Electromagnetic Compatibility Z370584 Rev. C
12-6
Specifications
Table 12.6 – Guidance and Manufacturer's Declaration: Electromagnetic Immunity
The VERITAS Vision System is intended for use in the electromagnetic environment specified below. The
customer or the user of the VERITAS Vision System should ensure it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance Level Electromagnetic
Environment: Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
(*)
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/ burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
±1 kV for input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV Differential
mode
±2 kV common mode
±1 kV
differential mode
±2 kV
common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
Voltage Dips 30%
reduction, 25/30
periods
At 0°
Voltage Dips > 95%
reduction, 0.5 period
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
Voltage Dips > 95%
reduction, 1 period At
Voltage Interruptions
> 95% reduction, 250/
300 periods
Voltage Dips 30%
reduction, 25/30
periods
At 0°
Voltage Dips > 95%
reduction, 0.5 period
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
Voltage Dips > 95%
reduction, 1 period At
Voltage Interruptions
> 95% reduction, 250/
300 periods
Mains power quality should be
that of a typical commercial or
hospital environment. If the user of
the VERITAS Vision System requires
continued operation during power
mains interruptions, it is
recommended that the VERITAS
Vision S
ystem be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m Power-frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.