List of Components ■ Standard Items MODEL Main Body Main Body with analog output Main Body with pressure sensor Main body with pressure sensor and analog output 6501-00 Qty - 6501-A0 6501-B0 Functions With analog output function One of these With pressure measurement function With pressure measurement and analog output functions 6501-C0 6531-06 1 Air velocity, air temperature, humidity sensor (directivity) Air velocity, air temperature sensor (directivity) Air velocity, air temperature sensor
Important Safety Information The symbols for the warnings used in this manual are defined below: Classifications Danger: To Prevent Serious Injury or Death Warnings in this classification indicate a danger that may result in serious injury or death if not observed. Caution: To Prevent Damage to the Product Warnings in this classification indicate a risk of damage to the product that may void the product warranty if not observed.
Danger Never bring the probe close to a flammable gas atmosphere. The heated sensor may cause a fire or explosion. Forbidden Never touch the sensor. Don’t touch Hot surface The sensor is heated during operation. Touching the heated sensor may cause burns, and may also damage the sensor itself. Do not disassemble or heat the batteries, or put them into a fire. Use properly Explosive Caution This may cause burns and the batteries may burst.
Forbidden Do not use or leave the instrument in a high temperature, high humidity or dusty environment. Do not leave this instrument under direct sunlight for a prolonged period. The instrument may not function properly out of the specified operating conditions. Do not subject the instrument or the probe to strong impacts. Forbidden Dropping the unit or placing heavy objects on it may cause damage or malfunction to the instrument. Never disassemble, modify or repair the product.
Do not move the main unit and the probe from a cold place to a warm place quickly. It will cause condensation. Forbidden Even when used in an environment within the specified operating temperature and humidity, a sudden temperature change may cause condensation. Condensation generated on the sensor may cause inaccurate measurements. Condensation on metal parts may cause rusting and lead to a malfunction. Do not touch the LCD screen with a sharp-pointed object or with excessive pressure.
Table of Contents §1 Part Names and Functions ....................................................................... 7 Main Body ............................................................................................................. 7 Touch Panel ........................................................................................................... 8 Probe .................................................................................................................... 9 Probe Cable ..........
§1 Part Names and Functions Main Body Pressure Port ■ Pressure port is only available on MODEL6501-B0/-C0. Probe Socket ■ Analog output is only available on MODEL6501-A0/-C0.
Touch Panel MENU Key Enters the Main Menu to select a function. *Press the MENU Key to CANCEL out of the current measurement or settings and return to the Main Menu. MENUaa1.NORMAL …………………… Normal Measurement Mode aaaaaa2.DUCTaTYPE …………… To input duct shape and size aaaaaa3.CALCULATION ……… Calculation Mode aaaaaa4.FLOWaRATE …………… To measure flow rate aaaaaa5.DATAaOUTPUT ……… To output stored data aaaaaa6.DATAaCLEAR ………… To delete stored data aaaaaa7.UTILITY ………………… To set the date, time and units aaaaaa8.
Probe There are 8 different types of probes available for CLIMOMASTER. The model number and specifications depends on the type of the probe you have. The compatibility feature of the probes allows you to swap out probes freely without having to recalibrate the main unit.
Probe Cable The probe can be extended by using a probe cable between the main unit and the probe. Probe Connector Approx.43 Approx. Φ14 Approx. Φ15 Main Body Connector Approx.43 Approx.2000 *Unit: mm Extension Rod (Optional) Straight Approx.130 Approx.Φ16 Approx.Φ23 Approx.Φ15 Cable length: Approx.2000 including the rod Approx.50 Approx.230 to 870 *Unit: mm Approx.Φ16 Approx.Φ23 Cable length: Approx.2000 including the rod Approx.Φ15 Flexible Approx.50 Approx.130 Approx.
§2 Getting Started Installing Batteries Backside of the body 1. Press down on the battery cover. 2. Slide the cover until it stops. 3. Lift the cover away from the body. 4. Insert the batteries according to the indicated polarity chart. This instrument requires six (6) AA size batteries that must be of the same type. DO NOT mix different types of batteries. Failure to observe this may cause battery leakage or damage the instrument. *Batteries CANNOT be recharged by the AC adapter.
Connecting the Probe and the Probe Cable This section describes how to connect the probe or the probe cable to the main unit. Make sure that the power is turned OFF when connecting or disconnecting the probe or the probe cable. Probe Probe No. Number 1. Align the connector of the probe or the probe cable with the probe terminal on the body. *When connecting the probe, make sure the probe number faces towards the front side of the main unit. 2.
