List of Components ■ Standard Items MODEL QTY 6035-00 6035-A0 ANEMOMASTER Standard Main Body 6035-B0 6035-C0 6036-00 One of these 6036-A0 ANEMOMASTER Professional Main Body 6036-B0 6036-C0 Functions With analog output function Straight Probe With pressure To measure air velocity, air temp, and sensor With analog volumetric flow rate.
Important Safety Information The symbols for the warnings used in this manual are defined below. Classifications Danger: To Prevent Serious Injury or Death Warnings in this classification indicate danger that may result in serious injury or death if not observed. Caution: To Prevent Damage to the Product Warnings in this classification indicate risks of damage to the product that may void the product warranty if not observed.
Danger Never bring the probe close to a flammable gas atmosphere. The heated sensor may cause a fire or explosion. Forbidden Never touch the sensor. Don’t touch Hot surface The sensor is heated during operation. Touching the heated sensor may cause burns, and may also damage the sensor itself. Do not disassemble or heat the batteries, or put them into a fire. Use properly Explosive Caution This may cause burns and the batteries may burst.
Forbidden Do not use or leave the instrument in a high temperature, high humidity or dusty environment. Do not leave this instrument under direct sunlight for a prolonged period. The instrument may not function properly out of the specified operating conditions. Do not subject the instrument or the probe to strong impacts. Forbidden Dropping the unit or placing heavy objects on it may cause damage or malfunction to the instrument. Never disassemble, modify or repair the product.
Do not move the main unit and the probe from a cold place to a warm place quickly. It will cause dew condensation. Forbidden Even when used in an environment within the specified operating temperature and humidity, a sudden temperature change may cause condensation. Condensation generated on the sensor may cause inaccurate measurements. Condensation on metal parts may cause rusting and lead to a malfunction. Do not touch the LCD screen with a sharp-pointed object or with excessive pressure.
Table of Contents §1 Part Names and Functions .................................................... 7 Overview........................................................................................ 7 Main Body ...................................................................................... 8 Touch Panel .................................................................................... 9 Probe...........................................................................................
Part Names and Functions Overview Unit: mm Probe Approx.Φ6 Main Body Approx.294 to 1060 §1 Approx.
Main Body Unit: mm Approx.51 Analog Output Terminal ■ Analog output is available on MODEL6535-A0 /-C0 and MODEL6536-A0 /-C0 Approx.88 Approx.
Touch Panel CLEAR Key: To clear stored data. MODE Key: To change measurement mode. CALC Key: To obtain average, max, and min. *MODEL 6035-00/ 6035-A0/ 6036-00/ 6036-A0 is not equipped with ZERO Key STORE Key: To store current measurement data. *ZERO Key: To provide zero adjustment in pressure measurement mode ENTER/HOLD Key: 1. To put the current display on hold during a measurement. Press the key again to release hold. 2. To confirm current setting.
Probe Direction Mark 風向点 Air Velocity Sensor 風速素子 Temp. Sensor and 温度素子・温度補償素子 Temp. Compensation Sensor Approx.Φ6mm -How to extend the Articulating Probe- Caution: Do not excessively bend the flex-neck or apply excessive force. When the instrument is not in use, the probe must be returned to the original position and stored. 1. Hold the upper part 1 of the probe, and unscrew 2. 1 1 3 6 2 4 10 5 Part Names and Functions 2.
§2 Getting Started Installing Batteries Backside of the body 1. Press down on the battery cover with your finger as shown left. 2. Slide the cover toward the bottom of the instrument until it stops, 3. Lift the cover away from the body. 4. Insert the batteries according to the indicated polarity chart. The instrument requires six (6) AA size batteries. Types of batteries that can be used are: Manganese, Alkaline, Ni-Cd batteries, or Ni-MH batteries. The six (6) batteries must be of the same type.
Turning ON/OFF the Power The power switch to turn the power ON/OFF is located at the side of the instrument. When powered up, the LCD test screen will be displayed before it switches to the Air Velocity / Air Temperature measurement screen in approx. 2 seconds. |: Power on O: Power off Power Switch LCD Test Screen Power On Air Velocity / Air Temperature Measurement Screen Calibration Reminder One year after the last calibration, the Annual Calibration Reminder will display on the screen.
