04973 Series Conductivity Cells Installation and Maintenance Manual 70-82-25-17 10/07 Honeywell Field Solutions
Copyright, Notices, and Trademarks Printed in U.S.A. – © Copyright 2007 by Honeywell Inc. Revision– 10/07 Warranty/Remedy Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Contact your local sales office for warranty information. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace without charge those items it finds defective.
About This Document Abstract The purpose of this manual is to support the installation and maintenance of the 4973 Series Conductivity Cells. Revision Notes The following list provides notes concerning all revisions of this document. Rev. ID Date Notes 0 10/96 This document is the initial release of the Honeywell version of the 4973 Series Conductivity Cell Installation and Maintenance Manual. This publication was originally released under the L&N system as 277067 Rev. N1.
Contacts World Wide Web http://hpsweb.honeywell.com Telephone Contact us by telephone at the numbers listed below. Organization Phone Number United States and Canada Honeywell 1-800-423-9883 iv Installation and Maintenance Manual Tech.
Contents 1. INTRODUCTION ................................................................................................... 1 1.1 Overview ........................................................................................................................................ 1 1.2 Description ..................................................................................................................................... 1 2. SPECIFICATIONS..........................................................
Figures Figure 1-1 4973 Type Conductivity Cells, for Flow and Insertion-type Applications.
Introduction 1. Introduction 1.1 Overview The 4973 Type cells, Figure 1-1, have a rugged configuration for reliable, continuous measurements of electrolytic conductivity in industrial water processes at temperatures up to 140°C at 250 psig.
Introduction Potted Potted Connector Connector Housing Housing * 1 1/8" Hex Across Flats 3/4 NPT Insertion Depth (See Specifications) Flow Channel Molding Flow Holes Graphite Electrodes * This is the integral cable option. Quick disconnect option also available. Titanium Electrodes Teflon Sheath FLOW Cell Constant FLOW 0. 1 0.0 1 4973 Series Cells Only 1.
Introduction Process Preferred Cell Locations Cool er Pump a/n 23340 Figure 1-2 Recommended Locations for Mounting a Conductivity Cell *Maximum allowable operating temperature for piping arrangements depends on the type of pipe tee chosen.
Specifications 2. Specifications 2.1 Specifications for 04973 Series Parameter Description Cell Constant 04973 Series: 0.01, 0.1, 1.0 and 10 as specified Electrode Material 0.01 and 0.1 constant, titanium 1.
Installation 3. Installation 3.1 General Requirements Observe the following before installing a conductivity cell. Specific requirements for particular types of installation are given in Sections 3.3 and 3.4. • Do not remove the Teflon sheath on 1or 10 constant cells, as this will change the cell constant value. • Do not use the cell in solutions which can affect the fittings or the cell materials. If in doubt, contact Honeywell. • Avoid all chlorinated hydrocarbons. Titanium and PES (0.01 and 0.
Installation 3.3 In-Line Type Mounting In addition to the General Requirements outlined above, note the following with regard to insertion-type mounting: • Make certain the liquid head is above the cell location during measurement. A vertical insertion (from above) or a horizontal insertion can be used. • Allow at least one-half inch clearance beyond the end of the cell and 1/8 to 3/16 inch radius clearance to permit circulation of the solution.
Installation 3.5 Torque Recommendations For inserting a cell in metal fittings or bushings - 40 ft-lb maximum. For inserting a cell in plastic fittings or bushings - 10 ft-lb maximum. Always use pipe sealant (preferably Teflon tape). 3.6 Dimension Drawings 1 1/ 8" (38mm) Hexagon Flow Out 1 1/2" (38mm) 3/4" NPT (Female) 1 1/2" (38mm) Octagon 3/4" NPT (Male) IN CELL 1 1/2" (38mm) Flow Chamber Flow In 8 3/4" (222mm) 14 1/2" max (368mm) 3/4" Fitting Allow 4 1/8" (105mm) for re moval of cell Notes: 1.
Installation 4.25 107.9 8-32 UNC – 2B 250 DEEP 1.0 25.4 .875 22.22 .50 12.7 1.5 38.1 3.75 95.25 4.0 101.6 1/8 NPT 4.5 114.3 3/4 NPT Figure 3-2 Dimension Drawing for 31079198 Flow Chamber 0.593" Dia. (15mm) "X" See Table Dim "X" Table I Inch mm 0.01 3.3 84 0.1 66 2.6 1.0 89 3.5 10 3.5 89 5.875" (149.2mm) 3/4” (76mm) 3/4 NPT 3/4" 1.13" Hex (29mm) ¾” female NPT for user's flexible electrical conduit connection.. For insertion or connection removal of cell, disconnect conduit connections.
