; contains a fuse that has blown. Select PROCEED or CANCEL.
MI 3309 BT DeltaGT Instrument description Measurement is in progress. Test result can be saved. Connect the test lead to the S/EB test terminal. Flex the mains cable of appliance during the test. Check that the device under test is switched on (to ensure that the complete circuit is tested). Connect the cord to be tested to the IEC test terminal. Test passed. Test failed. Measurement was aborted due to improper conditions during the test. RCD must be switched on.
MI 3309 BT DeltaGT Technical specifications 3 Technical specifications 3.1 Earth continuity Range 0.00 19.99 20.0 199.9 200 1999 Resolution 0.01 0.1 1 Accuracy (5 % of reading + 3 digits) Indication only Powered by ............................... battery or mains Test currents ............................. 200 mA into 2.00 Open circuit voltage .................. <9 V AC Pass levels................................ 0.10 , 0.20 , 0.30 , 0.40 , 0.50 , 0.60 , 0.70 , 0.
MI 3309 BT DeltaGT Technical specifications LN (terminal) ↔ S/EB (probe) appliances) (for fixed installed 3.3 Substitute leakage current Range 0.00 mA 9.99 mA 10.0 mA 20.0 mA Resolution 0.01 mA 0.1 mA Accuracy (5 % of reading + 3 digits) Powered by ............................... battery or mains Open circuit voltage .................. <50 V AC at rated mains voltage Short circuit current ................... <40 mA Pass levels: Substitute leakage .................... 0.25 mA, 0.50 mA, 0.75 mA, 1.
MI 3309 BT DeltaGT Technical specifications 3.5 Polarity test Powered by ............................... battery or mains Test voltage .............................. <50 V AC Detects...................................... PASS, L OPEN, N OPEN, PE OPEN, L-N CROSS, L-N SHORT, L-PE SHORT, N-PE SHORT, MULTIPLE FAULT. Test terminals ........................... Test socket ↔ IEC test terminal 3.6 Differential leakage current Range 0.00 mA 19.99 mA Apparent power Resolution 0.
MI 3309 BT DeltaGT Technical specifications 3.8 PRCD and RCD testing 3.8.1 General RCD Trip-out time Complete measurement range corresponds to EN 61557-6 requirements. Maximum measuring times set according to selected reference for RCD testing. Range 0 ms 300 ms (½IN) 0 ms 300 ms (IN) 0 ms 40 ms (5IN) Resolution 0.1 ms 0.1 ms 0.1 ms Accuracy 3 ms 1 ms Powered by ..................... mains Test current..................... ½IN, IN, 5IN (IN = 10 mA, 15 mA, 30 mA) Start angle .........
MI 3309 BT DeltaGT Technical specifications 3.9 Power Apparent power Range 0.00 kVA 4.00 kVA Resolution 0.01 kVA Accuracy (5 % of reading + 3 digits) Resolution 0.01 A Accuracy (5 % of reading + 3 digits) Current Range 0.00 A 16.00 A Powered by ..................... mains Test duration: ............................ 2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, --- s Test terminals ........................... MAINS terminal * Measurement is limited to 120 s if (Iload > 10 A). 3.
MI 3309 BT DeltaGT Technical specifications 3.12 General data Power supply voltage ........................ 9 VDC (61.2 V NiMH or NiCd battery, type HR 6) (size AA) Operation .......................................... typical 8 h Battery charging current.................... 250 mA (internally regulated) Overvoltage category......................... CAT II / 300 V Protection classification .................... double insulation Pollution degree ................................ 2 Protection degree case ...
MI 3309 BT DeltaGT Main menu and test modes 4 Main menu and test modes 4.1 Instrument Main menu From the Main menu of the instrument there are five instrument operation modes, Help menu and Setup menu can be selected: Figure 4.1: Instrument Main menu Keys: / TAB TEST Select one of the following menu items: pre-programmed test sequences according to the VDE 0701-0702 norm, see chapter 6.