Battery Level Indicator Check the “Battery Level Indicator” to confirm the remaining battery level. The battery consumption rate largely depends on the measured air velocity. When the batteries need to be replaced (or recharged), the indicator will start blinking. When using a charging battery, the battery level may be indicated as low even right after recharging because of its low nominal voltage (1.2V). The screen may freeze if high velocity is measured when the battery level indicator is blinking.
Precautions for Measurements Air Velocity Measurement Precautions Direction Mark ● The probe has its own directivity characteristics. Make sure that the direction mark is facing into the airflow (for details of the directivity characteristics, refer to “Probe Directivity (Air Velocity)” on P.31).
Press key to enter the setting screen. Select with the and press the keys, key. Select with press the keys, and key. Set the height (H) with keys and press the key. Set the width (W) with keys and press the key. *Press and hold either of the keys over 2 seconds, and the increment speed will accelerate.
§3 Measurement When the instrument is turned on, the LCD test screen will be displayed for approx. 2 seconds. The test screen will then switch to the Air Velocity / Air Temperature measurement screen. The display is updated every 1 second. Changing the Measurement Mode To change the measurement mode, press the key while each measurement screen is displayed. The screen changes as shown below.
Setting the Time Constant When there is rapid change in the measurement data, the readings may become difficult to read. In such case, the speed of updating the readings can be reduced by changing the time constant setting. *Time constant can be set only for Air Velocity and Volumetric Flow measurements. Time Constant determines the time span of the moving average.
§4 Data Storage and Statistical Calculation Storing Measured Data ■ Data storage function is only available on MODEL6036 Measured data can be stored in the built-in memory of the instrument. The instrument can hold up to 1500 data records. When storing the measured data, average, maximum and minimum values will be calculated for the data group to be stored. Each data group is stored with a storage number (shown as DNo xxx) which starts from 001 for each mode.
●Data Storage and Statistical Calculation Press the key to execute the statistical calculation (Average, Maximum and Minimum) for the group of data records stored in the temporary memory. Display will switch in the sequence shown as follows (AVG MAXMIN) as you press the key.
Viewing and Deleting Stored Data ■ The function of viewing and deleting stored data is available on MODEL6036 The average, maximum and minimum value of the stored data can be viewed on the display and deleted. Data viewing and deleting procedure will be described below by taking an example of an operation in the “Air Velocity / Air Temperature Measurement Mode”. ●Viewing the Stored Data Press the key when the instrument is in the measurement mode in which the data is stored.
§5 Setting the Measurement Unit and Baud Rate List of Measurement Units: ●Air velocity: m/s, FPM ●Volumetric flow rate: m3/h, ft3/min, m3/min ●Air temperature: °C, °F ●Length: mm, inch Air Velocity: 1m/s=196FPM Air Temp: T (OF)=1.8×T(°C)+32 Volumetric Flow: 1 m3/h=35.32 ft3/h Length: 1inch=25.4 mm ●Measurement Mode Press the key to enter the unit setting mode. ●Length Unit The unit to be set will blink. Use Press the keys to select the unit.
§6 Data Output Printing Out the Measurement Data (MODEL6036) ■ Print function is only available on MODEL6036 To print out the stored measurement Serial Communication data, an optional printer and printer Terminal cable are required. The printer cable must be connected to the serial communication terminal located on the side of the instrument.
Printout Examples (Measurement Mode) Air Velocity / Air Temperature ●EXAMPLE Measurement Mode Volumetric Flow / Air Temperature Pressure/Air Temperature ●EXAMPLE (Rectangular duct) ●EXAMPLE Duct Shape R Size 2550*2550mm Vel Temp 0.15m/s 21.8°C FlowRate 407.3 m3/h Temp 22.5oC Press 1.03kPa Temp 20.1°C ●EXAMPLE (Circular duct) Duct Shape C Size 2550mm FlowRate 407.3m3/h Temp 23.
Printout Examples (Stored Data) Measurement Mode Air Velocity / Air Temperature Volumetric Flow / Air Temperature ●EXAMPLE for group DNo006 ●EXAMPLE for group DNo006 (When the duct shape is Rectangular.) DNo.006 001 Vel Temp 002 Vel Temp 003 Vel Temp 004 Vel 0.15m/s 21.8°C 0.16m/s o 21.8 C 0.14m/s o 21.8 C 0.13m/s 001 FlowRate 407.3 m3/h Temp 22.5oC 002 FlowRate 405.6 m3/h Temp 22.5oC o 003 FlowRate 400.9 m3/h Temp 22.5 C 004 FlowRate 401.4 m3/h Temp 22.5oC AVG FlowRate 403.8 m3/h Temp 22.