Turning ON/OFF the Power The power switch to turn the power ON/OFF is located at the side of the instrument. When you turn the power on after connecting the probe/probe cable, a test screen with the Kanomax logo, model, and the version of the firmware will be displayed for a few seconds before it switches to the Normal Mode screen.
Precautions for Measurements Air Velocity ●MODEL6531/6541/6561: The probe has its own directivity characteristics. Make sure that the direction mark is facing into the airflow. If you are not sure of the airflow direction, slowly rotate the probe and select the point where you get the maximum velocity reading. Direction Mark Airflow ●MODEL6533/6542/6543/6551/6552: This omnidirectional probe has a horizontal, 360 degree dynamic characteristic (Refer to “Directivity of the Probe” on P.57 for details).
Pressure ■ Pressure sensor is only available on MODEL6501-B0/-C0. ●Do not expose the pressure sensor to more than 75kPa of pressure. Excess pressure may cause serious damage to the sensor. ●The operating temperature must be between 5oC and 40oC or 41oF and 104oF when measuring pressure. The instrument may not operate properly outside of this temperature range. ●Make sure to perform a zero adjustment before measuring pressure. When performing a zero adjustment, leave both pressure ports (+) and (-) open.
Duct Shape / Size Input Before measuring the volumetric flow rate, the duct shape and size settings must be entered. Up to 25 different duct types can be registered in this instrument. Select a duct type from the registered duct types to measure the air flow rate. MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO Press to enter the Main Menu. Use to select ”2. DUCT TYPE”. Press .
Registering a Rectangular Duct aaaaaaaDUCTaTYPE to select “3. WIDTH” and Use 1.ENTRYaNo.aaaaaaaa1 2.SHAPEaaaaRECTANGLE 3.WIDTHoooooooooo800 4.HEIGHTooooooooo500 5.UNIT(mm/inch)aaamm 6.SAVEaINFO press . Use width and press to enter the . to select “4. HEIGHT” and Use press . Use height and press to enter the . The maximum is 9999mm or 999.9 inch. Registering a Circular Duct aaaaaaaDUCTaTYPE Use 1.ENTRYaNo.aaaaaaaa1 2.SHAPEaaaaCIRCLE 3.DIAMETERaaaaaaa300 4.
§3 Normal Measurement NORMAL MODE This is the mode that you will be in, when you first turn on the instrument. In this mode you cannot save any data. The display is updated every 1 second. To move to NORMAL Mode from another measuring mode, press . Select “1. NORMAL” and press . MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.
Hold the Reading ■ Humidity is only available on MODEL6531/6533. '12/04/12a15:40:23 ~ TCC1 ■ Air temperature is only available on MODEL6531/6541 /6542/6533/6543/ 6561. 0.87f 25.3e 64.1g '12/04/12a15:40:23 ~ TCC1 HOLD 0.87f 25.3e 64.1g While measuring (in NORMAL Mode), press the key to hold the current reading. This is applicable in both the flow rate and pressure measurement screens as well. The ”HOLD” indicator appears on the display to indicate that the reading shown is held.
Setting the Time Constant ■ Humidity is only available on MODEL6531/6533. ■ Air temperature is only available on MODEL6531/6541 /6542/6533/6543/ 6561. ■ Pressure is only available on MODEL6501-B0/-C0. '12/04/12a15:40:23 ~ TCC1 0.87f 25.3e 64.1g In NORMAL Mode, you can change the Time Constant (TC) by pressing . Time Constant You can select the Time Constant from 1, 5 or 10seconds.
Changing the Time Constant Application ■ Pressure is only available on MODEL6501-B0 /-C0. When you first turn on the instrument, the Time Constant is only effective for AIR VELOCITY and AIR FLOW in Normal Mode. If you want to make it effective for AIR TEMPERATURE, HUMIDITY and PRESSURE, the following change must be made. MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO to enter the Main Menu.
§4 Measuring Max, Min, and Mean: CALCULATION Mode CALCULATION Mode will automatically calculate the maximum, minimum and mean of the measured data. The 1st trial TRIAL(1) The 2nd trial TRIAL(2) The Nth trial …………… TRIAL(N) Sampling Time Measuring Time……Sampling Time x N (Number of Trial) MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO Press to enter the Main Menu. Use to select “3.
