Product Info

158
17 Changing Settings
provider prior to changing the Pod Shut-O setting. is feature requires active
participation on your part.
To enable or disable Pod Shut-O:
1. Navigate to: Menu icon ( ) > Settings > Reminders > Pod Shut-O.
2. Tap the Pod Shut-O toggle to enable or disable the Pod Shut-O feature.
3. If Pod Shut-O is enabled, tap the Inactivity Timer eld and select the length
of time for the countdown timer. is setting can range from 1 to 24 hours.
Example: If you choose 10 hours, you must wake up your Omnipod 5 app
and unlock it at least once every 10 hours, day and night, to prevent the Pod
Shut-O alarm.
4. Tap SAVE.
Check BG after bolus
Turn ON the Check BG aer Bolus reminder if you want a reminder to check
your blood glucose aer you deliver a bolus. If BG reminders are ON, you can
dene the timing of the BG reminder at the time that you give a bolus.
To enable BG reminders:
1. Navigate to: Menu icon ( ) > Settings > Reminders.
2. Tap the Check BG after Bolus toggle to enable or disable the Check BG aer
Bolus reminder.
Missed bolus
If a meal bolus or a manual bolus are not delivered during the Missed Bolus time
period, the Omnipod 5 app vibrates or beeps and displays a reminder. You can
set time intervals from 30 minutes to 4 hours. You can set up to six Missed Bolus
reminders per day.
Note: To add, edit, or delete Missed Bolus reminders, the Missed Bolus reminder
toggle must be set to ON.
Enable or disable all Missed Bolus reminders
To enable or disable the ability to use Missed Bolus reminders:
1. Navigate to: Menu icon ( ) > Settings > Reminders.
2. Tap the Missed Bolus toggle to enable or disable all Missed Bolus reminders.
If you disable these reminders, the Omnipod 5 app saves any previously set
reminders for later use.
Settings
History
Updates
Effective Date: 22DEC2020, PCO-000267