Owner's manual

PASS-FAIL MODE
77
7777
77
Set frequency
Move the cursor to Freq column.
Enter the frequency using the
numerical keys and unit keys.
Func
Freq
500.00
Ls
2.00 V
Volt
Range
20Hz ~ 1MHz/5MHz/10MHz
5 digit resolution
Example:
0.5kHz (500Hz)
k
Enter
Set voltage
Move the cursor to Volt column.
Enter the voltage using the
numerical keys and unit keys.
Freq
Volt
2.00 V
500.00
Bias
Range
(1mV step)
10mV ~ 2V (DC or AC≤3 MHz)
10mV ~ 1V (AC>3 MHz)
Example:
100mV
m
V/A
Enter
Select data
capture rate
Move the cursor to Spd column.
Select the capture rate by
pressing F1 (Prog) repeatedly.
Bias
Spd
MAX
1.0000H
Hi
F 1
DC
AC≤
100Hz
AC≤
2kHz
AC>
2kHz
AC≥
1MHz
Slow
900ms 1.3s 600ms 600ms 620ms
Med
120ms 1.2s 470ms 450ms 470ms
Fast
60ms 650ms 180ms 150ms 150ms
Max
30ms 600ms 120ms 75ms 120ms
LCR-8000G Series User Manual
78
7878
78
Set Hi limit
Move the cursor to Hi column.
Enter the Hi limit using the
numerical keys and unit keys.
Spd
Hi
1.0000H
0.0000H
Lo
MAX
Range
follows the specification for each
measurement item
Example: 1.5kH
(for Ls)
Enter
H
k
Set Lo limit
Move the cursor to Lo column.
Enter the Lo limit using the
numerical keys and unit keys.
Spd
Hi
0.0000H
Lo
MAX
1.0000H
Range
follows the specification for each
measurement item
Example: 1.0kH
(for Ls)
Enter
H
k
Set trigger delay
Move the cursor to Dly (Delay)
column. Enter the amount of
trigger delay time using the
numerical keys and unit keys.
Hi
Lo
10 ms
Dly
1.0000H
0.0000H
Range
0 (no delay) ~ 1000ms
Example: 10ms
m