Install Instructions

P7660/P7645 SERIES DRY PRESSURE SENSORS
Honeywell Building Technologies
In the U.S.:
Honeywell
715 Peachtree Street NE
Atlanta, GA 30308
customer.honeywell.com
® U.S. Registered Trademark
©2020 Honeywell International Inc.
31-00422—01 M.S. 09-20
Printed in United States
TROUBLESHOOTING
RECYCLING
Product should not be thrown away in regular trash.
Instead, it should be recycled according to local
municipality.
SPECIFICATION
Power supply: 12-30 VDC/24 VAC
1
, 30 mA max.
Outputs:
P7660: 3-wire 0-5 V/10 V, 4-20 mA loop powered, 3-wire
4-20 mA (Selectable)
P7645: 4-20 mA loop powered
Output Scaling:
P7660 Selectable Ranges:
0-5” (0.1/0.25/2.5/5.0 “W.C.)
0-10” (1.0/2.5/5.0/10.0 “W.C.)
P7645 Models: 0-0.1” 0.-0.25”, 0-1.0”, 0-5.0”
Operating Temperature:
Operating Range: -4°F to 185°F (-20°C to 85ºC)
Compensated Range: 32°F to 122°F (0°C to 50°F)
Media Compatibility: Dry, oil-free air. Nitrogen.
Sensor Type:
Precision calibrated: MEMS silicone piezoresistive
Accuracy:
Combined linearity and hysteresis: +/- 1.0% of selected
range
BFSL: +/- 0.25% BFSL
Auto-zero input: Push-button and contact closure input
Enclosure:
Material: ABS
Dimensions: 4.75”h x 1.82”w x 0.90”d
Optional Probe: P7600PROBE
Table 1.
Symptom Solution
No output Check wiring. Ensure power supply meets
requirements.
Pressure reading error
Verify control panel software is configured for
correct output scaling.
Verify switch settings
Verify tubing is not pinched or leaking.
Possible contamination. Ensure sensor is used
only on dry air or nitrogen.
Table 2. Specification by Product
Product number
Pressure
range Output
LCD
display Probe
P7645C0010X 0-0.10 “W.C. 4-20 mA
Loop
No No
P7645C0025X 0-0.25 “W.C. 4-20 mA
Loop
No No
P7645C0100X 0-1.0 “W.C. 4-20 mA
Loop
No No
P7645C0500X 0-5.0 “W.C. 4-20 mA
Loop
No No
P7660P0500D 0.1/0.25/2.5/
5.0 “W.C.
Universal Yes No
P7660P0500X 0.1/0.25/2.5/
5.0 “W.C.
Universal No No
P7660P1000D 1.0/2.5/5.0/1
0.0 “W.C.
Universal Yes No
P7660P1000X 1.0/2.5/5.0/1
0.0 “W.C.
Universal No No
P7660D0500D 0.1/0.25/2.5/
5.0 “W.C.
Universal Yes Yes
P7660D0500X 0.1/0.25/2.5/
5.0 “W.C.
Universal No Yes
P7660D1000D 1.0/2.5/5.0/1
0.0 “W.C.
Universal Yes Yes
P7660D1000X 1.0/2.5/5.0/1
0.0 “W.C.
Universal No Yes
Table 2. Specification by Product
Product number
Pressure
range Output
LCD