CALCULATE R20000/20000 Select “1. MODE“ and press 1.MODE11111111AVERAGE 2.SAMPLING(TIME)0012s 3.SAMPLING(n)110050 4.DATA2STORAGE2?22YES 5.SET2TO2START Use . to select AVERAGE or INSTANT and press . CALCULATE R20000/20000 1.MODE11111111AVERAGE 2.SAMPLING(TIME)0012s 3.SAMPLING(n)110050 4.DATA2STORAGE2?22YES 5.SET2TO2START Use to select “2. SAMPLING (TIME)” and press Use . to select SAMPLING TIME (1 to 999 sec) and press .
■ Humidity is only available on MODEL6531/6533. ■ Air temperature is only available on MODEL6531/6541 /6542/6533/6543/ 6561. ■ Pressure is only available on MODEL6501-B0 /-C0. '12/04/12a15:40:23 ~ RDY. N2221 /2250 0.87f 25.3e 64.1g The instrument is ready for measuring. Press to change the applied parameters. (Air Velocity, Air Temp, Humidity – Flow Rate, Air Temp, Humidity – Pressure) Press to start. '12/04/12a15:42:57 ~ SMP . N2223 /2250 2.11f 25.3e 64.2g Press to stop.
§5 Measurement of Air Flow FLOW RATE Mode CLIMOMASTER features an accurate Flow Rate Mode which corresponds to the industry measurement standards such as ASHRAE. SINGLE FLOW RATE Mode This mode is useful for performing a measurement of the airflow of duct (i.e. a duct traverse). Following the steps listed below, you can program Climomaster with the following parameters: Sampling Time: The instrument will take a sample at the end of this entered sampling time.
Air Flow: Air Volume per Time Unit [m3/min, m3/h, ft3/min, ft3/h] Air Flow (Q) = Average Air Velocity (U) x Cross Sectional Area (A) Measurement Point r1 r2 r3 R r4 r5 Rectangular Duct r1=0.316R r2=0.548R r3=0.707R r4=0.837R r5=0.949R Circular Duct MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO SAMPLING MODE 1.SINGLEoFLOWoRATE 2.MULTIoFLOWoRATE Press to enter the Main Menu. Use to select “4.
SINGLE (R20000/20000 1.SAMPLING(TIME)001os 2.SAMPLING(n)110050 3.MEAS.2POINT220010 4.DATA2STORAGE2?22YES 5.DUCT2ENTRY2NO.2221 6.SET2TO2START Use to select “2. SAMPLING(n)”and press Use . to select No. TRIAL (1 to 9999) and press . SINGLE (R20000/20000 1.SAMPLING(TIME)001os 2.SAMPLING(n)110050 3.MEAS.2POINT220010 4.DATA2STORAGE2?22YES 5.DUCT2ENTRY2NO.2221 6.SET2TO2START Use to select “3. MEAS. POINT” and press Use . to select the number of MEAS.
'12/04/12a16:11:59 ~ 1 2 3 4 5 RDY. P 2221 N2221 D2223 C2300 The ready screen will be displayed. Use --6037a 25.1e 63.7g while pressing to change the entry no. of the duct type. Press to start measuring. Display Icons 1. RDY: Current Status (READY/SAMPLE) 2. P 1: Current Number of Points 3. N 1: Current Number of Trial 4. D 3: Selected Duct Entry Number 5. C300: Duct Shape and Size (In this case, the duct is circular and 300mm in diameter.
MULTI FLOW RATE Mode MULTI FLOW RATE is very similar to the SINGLE FLOW RATE MODE, except it adds an additional parameter for the number of locations. This allows you to take several air flow measurements (i.e. traverses) and get a total average, for example measuring flow rate from multiple ducts to get an overall average. The value of each point is the average of TRIAL(1) to TRIAL(N). The value of each location is the average of POINT(1) to POINT(M).
MULTI (R20000/20000 1.SAMPLING(TIME)001ps 2.SAMPLING(n)110050 3.MEAS./LOC.oo010/010 4.DATA2STORAGE2?22YES 5.DUCT2ENTRY2NO.2221 6.SET2TO2START If you press Use to select “6. SET TO START” and press . before saving the settings you will return back to the Main Menu and any setting changes you made will not be saved. The ready screen will be displayed. ① ② ③ ④ ⑤ Use '12/04/12a16:11:59 ~ RDY. P2221 N2221 L2221 C2300 ---502a 27.1e 62.