MI 3309 BT DeltaGT Main menu and test modes 4.3 Single test menu In single test menu individual tests can be performed. Figure 4.3: Single test menu See chapter 5 Single test for more information. 4.4 Custom Autotest menu This menu contains a list of custom prepared autosequences. The commonly used autotest sequences are added to the list by default. Up to 50 custom autotest sequences can be pre-programmed in this autotest mode. Custom autotest can also be uploaded from the PC SW PATLink PRO Plus.
MI 3309 BT DeltaGT Main menu and test modes Figure 4.6: Code Autotest menu See chapter 6.4 Code Autotest for more information. 4.7 Help menu Help menu contains schematic diagrams to illustrate how to correctly connect a device under test to the PAT testing instrument. Figure 4.7: Example of help screens Keys: / TEST, ESC Selects next / previous help screen. Returns to Main menu. 4.8 Setup menu In the Setup menu different parameters and settings of the instrument can be viewed or set. Figure 4.
MI 3309 BT DeltaGT TEST ESC Main menu and test modes factory settings; sound control. Confirms selection. Returns to the Main menu. 4.8.1 Memory In the Memory menu stored results can be recalled, printed or deleted. Labels can be printed and RFID tags can be written in this menu. Figure 4.9: Memory menu See chapter 7 Working with autotest results for more information. 4.8.2 Language selection The instrument language can be set in this menu. Figure 4.
MI 3309 BT DeltaGT Main menu and test modes Options: COM PORT USB: RS232: PRINTER PRN NAME INIT. BT DONGLE (PRN) communication with PC communication with external devices (printer, scanner, RFID reader/ writer, PC) Selects the printer (can be a RS232 or Bluetooth printer). Enters menu for searching Bluetooth printer. Initializes Bluetooth dongle for the printer. Keys: / TEST ESC Selects item to be changed. Selects option and confirms. Returns to Setup menu. Displayed settings are saved.
MI 3309 BT DeltaGT Main menu and test modes 4.8.3.2 Initialization of the Bluetooth dongle Initialization procedure (Bluetooth dongle for the printer): 1. Connect printer’s Bluetooth dongle A 1436 to the instrument’s PS/2 port. 2. Press RESET key on the Bluetooth dongle A 1436 for at least 10 seconds. 3. Select INIT. BT DONGLE (PRN) in Communication menu and press TEST. 4. Wait for confirmation message and beep.
MI 3309 BT DeltaGT Main menu and test modes 4.8.5 Test speed setup In this menu the instrument test speed can be set: Figure 4.14: Test speed menu Options: STANDARD Default option. No pauses during tests (default). FAST Keys: / TEST ESC Selects the speed mode. Confirms selection and returns to Setup menu. Returns to Setup menu without changes. Note: When enabling the fast mode then Visual inspection and Functional Test will be automatically set to PASS. 4.8.
MI 3309 BT DeltaGT Main menu and test modes Figure 4.16: User data menu Keys: / TEST ESC TAB Selects the user name. Confirms selection and returns to Setup menu. Returns to Setup menu without changes. Enters Edit user data menu. Edit user data: Figure 4.17: Edit user data menu Keys: / TEST MEM ESC Selects a letter. Selects the next letter. Confirms name and returns to User data menu. Deletes last letter. Returns to User data menu without changes.
MI 3309 BT DeltaGT Keys: / TEST, ESC Main menu and test modes Switches between Instrument data screens. Returns to Setup menu. 4.8.9 Initial settings In this menu the following instrument parameters can be set to their initial values: all measurement parameters in single test mode; LCD settings; language; communication settings; internal Bluetooth module is initialized; custom autotest sequences are replaced by factory pre-programmed ones. Figure 4.
MI 3309 BT DeltaGT Single test 5 Single test In a Single test mode individual tests can be performed. This is especially helpful for troubleshooting. 5.1 Performing measurements in single test mode Select appropriate test in Single test menu. Figure 5.1: Single test menu Keys: / TEST ESC Selects a single test. Enters Single test measuring menu. Returns to Main menu. A single test can be started from any Single test measuring menu. Before carrying out a test the parameters / limits can be edited.
