Installation Instructions Part 1
Table Of Contents
- Table of Contents
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System 11
- CHAPTER 2 Indications and Contraindications 21
- CHAPTER 3 Personal Pump Communicator (PPC) 25
- CHAPTER 4 Pump Implantation 77
- CHAPTER 5 Pump Refill Procedure 97
- CHAPTER 6 Explanting the Pump System 107
- CHAPTER 7 Warnings And Precautions 109
- CHAPTER 8 Adverse Reactions 115
- CHAPTER 9 System Alarms and Messages 117
- CHAPTER 10 Troubleshooting Pump System Under-delivery 131
- CHAPTER 11 MiniMed 2007C Implantable Insulin Pump System 133
- APPENDIX A Label Information 137
- APPENDIX B Implant Worksheet 141
- APPENDIX C Refill Form 143
- APPENDIX D Precautions and General Procedures 145
- APPENDIX E Pump Rinse Procedure 151
- APPENDIX F Side Port Catheter Flush Procedure 163
- APPENDIX G Stroke Volume Measurement 175
- APPENDIX H Pressure Measurement Using The Side Port Catheter 181
- CHAPTER 1 Description of the Medtronic MiniMed 2007C Implantable Insulin Pump System
- CHAPTER 2 Indications and Contraindications
- CHAPTER 3 Personal Pump Communicator (PPC)
- Introduction
- Part 1: PPC/Pump system initialization
- Part 2: Additional PPC programming features
- Main menu
- Program a bolus
- Set a normal bolus with the variable bolus feature turned off
- Set a normal bolus with the variable bolus feature turned on
- Set a square wave bolus
- Set a dual wave bolus
- Review bolus history
- Suspend mode
- Programming a basal rate
- Programming basal delivery pattern
- Setting basal rate profiles in each delivery pattern
- Program a temporary basal rate
- Set a temporary basal rate
- Stop a temporary basal rate
- Personal events
- History
- Main menu
- Setup Pump
- Setup II
- Audio bolus
- Activating the audio bolus feature
- Set an audio bolus from the main menu bolus screen
- Variable bolus
- Refill
- Priming
- Diagnostic rate
- Initialize to factory defaults
- Download software
- Stop Pump
- Supervisor password
- Exit supervisor
- Personal Pump communicator messages
- Clinical history codes
- Clinical history PPC/Pump codes
- Audio bolus
- CHAPTER 4 Pump Implantation
Part 2: Additional PPC programming features
45
Set a square wave bolus
A Square Wave Bolus of insulin is delivered evenly over a preset period
of time, from 30 minutes to 4 hours. A Square Wave Bolus may be desir-
able when eating long meals such as banquets or receptions, high fat
meals, or to compensate for gastroparesis. During a Square Wave Bolus,
the programmed basal rate is also delivered.
To access this feature you must first turn the Variable Bolus feature “ON”
in the “SETUP II” menu.
1. From the Time/Date screen, press SEL.
The “BOLUS” screen is displayed, show-
ing the last bolus programmed with the
time and date flashing.
2. Press ACT and the “SET BOLUS TYPE”
screen appears. Press the ▲
▲▲
▲ and ▼
▼▼
▼ but-
tons to select “SQUARE”. Press ACT.
3. The “BOLUS” screen appears, with
dashes under “EXT” flashing. Use the ▲
▲▲
▲
and ▼
▼▼
▼ buttons to enter an extended
bolus
amount.
Press ACT.
4. Blinking dashes will now appear under
the bolus amount entered.
Use the ▲
▲▲
▲ and ▼
▼▼
▼ buttons to enter a time
duration for the Square Wave Bolus, in
one-half hour increments from 30 min-
utes to four hours.
5. Press ACT and the “CONFIRM BOLUS”
screen is displayed, with the screen flash-
ing. Confirm the Square Wave Bolus by
pressing ACT again.
08:13 Jan 02
BOLUS
PROG
--
IMM EXT
10 U -- U
SET
SQUARE
BOLUS TYPE
SET BOLUS
IMM EXT
-- U 4.0U
SET BOLUS
IMM EXT
-- U 4.0U
2:00
CONFIRM BOLUS
IMM EXT
-- U 4.0U
2:00