§6 Store and Recall Measured Data Data which can be stored in the memory of the instrument is shown below.
CALCULATION 22222DAPAGEOSET ……………………… PAGE:0001 MODE:CALCULATION(A) ………… DATE:2012/04/12 ………………… TIME:2213:40:53 ………………… MEAS./LOC.:2220/2ooo- ……… SHAPE:C0ooo- ……………………… Use to select the page that you would like to recall and press .
You can select the range of calculation. If the data range is correct, press to calculate the range of data displayed on the screen. START:00100END:050 NUM.0m/s0000°C000%RH 001000.810025.40064.7 002000.950025.40064.7 003000.980025.60064.9 004001.050025.70065.1 005001.210025.70065.0 006000.990025.90065.2 If the range is not correct, press display a cursor on “START”. Use to select the first number of the data (START) that is to be calculated and press .
§7 Print Out Preparation for Printing Connect the printer cable to the Serial Communication terminal located on the side of the instrument to print out measurement data. Serial Communication Terminal ●Printer (optional)……Recommended: Seiko Instruments Model DPU-S245 ●Printer Cable (optional) You need to enter the same baud rate and data transmission conditions on both the Main Body and the printer.
Print Out in CALCULATION Mode and FLOW RATE Mode aNEXCALCULATION MAX AVG MIN 102.76f 081.43f 060.81f Press after the measurement and calculation has been performed to print out the result. aNEXT1FLOW2RATE MAX AVG MIN 106433a 085134a 066797a Press after the measurement and calculation has been performed to print out the result.
Print Out Stored Data(Single Page) MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO DATA OUTPUT 1.DISPLAY 2.PRINTER(SINGLE) 3.PRINTER(MULTI) Press to enter the Main Menu. Use to select “5. DATA OUTPUT” and press . Use to select “2. PRINTER(SINGLE)” and press Use 22222DAPAGEOSET ……………………… PAGE:0002 MODE:SINGLEoFLOW2RATE……… DATE:2012/04/12 ……………… …………… TIME:2215:42:53 DATA/DIV.
AaaaaPRINT OUTPUT Press 1.RESULT 2.DATA 3.ALL Print Out. Use and press to select the contents of the to select from items 1 to 3 to print out. Please refer to the printing example shown below. 1. RESULT… 2. DATA…… 3. ALL……… Conditions and Calculation Results Conditions and Stored Data Conditions, Calculation Results and Stored Data Example of Print Out PAGE SET PAGE:001 MODE:CALCULATION(A) DATE:2001/05/19 TIME:11:32:26 DATA:010 SAMPLING TIME:001 START:001 END:010 MAX 0.75 m/s AVG 0.40 m/s MIN 0.
Print Out Stored Data(Plural Pages) MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO DATA OUTPUT 1.DISPLAY 2.PRINTER(SINGLE) 3.PRINTER(MULTI) Press to enter the Main Menu. Use to select “5. DATA OUTPUT” and press . Use to select “3. PRINTER(MULTI)” and press . ooDATAoOUTPUT(MULTI) 1.SELECToSTART:o0001 oooooooooENDoo:o0003 oooooooooSEToo:oooNO 2.ALLoDATAooooooooNO Use to select “1.
§8 Digital Output Preparation Serial Communication Terminal You can download the data stored in the CLIMOMASTER to your PC, by connecting the CLIMOMASTER to your PC with the USB cable. <Requirements> ● Personal Computer ● USB Cable (supplied) ● Communication Software (such as Hyper Terminal for Windows) <Connecting PC> 1. Connect the PC to the Climomaster using the USB cable. 2. Turn ON the CLIMOMASTER. 3. Make sure that the CLIMOMASTER is displaying NORMAL Mode.
Transmission of On-Time Data (Measured at every 1sec) 1. Type “D**** ”. *Type in the number of data(4 digits) after “D”. 2. After “AD”, the data will be displayed. 3. Each data represents 1 second of measurement. If you ask for 20 data, it takes approximately 20 seconds to display. 4. The maximum number that can be set is 9999. If you need more, re-send the command Output Content ■ ”0000000” will be displayed for humidity on MODEL6541/ 6542/6543/6561.
Transmission of Stored Memory 1. Type “K ” 2. After “AK”, the data will be displayed. 3. All stored duct settings (1 through 25) will be downloaded. Example Output Content Entry No. (Data Location); Width; Height; Diameter; Units of Duct Size Example AP 1. Type “P”. P0011 2. After “AP”, the data will be displayed. 3. Number of data stored pages will be downloaded. 1. Type “T**** ”.