MI 3309 BT DeltaGT Single test 5.2 Measurements and inspections 5.2.1 Visual inspection A thorough visual check must be carried out before each electrical safety test. The following items should be checked: Inspection of the device under test for sign of damage. Inspection of the flexible power supply cable for damage. Any signs of pollution, moisture, dirt that can jeopardize safety.
MI 3309 BT DeltaGT LIMIT TIME Single test Maximum resistance [0.10 0.20 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 2.00] Measuring time [2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, --- s (continuous measurement)] Typical test circuits for Earth continuity resistance measurement Figure 5.5: Measurement of Earth continuity Figure 5.
MI 3309 BT DeltaGT Single test Figure 5.7: Examples of Earth continuity resistance measurement results Displayed results: Main result ............. Earth continuity resistance Note: Consider displayed warnings before starting measurement! If the PRCD test is set in the Autotest procedure power is applied to the mains test socket during the earth continuity test.
MI 3309 BT DeltaGT Single test Test circuits for Insulation resistance measurement Figure 5.9: Measurement of insulation resistance Figure 5.10: Measurement of insulation resistance of fixed installed DUTs of Class I Insulation resistance measurement procedure Select the INSULATION function. Set the test parameters. Connect device under test to the instrument (see Figure 5.9 and Figure 5.10). Press the TEST key for measurement. To stop continuous measurement press TEST key once again.
MI 3309 BT DeltaGT Single test Notes: When S/EB probe is connected during the test then the current through it is also considered. Consider any warning on the display before starting the measurement! Do not touch or disconnect the device under test during the measurement or before it is fully discharged! The message »Udisch …« will be displayed while the voltage on the device is higher than 10 V! 5.2.
MI 3309 BT DeltaGT Single test Figure 5.14: Measurement of insulation resistance of fixed installed DUTs Insulation resistance - P measurement procedure Select the INSULATION-P function. Set the test parameters. Connect device under test to the instrument (see Figure 5.13 and Figure 5.14). Press the TEST key for measurement. To stop continuous measurement press TEST key once again. Store the result by pressing MEM key (optional). Figure 5.
MI 3309 BT DeltaGT Single test Figure 5.16: Substitute leakage menu Test parameters for Substitute leakage current measurement OUTPUT LIMIT TIME Test voltage [30 V] Maximum current [0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.00 mA, 2.25 mA, 2.50 mA, 3.50 mA, 4.00 mA, 4.50 mA, 5.00 mA, 5.50 mA, 6.00 mA, 7.00 mA, 8.00 mA, 9.00 mA, 10.0 mA, 15.
MI 3309 BT DeltaGT Single test Substitute leakage measurement procedure Select the SUB. LEAKAGE function. Set the test parameters. Connect device under test to the instrument (see Figure 5.17 and Figure 5.18). Press the TEST key for measurement. To stop continuous measurement press TEST key once again. Store the result by pressing MEM key (optional). Figure 5.19: Example of substitute leakage current measurement results Displayed results: Main result .............
MI 3309 BT DeltaGT Single test Test parameters for substitute leakage - P current measurement OUTPUT LIMIT TIME Test voltage [30 V] Maximum current [0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.00 mA, --- mA] Measuring time [2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, --- s (continuous measurement)] Test circuits for substitute leakage - P measurement Figure 5.21: Measurement of Substitute leakage - P current Figure 5.
MI 3309 BT DeltaGT Single test Figure 5.23: Example of substitute leakage - P current measurement results Displayed results: Main result…………..Substitute leakage current (LN – P) Notes: Consider any displayed warning before starting measurement! The currents flowing through the PE terminal or PE of the test socket will not be considered. 5.2.7 Polarity test This test checks the polarity of supply cords.
MI 3309 BT DeltaGT Single test Figure 5.26: Examples of polarity test result Displayed results: Main result ............. PASS/FAIL, description of fault Note: Consider any displayed warnings before starting test! 5.2.8 Differential leakage The purpose of this test is to determine the sum of all leakages flowing from the live conductors to the earth. The differential method allows measuring the full and true leakage current, even if there are parallel current paths from the DUT to ground.