Installation “X” ± 0.031 ± 0.787 2.771" ±.035 See Table "Z" (154mm ±.889) "Y" 3/4" NPT 1 .1" Dia. (27.9mm) 1.13" Hex (29mm) Suffix A 0.01 00.1 0.1 1.0 10 Dim "X" mm Inch 3.3 84 66 2.6 89 3.5 89 3.5 Dim "Y" Inch mm .703 17.85 .703 17.85 .593 1 5.06 .593 1 5.06 Cable iscable approx. 0.250" Internet option is (6.4mm) O.D. with 4 (4973 SeriesO.D. Cells) approx. 0.250” (6,4mm) or 64 (DL4000 DirectLine with 4973 Series Cells Cells) conductors Dim "Z" mm Inch .542 13.76 .542 13.76 .625 15.87 .
Electrical Connections 4. Electrical Connections 4.1 Overview The terminal board connections for the various Honeywell measuring instruments are given in the appropriate Figures in this section. To avoid the possibility of AC pickup in the cell leads, separate them from all AC line-voltage wiring or run them in a separate grounded conduit. ATTENTION Do not use shielded cable except where shown in the following figures. 4.2 Instrument Wiring for 4973 Cells with Integral Cable 4.2.
Electrical Connections 50 YELLOW COAX 20 OR 50 FT. CABLE LENGTH. Direct Cell to Analyzer Installation 50 Cell to Analyzer through Junction Box NOTES 5. FOR PURE WATER SAMPLES IN NON-CONDUCTIVE (PLASTIC, GLASS, ETC.) PIPING, GROUND THE BLACK CELL ELECTRODE LEAD NEAR THE CELL. ALTERNATIVELY, CONNECT TO THE UDA GROUND SCREW AS SHOWN DOTTED. DO NOT GROUND 10, 25, OR 50 CONSTANT CELLS. 6. FOR CELL LEADS BLACK AND WHITE, USE 16 TO 22 AWG CABLE, SHIELDED TWISTED PAIR, WITH 30 pF MAX.
Electrical Connections 4.2.
Electrical Connections 4.3 Instrument Wiring for 4973 Cells with Quick Disconnect Cable 4.3.1 Wiring Model 4973 with Quick Disconnect Cable to UDA2182 Wire Color Signal Name Yellow 10 Cell Low 9 8 7 Coax Cell High 6 RTH 3rd Wire Green 5 RTH Low Red 4 RTH High Brown 3 Blue 2 Jumper Wire to chassis ground screw 1 Earth Ground Figure 4-4 Wiring Diagram for 4973 Cells with Quick Disconnect Cable Connected to UDA2182 Analyzer 4.3.
Electrical Connections YELLOW NOTE: IGNORE BLUE AND BROWN WIRES. COAX 4.3.3 Wiring Model 4973 with Quick Disconnect Cable to APT COAX NOTE: IGNORE BLUE AND BLACK WIRES.
Maintenance 5. Maintenance 5.1 Overview The only maintenance that may be required is occasional cleaning. When cleaning, avoid scratching electrode surfaces. Do not use a brush or pipe cleaner. 5.2 To Clean the Cell The cell will require cleaning if sludge, slime, etc., accumulate in the flow channels. Since the materials of construction are chemically inert, chemical agents may be used and are recommended for cleaning the cells.
Maintenance 5.5 Air Entrapped in Cell Flow Channel If measurement errors appear for horizontal mountings of a 1 or 10 constant cell, it may be that air is entrapped in the cell flow channel. Take one of the following actions to eliminate this problem: 16 • Increase flow to at least 1 gpm. • Rotate the cell mounting so that its flow channel faces the same direction as the pipeline flow exit. • Install the cell and/or flow chamber vertically.
Accessories and Spare Parts 6. Accessories and Spare Parts Description Part Number Flow Cell Housing (PES) 055919 Flow Cell Housing (316 SS) 31079198 Junction Box 31316260 Teflon Shield White for 1 const. cell 31021599 Clear for 10 const. cell (see note below) 31018760 Extension Cables for Sensors with Quick Disconnect Option 2 m (6.56 ft.) 50024092-001 3 m (9.84 ft.) 50024092-002 6 m (19.69 ft.) 50024092-003 15 m (49.21 ft.
Accessories and Spare Parts 18 Installation and Maintenance Manual 10/07
Honeywell Field Solutions 512 Virginia Drive Fort Washington, PA 19034 70-82-25-17 10 07 Printed in USA http://hpsweb.honeywell.