User manual

For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration - electromagnetic emissions
The 4508 is intended for use in the electromagnetic environment specied below.
The customer or the user of the 4508 should assure that it is used in such an environ-
ment.
Emissions test Compliance
Electromagnetic
environment-guidance
RF emissions
CISPR 11
Group 1
The 4508 uses RF energy only for its inter-
nal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electro-
nic equipment
RF emissions CISPR 11 Class B
The 4508 is suitable for use in all
establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not
applicable
Voltage uctuations/
icker emissions IEC
61000-3-3
Not
applicable
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration - electromagnetic immunity
The 4508 is intended for use in the electromagnetic environment specied below.
The customer or the user of the 4508 should assure that it is used in such an environ-
ment.
IMMUNITY test IEC 60601 test
level
Compliance
level
Electromagnetic
environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
oors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
not applicable not applicable
(For INTERNALLY POWERED
ME EQUIPMENT)
ELECTROMAGNETIC COMPATIBILITY INFORMATION