Guidance and manufacture’s declaration - electromagnetic emissions
The 4508 is intended for use in the electromagnetic environment specied below.
The customer or the user of the 4508 should assure that it is used in such an environ-
ment.
Emissions test Compliance
Electromagnetic
environment-guidance
RF emissions
CISPR 11
Group 1
The 4508 uses RF energy only for its inter-
nal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electro-
nic equipment
RF emissions CISPR 11 Class B
The 4508 is suitable for use in all
establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not
applicable
Voltage uctuations/
icker emissions IEC
61000-3-3
Not
applicable
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration - electromagnetic immunity
The 4508 is intended for use in the electromagnetic environment specied below.
The customer or the user of the 4508 should assure that it is used in such an environ-
ment.
IMMUNITY test IEC 60601 test
level
Compliance
level
Electromagnetic
environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
oors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
not applicable not applicable
(For INTERNALLY POWERED
ME EQUIPMENT)
ELECTROMAGNETIC COMPATIBILITY INFORMATION