MI 3309 BT DeltaGT Single test Test circuit for differential leakage current measurement Figure 5.28: Measurement of differential current Differential current measurement procedure Select the DIFF. LEAKAGE function. Set the test parameters. Connect device under test to the instrument (see Figure 5.28). Press the TEST key for measurement. To stop continuous measurement press TEST key once again. Store the result by pressing MEM key (optional). Figure 5.
MI 3309 BT DeltaGT Single test 5.2.9 Touch leakage This test determines the current that would flow if a person touches accessible conductive parts of the DUT. The instrument measures: The touch leakage current flowing through the S/EB probe into earth. The DUT can be powered from the test socket or directly from the installation (fixed installed equipment). Figure 5.
MI 3309 BT DeltaGT Single test Figure 5.32: Measurement of touch leakage current on a fixed installed DUT Touch leakage current measurement procedure Select the TOUCH LEAKAGE function. Set the test parameters. Connect device under test to the instrument (see Figure 5.31 and Figure 5.32). Press the TEST key for measurement. To stop continuous measurement press TEST key once again. Store the result by pressing MEM key (optional). Figure 5.
MI 3309 BT DeltaGT Single test 5.2.10 (P)RCD test The purpose of this test is to ensure the proper operation of residual current devices (RCD) built into appliances / installations and portable residual current devices (PRCD). Trip-out time measurement verifies the sensitivity of a (P)RCD at selected residual currents. The RCD test is performed via the instrument mains supply input. Figure 5.
MI 3309 BT DeltaGT Single test b) Testing of PRCD via mains socket c) Testing of PRCD via test socket Figure 5.36: Testing of RCD and PRCD 5.2.10.1 (P)RCD single test In single test a fast (P)RCD test with selected test current and with one or both starting polarities is performed. Trip-out time measurement procedure Measuring PRCD Select the RCD test function. Select test mode. Set test parameters. Connect tested PRCD / device to an external voltage socket.
MI 3309 BT DeltaGT Single test Measuring RCD Select the RCD test function. Select test mode. Set test parameters. Connect the DeltaGT MAINS terminal to mains socket protected by tested RCD (see Figure 5.36a). Depending on the type of PRCD, it may be necessary to manually switch it on. Press the TEST key to perform measurement. If both starting polarities are selected: Reactivate tested RCD. Store the result by pressing MEM key (optional). Figure 5.
MI 3309 BT DeltaGT Single test Test with 5IN, 180 (step 4). Re-activate (P)RCD. Test with ½IN, 0 (step 5). Test with ½IN, 180 (step 6). (P)RCD should trip-out (P)RCD should not trip-out (P)RCD should not trip-out End of test. Displayed results: Main results ........... trip-out times at different currents / starting polarities U ............................
MI 3309 BT DeltaGT Single test Test circuit for the power test Figure 5.39: Power test Power test procedure Select the POWER function. Set test parameters. Connect device under test to the instrument’s test socket and switch it on (see figure 5.39). Connect the instrument to mains voltage. Press the TEST key for measurement. To stop continuous measurement press TEST key once again. Store the result by pressing MEM key (optional). Figure 5.
MI 3309 BT DeltaGT Single test 5.2.12 Voltage TRMS In this function the voltage across the MAINS terminal is measured continuously. Test circuit for voltage measurement Figure 5.41: Voltage measurement with the IEC cord Voltage TRMS procedure Select the VOLTAGE TRMS function. Connect the IEC cord to the instrument’s MAINS terminal and to the external mains socket as shown on Figure 5.41. Store the result by pressing MEM key (optional). Figure 5.
MI 3309 BT DeltaGT Single test Figure 5.43: Clamp current menu Test parameters for clamp current measurement LIMIT Maximum current [0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.00 mA, 2.25 mA, 2.50 mA, 3.50 mA, 4.00 mA, 4.50 mA, 5.00 mA, 5.50 mA, 6.00 mA, 7.00 mA, 8.00 mA, 9.00 mA, 10.0 mA, 15.0 mA, --- mA] TIME Measuring time [2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, none] Test circuit for clamp current measurement Figure 5.