1. Press “M**** ”. *Type in the desired page number(4 digits) after “M”. 2. After “AM”, the data will be displayed. ■ Pressure is only available on MODEL6501-B0 /-C0. Output Content A. Measuring Parameter WTH: Flow Rate; Temperature; Humidity WT: Flow Rate; Temperature VTH: Velocity; Temperature; Humidity VT: Velocity; Temperature PRS: Pressure B. Measuring Mode 000: Calculation Mode 001: Flow Rate Mode C. Sampling Time D. Number of Data E.
§9 Analog Output Analog Output ■ Analog output is only available on MODEL6501-A0/-C0. 1. Data Update……every 0.1 sec Terminal (Except Humidity; updated every 1sec) 2. Load Impedance……Above 5KΩ 3. Output Voltage……DC 0 to 1V For analog output, you must select one setting from the table below.
MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO Press to enter the Main Menu. Use to select “7. UTILITY”. Press . aaaaaaaUTILITY oooooooUTILITY Use 1.TIME AJUST 1.TIMEoADJUST 2.UNIT SELECT 2.UNIToSELECT 3.ANALOG OUTPUT 3.ANALOGoOUTPUT 4.TC(T.H.P)oSET 4.TC(T.H.P) SET 5.BACKLIGHToON/OFF to select “3. ANALOG OUTPUT”. Press Select “1.
§ 10 Other Settings Date MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO Press . Use to select “7. UTILITY” and . press oooooooUTILITY Use 1.TIMEoADJUST 2.UNIToSELECT 3.ANALOGoOUTPUT 4.TC(T.H.P)oSET 5.BACKLIGHToON/OFF to select “1. TIME ADJUST” and press .
Units and Baud Rate Velocity 1m/s = 196 FPM Temperature T(oF) = 1.8×T(°C)+32 Flow Rate 1m3/h = 35.32ft3/h MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO Press oooooooUTILITY Use to enter the Main Menu. Use Press 1.TIMEoADJUST 2.UNIToSELECT 3.ANALOGoOUTPUT 4.TC(T.H.P)oSET 5.BACKLIGHToON/OFF to select “7. UTILITY”. . to select “2. UNIT SELECT” and press .
Settings for LCD Backlight MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO Press oooooooUTILITY Use 1.TIMEoADJUST 2.UNIToSELECT 3.ANALOGoOUTPUT 4.TC(T.H.P)oSET 5.BACKLIGHToON/OFF “5. BACKLIGHT ON/OFF”. . Use Press Press to select “7. UTILITY”. . to select . Use ooooLCDoBACKLIGHT Press 1.BACKLIGHToON ooooFORoooooooo01omin to select “1. BACKLIGHT”. . Use to select BACKLIGHT ON/OFF. Press 2.
To Delete Data Press to enter the Main Menu. MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO Use to select “6. DATA CLEAR” and press . Use to select “1. CLEAR” (delete the selected data only) and press . AaaaaDATARCLEAR 1.CLEARLSTART:000001 00000000END00:000001 00000000SET00:002YES 2.
To Delete All the Data Press to enter the Main Menu. MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO Use Use Press t to select “6. DATA CLEAR”. . to select “2. ALL CLEAR”. Press . AaaaaDATARCLEAR 1.CLEARLSTART:000001 00000000END00:000001 00000000SET00:0020NO 2.
Contrast Adjustment If the LCD display of the CLIMOMASTER is too dark or too light, there is an adjustment dial at the back, bottom of the unit, inside the battery cover. Contrast Adjusting Dial You can adjust the contrast by using a precision screwdriver (-) (0.9 to 1.5mm). Turn the dial clockwise to darken the LCD and counterclockwise to lighten it.
§ 11 How to Clean a Probe Dust or particles attached to the velocity sensor would alter the amount of heat diffusion, which leads to a less precise reading. Also, deformation or clogging of the protective mesh around the sensor of the probe would also affect the accuracy of the instrument. Users are encouraged to clean the probe regularly to maintain the accuracy of measurements. Procedure Clean the sensor by soaking it in water in an ultrasonic cleaner for 10 to 20 seconds.
