5-2
5.3 Enclosure Leakage Current Tests
Leakage-HI:
300.00 uA
Leakage-LO:
0.00 uA
Voltage HI:
120.0V
Voltage LO:
100.0V
Probe: Ground
To Line
Neutral :
Closed
Reverse: Off Ground: Open
5.3.1 Connect the probe of the Electrical Safety Analyzer to the
Equipotential lug on the back of the monitor.
5.3.2 Set the Ground switch to the Open position on the Electrical
Safety Analyzer.
5.3.3 Turn the monitor on.
5.3.4 Read the current leakage on the Electrical Safety Analyzer.
5.3.5 Set the polarity on the Electrical Safety Analyzer to Reverse.
5.3.6 Read the current leakage on the Electrical Safety Analyzer.
Note: The readings should not exceed 300 µA or 0.3 volts at 100-120V /50-
60Hz on the Electrical Safety Analyzer with the monitor powered on.
Leakage-HI:
100.00 uA
Leakage-LO:
0.00 uA
Voltage HI:
120.0V
Voltage LO:
100.0V
Probe: Ground
To Line
Neutral :
Closed
Reverse: Off Ground:
Closed
5.3.7 Set the Ground switch to the Closed position on the Electrical
Safety Analyzer.
5.3.8 Read the current leakage on the Electrical Safety Analyzer.
5.3.9 Set the polarity on the Electrical Safety Analyzer to Reverse.
5.3.10 Read the current leakage on the Electrical Safety Analyzer.
Note: The readings should not exceed 100 µA or 0.1 volts at 100-120V /50-
60Hz on the Electrical Safety Analyzer with the monitor powered on.
5.3.11 Turn the Monitor off.