MI 3309 BT DeltaGT Single test Figure 5.45: Examples of clamp current measurement result Displayed results: Main result ............. clamp current Notes: When measuring leakage currents, the neighbouring magnetic fields and capacitive coupling (especially from the L and N conductors) can disturb the results. It is recommended that the clamp is as close as possible to the grounded surface and away from wires and other objects under voltage or carrying current.
MI 3309 BT DeltaGT Autotest sequences 6 Autotest sequences Autotest is the fastest and easiest way to test appliances. During the autotest, preprogrammed measurements run automatically in a sequential order. The complete autotest results can be stored together with the associated appliance ID, appliance NAME Re-test period, and Location. 6.1 VDE organizer setup menu Select VDE Organizer in Main menu. In the first step the appliance type, protective measures and additional protection should be set.
MI 3309 BT DeltaGT Autotest sequences 6.1.1 Carrying out a test sequence set with VDE Organizer General meaning of keys during VDE Organizer autotest sequence: Sets Organizer option or set value in selected (highlighted) item. / Cancels VDE sequence and returns to Main menu. ESC Starts / repeats selected measurement or proceeds to next step. TEST After the test sequence is finished the instrument proceeds to the ‘Autotest Result’ menu.
MI 3309 BT DeltaGT Autotest sequences 6.1.1.3 Insulation resistance measurement This test is offered if it is applicable according to the VDE Organizer setting. The measurement is described in chapter 5.2.3 Insulation resistance. Figure 6.
MI 3309 BT DeltaGT Autotest sequences Note: The leakage current limit is automatically set on base of set appliance power. There are no special options to be set if other Appliance type is set. Figure 6.9: VDE Organizer – Substitute leakage result screen There are no special options to be set in the Substitute leakage result screen. 6.1.1.6 Differential leakage measurement The measurement is described in chapter 5.2.8 Differential leakage. Figure 6.
MI 3309 BT DeltaGT 250 V Autotest sequences To be set if overvoltage protection devices are installed or SELV/ PELV protection measures. Figure 6.13: VDE Organizer – Insulation resistance - P result screen Options in Insulation resistance - P result screen: Proceeds with the next measurement. NEXT REPEAT Repeats the test (use in case of multiple isolated / SELV / PELV accessible points). Lowest result will be stored. 6.1.1.
MI 3309 BT DeltaGT Autotest sequences 6.1.1.10 Touch leakage measurement The measurement is described in chapter 5.2.9 Touch Leakage. Figure 6.17: VDE Organizer – Touch leakage starting screen There are no special options in the Touch leakage - starting screen. Figure 6.18: VDE Organizer – Touch leakage result screen Options in Touch leakage result screen: Proceeds with the next measurement. NEXT REPEAT Repeats the test (use in case of multiple isolated / SELV / PELV accessible points).
MI 3309 BT DeltaGT Autotest sequences There are no special options in the RCD result screen. 6.1.1.12 Functional test Figure 6.21: VDE Organizer – Functional test result screens Options in Functional test result screen: To be applied manually. PASS / FAIL Option if General or Appliance with heating elements is set: Starts Power test. The measurement is described in chapter 5.2.11 POWER Power test. Option if IEC leads, multiple outlets without electronic parts is set: Starts Polarity test.
MI 3309 BT DeltaGT Autotest sequences Note: If more than 50 autotests are saved, »Out of custom autotest memory« message is displayed. 6.3 Simple test Simple test sequences are commonly used pre-programed autotest sequences with possibility of fast testing. Fast testing mode can be enabled in Test speed setup function in Setup menu. Refer to chapter 4.8.5 Test speed setup for more information. Select Simple test function in Main menu. Figure 6.
