Owner's manual

APPENDIX
131
131131
131
O/S Trimming Points
All-Frequency O/S Trimming Point List
The frequency of the extra trimming point is the measurement
frequency you were using before entering the calibration mode.
An “O/S Trim Error!” warning should not appear after running an
all frequency O/S trimming.
Model LCR-8101G LCR-8105G LCR-8110G
Trimming Point
13 points 17 points 22 points
1 20 Hz 20 Hz 20 Hz
2 100 Hz 100 Hz 100 Hz
3 300 Hz 300 Hz 300 Hz
4 1 kHz 1 kHz 1 kHz
5 3.3 kHz 3.3 kHz 3.3 kHz
6 10 kHz 10 kHz 10 kHz
7 20 kHz 20 kHz 20 kHz
8 50 kHz 50 kHz 50 kHz
9 100 kHz 100 kHz 100 kHz
10 200 kHz 200 kHz 200 kHz
11 500 kHz 500 kHz 500 kHz
12 1 MHz 1 MHz 1 MHz
13 Extra Trimming
Point
2 MHz 2 MHz
14 3 MHz 3 MHz
15 4 MHz 4 MHz
16 5 MHz 5 MHz
17 Extra Trimming
Point
6 MHz
18 7 MHz
19 8 MHz
20 9 MHz
21 10 MHz
22 Extra Trimming
Point
LCR-8000G Series User Manual
132
132132
132
100 kHz O/S Trimming Point List
The frequency of the extra trimming point is the measurement
frequency you were using before entering the calibration mode.
“O/S Trim Error!” warning will still appear after running a 100
kHz trimming if you choose to use a measuring frequency higher
than 100 kHz or other than where you were when you ran the 100
kHz trimming.
LCR Models LCR-8101G LCR-8105G LCR-8110G
Trimming Point
10 points 10 points 10 points
1 20 Hz 20 Hz 20 Hz
2 100 Hz 100 Hz 100 Hz
3 300 Hz 300 Hz 300 Hz
4 1 kHz 1 kHz 1 kHz
5 3.3 kHz 3.3 kHz 3.3 kHz
6 10 kHz 10 kHz 10 kHz
7 20 kHz 20 kHz 20 kHz
8 50 kHz 50 kHz 50 kHz
9 100 kHz 100 kHz 100 kHz
10 Extra Trimming
Point
Extra Trimming
Point
Extra Trimming
Point