M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 M130 – THERMOCOUPLE CALIBRATOR USER MANUAL Meriam Process Technologies’ M130 is a handheld calibrator for thermocouple (T/C) measurement or simulation. Eight T/C types are supported referencing NIST175 and ITS90 standards. Temperature can be displayed in selectable engineering units of ºC, °F, ºR, K www.meriam.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 Table of Contents Introduction .............................................................................3 User Interface...........................................................................4 Keypad Functions ....................................................................4 ON/OFF Key ...........................................................................4 BACKLIGHT Key..................................................
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 Introduction Thank you for purchasing this Meriam product. Meriam has been providing innovative, reliable, cost-effective measurement and calibration solutions since 1911. The M130 Thermocouple calibrator has been designed using the latest technology to provide reliable accuracy throughout its entire operating temperature range, dependable operation, and exceptional battery life.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 User Interface Keypad Functions ON/OFF Key The ON/OFF key, represented by the standard ON/OFF symbol, turns the M130 ON or OFF (note: the ON/OFF key must be held until the display turns on or off). Upon power on, the M130 will perform a self-test, displaying the Meriam product ID, and all segments for approximately 1-2 seconds, then go into the mode in use when unit was powered off.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 FUNCTION Key The FUNCTION key, represented by the FUNC symbol, selects one of 3 functional modes: Measure, Simulate (manual), or Simulate (auto), and displays the corresponding icons (MEAS, SIM). This key can also be used to advance one menu item or will advance to next recorded value while in RECORD view mode.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 adjustment for changing the values. The positive sign (+) indicates the Group Keys will increment the displayed value. DECREMENT ( – ) Group Keys The three keys beneath this symbol allow the user to scroll through edit selections when EDIT icon is lit. The keys are also used to decrement numeric simulation values.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 During an over-range/under-range condition, the red backlight will over-ride the green backlight if it was already on. Once the condition is corrected, the green backlight will be restored to its previous state (assuming the backlight auto-off timer did not expire). Notes: 1. The backlight should be turned off when not needed to conserve battery power. 2. The backlight will be automatically disabled when the M130 reaches battery mode.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 FUNCTIONAL MODES, STATES, AND KEY FUNCTIONS Keys SETUP FUNC X W UNITS ENTER Increment (+) Group Decrement (-) Group Run Setup Edit Setup MEAS or SIM Enters Setup state, or ends simulation activity Next Functional mode Next engineering unit Toggles lower display line between TC type / temp.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 • ENTER key accepts the menu item to edit and saves the selection. During the SETUP EDIT state the W UNITS key goes back one increment value, and the FUNC X key advances one increment value. • T/C measurements can be recorded to memory either manually or automatically. See the Record Mode section of this manual for more information.
M130 – T/C Calibrator MEAS SETUP: Measure Mode Setup menu structure T/C TYPE mV B E J K N R S T OFF TMR 30 MIN OFF F RECAL ZERO VIEW RECORD MANUAL AUTO Note: Instruction Manual Z1017-IR E.O. 6566 Sept.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 SIM (AUTO) SETUP: Simulation Mode’s Manual menu structure AUTO A.STEP 1 START UNITS C F R K TC TYPE mV B E J K N R S T URV LRV Press SETUP to enter Title of Auto Step 1 simulation; setup saved here for subsequent re-use Starts A.
M130 – T/C Calibrator K TC TYPE mV B E J K N R S T URV LRV UP TIME DN TIME DELAY Instruction Manual Z1017-IR E.O. 6566 Sept.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 Engineering Units Select Engineering Units are conveniently selected when actively measuring or in Manual Simulation mode. Press the W UNITS key to display any of the following: ºC, °F, ºR, K. NOTE: to display mV units, select the mV option under MEAS SETUP / TC TYPE. The W UNITS key will not change display units for this TC TYPE. Record Mode The M130 can store up to 240 temperature measurements in a single record (REC) session.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 session will stop and the RECORD Data mode will automatically exit to Measure Mode. VIEW Recorded Data: To view the results of a REC session from Measure Mode, press SETUP key and then W or X key until VIEW is displayed. Press ENTER. The REC icon is shown in upper right. Display will show recorded value 1 of X (1/X). Pressing ENTER will toggle to show the T/C type and engineering unit.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 Specifications Type and Range: T/C types B, E, J, K, N, R, S, T & milli-volts Milli-volts range -15mV to +80.00 mV T/C Measurement & Simulation Accuracy: See Table below mV Accuracy: 0.005% of reading ±6µV mV Temp. Effect: ±2 µV /°C (Factory Ref. Temp. = 23°C) Cold Junction Effect: ±0.05°C/°C Cold Junction Accuracy: ±0.1°C at 25°C T/C Resolution: T/C types E, J, K, N, T: 0.1°C T/C types B, R, S: 1°C mV: 0.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 T/C Measurement & Simulation Accuracy Table* Find T/C type in left column and follow the row across to the temperature range of interest. Then follow the column up to determine the potential error in degrees C. T/C ±3.0° C ±0.3° C ±0.2° C ±0.1° C E -200 to -179°C -179 to 15°C 15 to 1000°C J -200 to -165°C -165 to 1200°C -200 to 110°C 110 to 1300°C -168 to -86°C -86 to 346°C Type ±2.0° C ±1.0° C ±0.5° C K ±0.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 Certification The M130 is EMI/ESD compliant to the test specification EN61326. The M130 is for general purpose use only. Do not use in potentially hazardous atmospheres. Changing the Batteries The M130 is powered by three, 1.5 volt AA size batteries. The “BAT” icon in upper left of display will activate when battery output drops below a threshold level. Approximately 2 hours of life remain from that point.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 Install the three batteries by sliding them into the bottom of the battery slots, positive (+) end first, making sure the polarity markings on the batteries align with the markings molded into the battery compartment. When a battery is installed with the polarity reversed, the stand-offs prevent the negative battery terminal from contacting the positive terminal in the battery compartment.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 User Connections Connection: Standard mini-T/C Connector (Measure and Simulate) Warning: The copper receiver blades of the mini-T/C connector in the M130 can be damaged if the mating plugs are misaligned during installation. Take care to properly align the mating plug prior to pressing it into the M130’s connector. www.meriam.
M130 – T/C Calibrator Instruction Manual Z1017-IR E.O. 6566 Sept. 2008 Contact Information If the M130 is damaged, it should be returned to the factory for servicing. Contact the Meriam Process Technologies representative in your area or call the factory at the numbers listed below for a Return Material Authorization (RMA) number. Meriam Process Technologies 10920 Madison Ave. Cleveland, OH 44102 Ph.