Analog Output Analog output is limited to the output of air ■ Analog output is velocity in the Air Velocity / Air Temperature available on MODEL 6035-A0/ Mode. 6035-C0/6036-A0/ 6036-C0. 1. Data update interval ……0.1sec Analog Output Terminal 2. Load impedance………..…5KΩ and above 3. Output voltage………….DC 0~3V (0~30m/s) 4.
§7 Cleaning the Probe When the sensor is contaminated with impurities such as dust, particles, soot or machine oil, the heat dissipation rate will change. In most cases, heat dissipation will decrease, resulting in lower air velocity readings. This is same for the probes which are equipped with a mesh cover. The same problem will occur if the mesh is deformed, or clogged with impurities.
§8 Specifications Product ANEMOMASTER Standard ANEMOMASTER Professional 6035 6036 Model Measuring Object Measuring Range (Resolution) Air Velocity Response Time 0.01 to 30.0 m/s (0.00 to 9.99m/s: 0.01m/s, 10.0 to 30.0m/s: 0.1m/s) Air Temperature -20.0 to 70.0 °C Pressure*2 -5.00 to 5.00 kPa (0,01 kPa) (0.1oC) 0 to 2550 mm 0 to 255 inch Duct Size Range Accuracy Clean Air Flow Air Velocity ±3% of reading or ±0.015 m/s (±3 FPM), whichever is greater Air Temperature ± 0.
§9 Measurement Principles Heat Dissipation [H] Current Hot-wire Anemometer Principle When the heated air velocity sensor is exposed to airflow, the sensor Current temperature will change by the heat drawn by the airflow. Cooling -off Accordingly, the sensor resistance value will change. This change in Wind the resistance value will vary largely as the air velocity increases.
- Temperature Compensation When the air temperature changes, the amount of heat dissipation will change accordingly even the air velocity is constant. Thus, Anemomaster employs a temperature compensation circuit to enable accurate air velocity measurement by eliminating the influence to the temperature change.
§ 10 Air Velocity Compensation When the heated air velocity sensor of the instrument is exposed to airflow, the heat is drawn from the sensor. The instrument obtains air velocity readings by using this relationship between the amount of heat removed (heat dissipation) and air velocity.
§ 11 Probe Directivity (Air Velocity) Horizontal Directivity Direction Mark -180° 90° -90° 0° At 5m/s Vertical Directivity 0° Direction Mark -90° +90° -180° At 5m/s 31
§ 12 Troubleshooting Problems Display does not appear when power is turned ON. blinks Reading is displayed as ”-----” Incorrect air velocity reading. High air temperature readings. Air velocity reading does not change from “0.0” when sampling in Volumetric Flow / Air Temperature measurement mode. Printing Failure Data Transfer Failure Analog Output Failure Incorrect Output Value 32 Troubleshooting Possible Cause / Solution Battery is inserted in wrong polarity.
Various Status Displays The number of stored data records has exceeded the maximum limit of 1500 data records. Stored data must be deleted to enable further data storage. The screen on the right is displayed during printing. When you want to cancel printing, press and hold the key.
§ 13 Warranty and After Service Kanomax Limited Warranty The limited warranty set forth below is given by KANOMAX JAPAN, Inc. (hereafter referred to as “KJI”) with respect to the KANOMAX brand anemometer, and its attachment parts including probe and other accessories (hereafter referred to as “PRODUCT”) purchased directly from KJI or from and authorized distributor.
§ 14 Contact Information U.S.A. KANOMAX USA, INC. PO Box 372, 219 Route 206, Andover, NJ 07821 U.S.A. TEL: (800)-247-8887 / (973)-786-6386 FAX: (973)-786-7586 URL: http://www.kanomax-usa.com/ E-Mail: info@kanomax-usa.com JAPAN KANOMAX JAPAN, INC. 2-1 Shimizu, Suita City, Osaka 565-0805, Japan TEL: 81-6-6877-0183 FAX: 81-6-6879-5570 URL: http://www.kanomax.co.jp E-Mail: sales@kanomax.co.jp CHINA Shenyang Kano Scientific Instrument Co., Ltd. #2610, 51 Wulihe Street Heping District Shenyang P.R.C.