MI 3309 BT DeltaGT Autotest sequences Reading a Code autotest sequence (with barcode scanner, RFID reader/writer or manually) Connect barcode scanner or RFID reader/writer to the instrument RS232 / PS2 connector first. Figure 6.24: Connection of the barcode scanner and RFID reader/writer Set RS 232 communication port using Communication function in Setup menu. Select Code Autotest in Main menu. The latest received or set autotest sequence name and its code is displayed.
MI 3309 BT DeltaGT Autotest sequences For more information refer to Chapter 8.3 Bluetooth communication and PATLink Android instructions. 6.5 Carrying out autotest sequences (Code, Simple and Custom) General meaning of keys during Code, Simple and Custom autotest sequences: TAB, / ESC TEST Set option. Set limit value in selected (highlighted) item. Cancels autotest sequence and returns to the autotest menu (Code, Simple or Custom) without changes.
MI 3309 BT DeltaGT Autotest sequences Options in Earth continuity result screen: Proceeds to next step. NEXT Repeat the test (use in case of multiple earthed points). Highest result will REPEAT be stored. 6.5.3 Insulation resistance measurement The test is offered if it is applicable according to the autotest setting. The Insulation starting screen is displayed first. Measurement and options in Insulation starting screen are described in chapter 5.2.3 Insulation resistance. Figure 6.
MI 3309 BT DeltaGT Autotest sequences Figure 6.33: Differential leakage starting screen After the measurement is carried out the Differential leakage result screen is displayed. Figure 6.34: Differential leakage result screen There are no special options to be set in the Differential leakage result screen. 6.5.6 Insulation resistance - P measurement The test is offered if it is applicable according to the autotest setting. The Insulation resistance - P starting screen is displayed first.
MI 3309 BT DeltaGT Autotest sequences Figure 6.37: Substitute leakage - P starting screen After the measurement is carried out the Substitute leakage - P result screen is displayed. Figure 6.38: Substitute leakage - P result screen Options in Substitute leakage - P result screen: Proceeds with the next measurement. NEXT REPEAT Repeats the test (use in case of multiple isolated / SELV / PELV accessible points). Highest result will be stored. 6.5.
MI 3309 BT DeltaGT Autotest sequences (P)RCD single test menus (P)RCD autotest menu Figure 6.41: (P)RCD test starting screen After the measurement is carried out the (P)RCD test result screen is displayed. Figure 6.42: Example of Custom / Code Autotest – (P)RCD test result screen There are no special options in the (P)RCD result screen. 6.5.10 Polarity test The test is offered if it is applicable according to the autotest setting. The Polarity test starting screen is displayed first.
MI 3309 BT DeltaGT Autotest sequences Figure 6.45: Power starting screen After the measurement is carried out the Power result screen is displayed. Figure 6.46: Power result screen There are no special options to be set in the Power result screen. 6.5.12 Clamp current measurement The test is offered if it is applicable according to the autotest setting. The Clamp current starting screen is displayed first. Measurement and options in Clamp current starting screen are described in chapter 5.2.
MI 3309 BT DeltaGT Autotest sequences 6.6 Handling autotest results After the Code, Simple or Custom autotest is finished, the Main autotest result screen will be displayed including an overall / (PASS / FAIL) indication. Figure 6.50: Main autotest result screen Options in Autotest results screen: Views individual results. VIEW RESULTS Returns to Code, Simple or Custom menu. NEW TEST Saves Autotest results. SAVE RESULTS Refer to chapter 7.
MI 3309 BT DeltaGT Autotest sequences Saving autotest as Custom autotest In the Save as custom screen last autotest can be stored as CUSTOM AUTOTEST. Options in Save as custom screen: Edit autotest name. / , TEST MEM (SAVE) Saves autotest name. ESC (DEL) Deletes last character of autotest name. ESC (CANCEL) Returns to the previous menu. Figure 6.
MI 3309 BT DeltaGT Working with autotest results 7 Working with autotest results 7.1 Saving autotest results After selecting Save results in Autotest results menu, the autotest results will be stored in the internal memory of the instrument. The appliance ID number, NAME, Re-Test period and LOCATION can be added to the test results before the results are saved: Figure 7.1: Save results menu (Appliance ID) Keys: / , TEST MEM (OK) ESC (DEL) ESC (CANCEL) Edit Appliance ID data. Saves Appliance ID.