§ 12 Specification Product Model No. of the Main Body Model No. of the Probe Measuring Object Measuring Range Resolution Air Velocity [m/s] Accuracy Response Time [sec.] Air Temperature [oC] 6541 -21 6561 -21 ±2% of reading or ±0.015 (±3FPM), whichever is greater Approx. 1 Approx. 4 Approx. 7 - - - Response Time [sec.] Approx. 30 (at 1m/s, 90% response time) - Accuracy Response Time [sec.] Measuring Range Resolution Pressure ※2 Accuracy [kPa] Response Time [sec.
§ 13 Principle of Measurement Principle of Hot-wire Anemometer Wind Cooling -off Sensor (Platinum Coil) Current (i) i Air Velocity Sensor Current Current The principle of the thermal Probe is based on a heated element from which heat is extracted by the colder impact flow. The temperature is kept constant via a regulating switch. The controlling current is directly proportional to the velocity.
Wind Wind Temp Sensor (Platinum Resistor) Measurement of Wind-Temperature (MODEL 6531/6541/6542/6533/6543/6561) The coefficient of resistance of the temperature sensor has a direct proportional relationship with the temperature. We can measure the wind temperature by adjusting the sensor’s temperature to the wind-temperature, and measuring its coefficient of resistance.
What are Discomfort Index (DI) and Dew Point Temperature (DT)? **What is DI and DT?** ● Discomfort Index (DI) Discomfort Index is an index to show how uncomfortable it is in the summer time. CLIMOMASTER use the formula, shown below, which is used by Meteorological Agencies. DI 0.81T 0.01H (0.99T 14.3) 46.
About Compensation This instrument has been calibrated at normal temperature and pressure. Therefore, if you are going to measure velocity at a different temperature and pressure, the indicating value will be affected. Influence of Measuring Temperature This instrument has been calibrated at normal temperature. Because of theoretical fundamentals of hot-wire anemometers, it can easily be effected by the ambient temperature. To prevent such influence, temperature compensation is needed.
§ 14 Probe Directivity (Air Velocity) MODEL 6531-21 / 6541-21 / 6561-21 Horizontal Wind Direction Indicator -180° 90° 0° -90° At 5m/s Vertical Wind Direction Indicator 90° -90° At 5m/s 57
Probe Directivity (Air Velocity) MODEL 6542-21 Horizontal 0° -180° 90° -90° 90° -90° Vertical 0° -180° At 1m/s Probe Directivity (Air Velocity) MODEL 6533-21/6543-21/6551-21/6552-21 Horizontal -180° -90° 0° 90° 0° 90° -90° -180° At 1m/s 58 Vertical Probe Directivity (Air Velocity)
§ 15 Trouble Shooting Check battery Problem Possible cause(s) / Solution(s) Battery is inserted in wrong polarity. The instrument will not Turn off the power and insert the battery turn on correctly. Display contrast is not set up properly. The display is not visible Adjust the contrast volume switch. The battery is low. The battery level indicator Turn off the instrument and replace the flashes on the display batteries.
Output check (1) - Print out Problem The printer doesn’t operate. Cannot printout the display Want to cancel printing Possible cause(s) / Solution(s) The printer is not connected properly. Confirm the printer connection. The Baud Rate might not be set properly. Check both the instrument and printer settings. Make sure to use the optional printer (Model DPU-S245) and printer cable (Model 6000-31). Confirm the connection procedure.
§ 16 Warranty and After Service Kanomax Limited Warranty The limited warranty set forth below is given by KANOMAX JAPAN, Inc. (hereafter referred to as “KJI”) with respect to the KANOMAX brand anemometer, and its attachment parts including probe and other accessories (hereafter referred to as “PRODUCT”) purchased directly from KJI or from and authorized distributor.
§ 17 Contact Information U.S.A. KANOMAX USA, INC. PO Box 372, 219 Route 206, Andover, NJ 07821 U.S.A. TEL: (800)-247-8887 / (973)-786-6386 FAX: (973)-786-7586 URL: http://www.kanomax-usa.com/ E-Mail: info@kanomax-usa.com JAPAN KANOMAX JAPAN, INC. 2-1 Shimizu, Suita City, Osaka 565-0805, Japan TEL: 81-6-6877-0183 FAX: 81-6-6879-5570 URL: http://www.kanomax.co.jp E-Mail: sales@kanomax.co.jp CHINA Shenyang Kano Scientific Instrument Co., Ltd. #2610, 51 Wulihe Street Heping District Shenyang P.R.C.