MI 3309 BT DeltaGT Working with autotest results Figure 7.3: List menu (Appliance NAME) The Re-Test period can be entered. Figure 7.4: Save results menu (Re-Test period) Keys: / , TEST ESC (CANCEL) MEM (SAVE) Set Re-Test period in months. Returns to the previous menu. Saves Re-Test period and returns to Autotest result menu. The Re-Test period can also be scanned with RFID reader/writer or using PATLink android application on mobile devices (QR codes).
MI 3309 BT DeltaGT Working with autotest results 7.2 Recalling results Saved autotest results can be recalled, printed or deleted from the Memory menu. Enter the Memory menu from the Setup menu. Figure 7.6: Memory menu To enter the Recall results menu select Recall results in Memory menu. A list of Appliance ID’s and NAMES are displayed in a chronological order (last performed measurement is displayed at the bottom of the list).
MI 3309 BT DeltaGT Working with autotest results 7.3 Deleting individual autotest results To enter the Delete results menu select Delete results in Memory menu. A list of Appliance ID’s and NAMES are displayed in a chronological order (last performed measurement will be displayed at the bottom of the list). In the lower window of the display the following data is displayed: Appliance ID, NAME; date and time of the selected test; the overall / status of the selected test. Figure 7.
MI 3309 BT DeltaGT Working with autotest results 7.5.1 Printing labels / writing in RFID tags from Autotest Results menu To print a label or write data to a TAG an Autotest must be saved first. Refer to chapter 7.1 Saving Autotest results for more information. Figure 7.13: Autotest result screen Options in Autotest results screen (after Autotest is saved): VIEW RESULTS NEW TEST Views individual results. Returns to Code, Simple or Custom menu. SAVE AS CUSTOM Saves test setup as Custom autotest.
MI 3309 BT DeltaGT Working with autotest results Figure 7.15: Options in Printer menu Print Simple label A simple appliance label will be printed. Print label An appliance label with barcode will be printed. Print results All the data stored at the specified location will be printed. That includes Appliance ID, Appliance NAME, Test date and time, Overall and individual measurement result (Pass or Fail), individual measurement values, limits and other settings.
MI 3309 BT DeltaGT Communication 8 Communication There are three communication interfaces for communication with PC and other external devices: USB, RS232 and Bluetooth. See chapter 4.8.3 Communication for more information. 8.1 USB communication How to establish an USB link: COM PORT: USB should be selected in Communication menu. Connect the PC USB port to the instrument USB connector using the USB interface cable. Switch on the PC and the instrument. Run the PATLink PRO program.
MI 3309 BT DeltaGT Communication 8.3 Bluetooth communication The internal Bluetooth module enables easy communication via Bluetooth with PC and Android devices. How to configure a Bluetooth link between instrument and PC: Switch On the instrument. On PC configure a Standard Serial Port to enable communication over Bluetooth link between instrument and PC. Usually no code for pairing the devices is needed. Run the PATlinkPRO program. Set communication port and baud rate speed.
MI 3309 BT DeltaGT Communication If printing from Android device be sure that Zebra printer is selected in the Metrel Android application as the Bluetooth printer. The configuration tool is available in the Metrel's Android application. The instrument and printer are ready to communicate. Notes: The name of a correctly configured Bluetooth device must consist of the instrument type plus serial number, eg. MI 3309 BT-12240429D.
MI 3309 BT DeltaGT Maintenance 9 Maintenance 9.1 Periodic calibration It is essential that all measuring instruments are regularly calibrated in order for the technical specification listed in this manual to be guaranteed. We recommend an annual calibration. The calibration has to be done by an authorized technical person only. 9.
MI 3309 BT DeltaGT Instrument set and accessories 10 Instrument set and accessories Standard set of the instrument Instrument MI 3309 BT DeltaGT Test probe, black Crocodile clip, black Test lead, 1.5 m, black 2 x IEC cord, 1.
MI 3309 BT DeltaGT Appendix A – Barcode and QR code formats Appendix A – Barcode and QR code formats Barcode formats The instrument DeltaGT supports two barcode formats (single and double). Autotest code and appliance ID Autotest codes are represented as a three digits code. These autotest codes can also be represented by the barcode. Using the barcode scanner, the instruments can accept autotest code from barcode label. A0 1 Autotest code Also appliance ID can be read from barcode label.
MI 3309 BT DeltaGT Appendix B – Pre-programmed autotests (GER) Appendix B – Pre-programmed autotests (GER) Pre-programmed autotest sequences table Autotest shortcut code Visual test Output Limit Time Output Limit Insulation Time Output Insulation Limit (probe) Time Output Sub. leakage Limit Time Output Sub.
MI 3309 BT DeltaGT Appendix B – Pre-programmed autotests (GER) Pre-programmed autotest sequences table (cont’d) Autotest shortcut code Visual test Output Limit Time Output Limit Insulation Time Output Insulation Limit (probe) Time Output Sub. leakage Limit Time Output Sub.
MI 3309 BT DeltaGT Appendix B – Pre-programmed autotests (GER) Pre-programmed autotest sequences table (cont’d) Autotest shortcut code Visual test Earth continuity Insulation Insulation (probe) Sub. leakage Sub.
MI 3309 BT DeltaGT Appendix B – Pre-programmed autotests (GER) METREL GmbH VDE tester test type card Code Autotest sequence name and descriptions A01 Kl_1_Iso A02 Kl_1_Iso_BLT A03 Kl_1_Ia A04 Kl_1_Ia_BLT A05 Kl_2_Iso A06 Kl_2_Ibs Testing according to VDE. Class 1 device. Earth bond: 0.30 Insulation resistance and substitute Insulation: 1.00 M leakage current measurements are Sub leakage: 3.50 mA applicable. Testing according to VDE. Class 1 device with isolated accessible cinductive parts.
MI 3309 BT DeltaGT Appendix B – Pre-programmed autotests (GER) METREL GmbH VDE tester test type card (cont'd) A10 Kl_2 A11 Kl_3_Iso A12 Kl_3 Testing according to VDE. Class 2 device without any isolated accessible conductive parts. Testing according to VDE. Insulation - P: 0.25 M Class 3 device with isolated accessible conductive parts. Testing according to VDE. Class 3 device without any isolated accessible conductive parts.
MI 3309 BT DeltaGT Appendix C – Simple test codes (GER) Appendix C – Simple test codes (GER) Simple test codes Visual test Output Earth continuity Limit Time Output Limit Insulation Time Output Insulation Limit (probe) Time Output Limit Sub. leakage Time Output Sub. leakage Limit (probe) Time Output Differential Limit leakage Time Output Limit Touch leakage Time Functional test Output Limit Power* Time Output TRMS clamp Limit current Time Polarity test CLASS I CLASS II CLASS III 200 mA 0.
MI 3309 BT DeltaGT Appendix D – Pre-programmed autotests (NL) Appendix D – Pre-programmed autotests (NL) Pre-programmed autotest sequences table Autotest shortcut code Visual test Output Limit Time Output Limit Insulation Time Output Insulation Limit (probe) Time Output Sub. leakage Limit Time Output Sub.
MI 3309 BT DeltaGT Appendix D – Pre-programmed autotests (NL) Pre-programmed autotest sequences table (cont’d) Autotest shortcut code 05 06 07 08 KL_3_ALG KL_1_AGMD KABEL_5M_2.5MM KABEL_15M_2.5MM Visual test Output Limit Time Output Limit Insulation Time Output Insulation Limit (probe) Time Output Sub. leakage Limit Time Output Sub.
MI 3309 BT DeltaGT Appendix D – Pre-programmed autotests (NL) Pre-programmed autotest sequences table (cont’d) Autotest shortcut code Visual test Output Limit Time Output Limit Insulation Time Output Insulation Limit (probe) Time Output Limit Sub. leakage Time Output Sub.